Specifications; Mechanical Specifications; Supplemental Characteristics - HP 8517B Operating & Service Manual

S-parameter test set
Table of Contents

Advertisement

4
Table 4-2
Connector type
Connector torque
Impedance
Damage-input level
DC bias
Nominal source operating power level
Nominal connector-nut size

Specifications

Specifications describe the warranted performance of the instrument. The
electrical specifications of the test set when used with an HP 8510 network
analyzer, are defined in the HP 8510C On-Site Service Manual.

Mechanical Specifications

Table 4-1 HP 8510/HP 8517B Mechanical Specifications
Test Port Description
Connector Type:
Precision, rugged male test-port connector
2
Center Pin Recession
1. Precision rugged make test-port connector.
2. Refer to the calibration kit manual to learn about measuring the pin depth.

Supplemental Characteristics

Supplemental characteristics listed in Table 4-2 provide useful information
by giving typical, but non-warranted, performance parameters.
HP 8510/HP 8517B Supplemental Characteristics
Front Panel Test Ports
precision 2.4 mm, male
90 N-cm (8 in-lb), maximum
50
nominal
At Port 1 and Port 2 :
At Port 2 with Option 004:
500 mA, 40 Vdc, maximum
2 dBm to –20 dBm,
with flatness correction off, depending on frequency
20 mm
Specification
1
2.4 mm
+0.00254 mm to +0.0178 mm
+0.0001 inches to +0.007 inches
+17 dBm CW, RF
30 dBm CW (1 watt) CW, RF
HP 8517B S-Parameter Test Set Manual 4-2
Specifications