Siemens sitrans PROBE LU Instruction Manual page 39

2-wire loop powered ultrasonic transmitter for level, volume and flow monitoring of liquids
Hide thumbs Also See for sitrans PROBE LU:
Table of Contents

Advertisement

P050 Value
0
*
1
2
3
4
5
6
7
8
9
7ML19985HT01
Vessel Shape
----
A
L
SITRANS Probe LU – INSTRUCTION MANUAL
Description
no volume calculation required N/A
flat bottom
conical or pyramidal bottom
parabolic bottom
spherical bottom
angled bottom
flat end cylinder
parabolic end cylinder
sphere
universal linear
level/volume/flow breakpoints
Also required
P051
P051, P052
P051, P052
P051, P052
P051, P052
P051
P051, P052, P053
P051
P051, P054, P055
Page 33