This table shows which input range is used for each input type.
94
• The Calibration Positive Range box appears with instructions to connect the top limit of
the input range. Once this is done press the Calibrate button. Next, the Calibration Neg-
ative Range box appears instructing the user to apply the bottom range limit. When this
is done press the Calibrate button.
• When the calibration is complete the icon changes to the User cal icon under the chan-
nel to show that the channel is using the user cal.
Calibration Input Ranges
Table 4.6 : Calibration Input range table
Input Type
C (W5) Type TC
G (W) Type TC
Power up
Table 4.5 : Analogue In Calibration
Calibration Type
Factory
User
Recalibrate
Recalibrate All
Factory All
K Type TC
R Type TC
S Type TC
B Type TC
J Type TC
T Type TC
E Type TC
N Type TC
Input Range
50mV
25mV
25mV
10mV
50mV
25mV
100mV
50mV
50mV
50mV
43-TV-25-35 GLO Iss.4 Dec 06 UK
A5E01001767-04