Operation Modes

6.2 Manual Mode

Do display the Manual Mode screen, touch MENU→ MANUAL.

■ Checking Status

■ Setting the Light Modulation Ratio

18
The Manual screen consists of the following elements:
READY
The status of the main unit (will be either READY,
NO LED, or LED ON)
1 2 3 4
The LED status (inverted: connected; lit: not
connected; flashing: irradiating)
During irradiation, the LED status changes as follows:
Slowly flashing: Normal use at stable temperature
Slightly faster flashing: Temperature at warning
level (when using at higher output)
Rapidly flashing: Temperature a t danger level
(output must be lowered; Warning Screen appears)
1 100%
The light-modulation ratio setting (enabled when the
number is touched and inverted).
0.0s
The irradiation time
WARN
Inverted when there is an alert. Touch to display the
Warning Screen.
MENU
To the Menu screen
In Manual Irradiation mode, the light modulation ratio can be set
individually for each LED. The initial setting is 100%.
To set the light modulation ratio, from the Manual Mode screen, press
the 100% for the number of the LED you wish to set (the light
modulation ratio can be set between 0 and 120%, in increments of 1%).
Performing Setup
Enter a number by touching 0 to 9, then touch the Enter button ( ↵).
The actions of the other keys are as follows:
ESC
Cancel the configuration, and return to the Manual screen.
BS
Back space; delete 1 character.
CLR
Clear the settings.