Application planning
4.2 Operation with at least one OXYMAT 7
4.2
Operation with at least one OXYMAT 7
Overview
If the sample gas is pumped into the analyzer module, surges may occur through operation
which can result in an unstable display of measured values. Fitting a sample gas restrictor in
the sample gas inlet of an OXYMAT 7 remedies this situation.
A further measure is the connection of a "pneumatic low pass" consisting of a restrictor and
a damping vessel.
①
Sample gas restrictor for OXYMAT 7, version with
hoses
②
Wall-mounted device, bottom
Install a sample gas restrictor:
● Before the sample gas inlet, if sample gas is pumped into the analyzer module
● Before the sample gas outlet, if sample gas is pumped out
When using two analyzer modules:
● Two analyzer modules in series:
● Two analyzer modules in parallel: Sample gas restrictor before each sample gas
24
– Sample gas restrictor upstream of the sample gas inlet of the analyzer module into
which the sample gas is first introduced.
– Sample gas restrictor at the sample gas outlet of the analyzer module from which the
sample gas is first pumped.
inlet/outlet of an OXYMAT 7.
③
Sealing cap
④
Sample gas restrictor for OXYMAT 7, version with
pipes
Operating Instructions, 01/2013, A5E31930403-01
Wall-mounted device