Chapter 2 Operating Your Universal Counter
Using the Save and Recall Menus
Press Enter key to complete the 1.5V entry.
Press 50 /1M
key to set input impedance to 50 .
Press DC/AC key to set input coupling to DC.
To save these settings, which make up your measurement setup, simply
perform the following steps:
a. Press Save & Print key.
SAVE:
NO is displayed.
b. Press the appropriate arrow key until SAVE:
c. Press Enter key.
The measurement setup is now saved in register 1. To recall this
measurement setup, perform the recall procedure in the following
section.

To Use the Recall Function

This procedure describes how to RECALL the measurement setup you've
saved in the previous procedure.
You cannot save a setup to register 0 (RECALL
automatically saves the current setup to register 0 before executing a
recall function. You can, however, recall the setup in the register 0 if you
need to, which is the purpose of this "backup" storage register.
Cycle POWER key to preset the Counter.
Note the Counter is set to measure the frequency of a signal applied to
channel 1 as indicated by the Freq and Ch1 annunciators.
Press Recall key until RECALL 1 is displayed, and press Enter key or
let the Counter time out.
Note that the measurement setup has been recalled as indicated by the
Period and Ch1 annunciators, and 50
2-54
1 is displayed.
0). The Counter
and DC LEDs.
Operating Guide