Rate (P700 And P701) - Siemens sitrans PROBE LU Instruction Manual

2-wire loop powered ultrasonic transmitter for level, volume and flow monitoring of liquids
Hide thumbs Also See for sitrans PROBE LU:
Table of Contents

Advertisement

P664 Temperature
Displays the temperature (in
Values

Rate (P700 and P701)

These parameters determine how material level changes are reported.
P700 Maximum Fill Rate
Allows you to further adjust the SITRANS Probe LU response to increases in the actual
material level (or an advance to a higher Failsafe Material Level, P071). P700 is
automatically updated whenever Measurement Response (P003) is altered.
Values
Altered by
Related
Enter a value slightly greater than the maximum vessel-filling rate, in Units (P005) or
percent of Span (P007) per minute.
P003 Value
1
2
3
P701 Maximum Empty Rate
Adjusts the SITRANS Probe LU response to decreases in the actual material level (or an
advance to a lower Failsafe Material Level, P071). P701 is automatically updated
whenever Measurement Response (P003) is altered.
Values
Altered by
Related
7ML19985HT01
o
C) read from the transducer.
Range
Range
Default
P003 Measurement Response
P005 Units
P007 Span
P071 Failsafe Material Level
Meters/Minute
0.100
1.0000
10.000
Range
Default
P003 Measurement Response
P005 Units
P007 Span
P071 Failsafe Material Level
SITRANS Probe LU – INSTRUCTION MANUAL
o
–40 to 85 (
C)
0.0000 to 99999 m / min.
0.100
0.0000 to 99999 m / min.
0.100 (m)
Page 49

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents