S-parameter test set
Table of Contents

Advertisement

Specifications
Table 4-2
HP 8510/HP 8517B Supplemental Characteristics (Continued)
Connector type
Recommended connection torque
Damage input level
Nominal connector-nut size
User 1 Power Levels for Reference Channel Phase Lock
Minimum power level
Maximum power level
HP 8517B Power Requirements and Physical Characteristics
Operating temperature range
Input Power (AC)
Dimensions
Weight
1. This must be AC coupled.
4-3 HP 8517B S-Parameter Test Set Manual
Rear Panel RF Connectors
precision 2.4 mm female
90 N-cm (8 in-lb), maximum for precision-2.4 mm hardware
1
15 dBm
8 mm
–45 dBm
–10 dBm to –20 dBm, depending on frequency
0 C to 55 C
110, 120, 220, or 240 Vac, ±10%
Line frequency from 47 to 66 Hz
460 mm (High)
133 mm (Wide)
(18.1 inches H
5.25 inches W
15 kg (35 lb), net
609 mm (Deep)
24 inches D)