Note that the Upgrade Wizard automatically creates a log file, logfile.txt, in the
directory from which it was run. Rename this log file and save it together with the
Upgrade Wizard Report disk file in another directory.

Step 7. Verify and Configure Important ITS Settings via the APC Interface.

After adding the remaining Access Point Controllers to the wireless subnet, several
important ITS settings must be verified and configured via the APC web interface to
ensure proper operation of the system. These include:
• Verify Filter Settings
• Verify BootP Address Ranges
• Configure the AP Defaults (for 1.4 GHz or 2.4 GHz APs)
• Configure AP Groups
• Configure Basic Settings for each Group
• Configure Alerts for each Group

Verifying the Filter Settings

To verify the network filter settings:
1. In View Device tree, select System and then select the Filters tab.
The Filter Configuration screen (Figure 3-10) appears.
2. Verify that the following network filter settings are configured as follows:
Protocol Type Filters:
• IP/ARP - UNCHECKED
• All others - CHECKED
Multicast Address Filter:
• Enable Multicast Filter - CHECKED
• Multicast Address - 01:00:5E:00:00:00
• Multicast Mask - ff:ff:ff:ff:ff:ff
ARP Filter:
• Enable ARP Filter - UNCHECKED
• Network IP Address - 0.0.0.0 (N/A)
-
IPX Broadcast Filters:
• All filters - CHECKED
• Broadcast Bandwidth Allocation - 20%
3. If you made any changes to the Network filter settings, click SAVE and then
verify the results.