Chapter 3 Specifications
Measurement Specifications (Continued)
Measurement Specifications (Continued)
Time Interval
Measurement is specified over the full signal ranges
Results Range:
LSD:
RMS Resolution:
t
res
Systematic Uncertainty: ±(Time Base Error TI) ± Trigger Level Timing Error ± 1.5 ns Differential Channel Error (HP 53131A)
Time Interval Delay
After a Time Interval Measurement has begun by satisfying the trigger conditions on Channel 1, the instrument will wait for the
user-entered delay time to elapse before the end-of-measurement trigger will be accepted on Channel 2. Please refer to Measurement
Arming for additional information.
Ch1
Frequency Ratio:
Ch2
Measurement is specified over the full signal range of each input.
Results Range:
'Auto' Gate Time:
LSD:
1
Ratio 1 2 :
---------------------------------------------------------- -
Ch2 Freq Gate Time
Ch2 Freq
Ratio 2 1 :
------------------------------------------------------------------
2
Ch1 Freq
Gate Time
RMS Resolution:
2
1
Ch1 Freq Ch2 Trigger Error
+
Ratio 1 2 :
------------------------------------------------------------------------------------------------------------ -
Ch2 Freq Gate Time
2 Ch2 Freq
Ratio 2 1 :
------------------------------------------------------------------------------------------------------------------------------------------- -
Ch1 Freq
To minimize relative phase measurement error, connect the higher frequency signal to Channel 1 when possible.
7
See Specifications for Pulse Width and Rise/Fall Time measurements for additional restrictions on signal timing characteristics.
Operating Guide
7
of Channels 1 and 2.
5
–1 ns to 10
s
53131A
53132A
------------------------------------------------------------------------------------------- -
500 ps
150 ps
2
2
t
Start Trigger Error
Stop Trigger Error
+
+
re s
53131A
53132A
------------------------------------------------------------------------------------------- -
750 ps
300 ps
Ch1
Ch2
Ch3
–10
11
10
to 10
100 ms (or sufficient cycles on Channel 2 or 3 to make a valid measurement, whichever is longer)
2
1
+
Ch1 Freq Ch2 Trigger Error
2
Gate Time
2
± 900 ps Differential Channel Error (HP 53132A)
Ch3
Ch1
Ch1
1
Ratio 1 3 :
---------------------------------------------------------- -
Ch3 Freq Gate Time
Ch3 Freq
Ratio 3 1 :
------------------------------------------------------------------
2
Ch1 Freq
2
1
Ch1 Freq Ch3 Trigger Error
+
Ratio 1 3 :
------------------------------------------------------------------------------------------------------------ -
Ch3 Freq Gate Time
2
2 Ch3 Freq
Ratio 3 1 :
------------------------------------------------------------------------------------------------------------------------------------------- -
Gate Time
2
1
+
Ch1 Freq Ch3 Trigger Error
2
Ch1 Freq
Gate Time
2
3-9