P210 mA Setpoint Parameter
Used to set the process level that corresponds to the 4 mA value. 4 mA always defaults to
0, and P201 determines whether this is a level, space, or distance measurement.
P211 mA Setpoint Parameter
Used to set the process level that corresponds to the 20 mA value. 20 mA always defaults
to 100%, and P201 determines whether this is a level, space, or distance measurement.
P212 mA Output Minimum Limit
Values 2 to 22.00
P213 mA Output Maximum Limit
Values 2 to 22.00
P214 4 mA Output Trim
Used to calibrate the mA output for 4 mA. The device mA output is pre-calibrated;
however, this parameter can be used to trim remote displays or inputs.
Values
Related Parameters
Steps:
1. Set P201 to 0 (manual).
2. Set P911 to 4 mA.
3. Record remote reading in mA.
4. Enter this value in P214.
5. Set P201 to previous setting.
P215 20 mA Output Trim
Used to calibrate the output for 20 mA. The device mA output is pre-calibrated; however,
this parameter can be used to trim remote displays or inputs.
Values
Related Parameters
Steps:
1. Set P201 to 0 (manual).
2. Set P911 to 20 mA.
3. Record remote reading in mA. Attach calibrated meter.
4. Enter this value in P215.
5. Set P201 to previous setting.
7ML19985CL02
Range: 0 to 22.00. Display P911
P215 20 mA Output Trim
0 to 22.00. Display P911.
Range:
P214 4 mA Output Trim
SITRANS LR 300 – INSTRUCTION MANUAL
Page 65