Installing the analyzer modules and option module
5.4 Cabling
Connecting option module OM 2.1 to the analyzer module
1
2
①
②
Figure 5-7
1. Connect the option module OM 2.1 to the analyzer module PCB of OXYMAT 7 using the
2. Close door and tighten the six Torx screws. Tightening torque: 2.0 Nm.
5.4
Cabling
Each analyzer module has a cable set. The CAN cable which connects all components
always has all connections.
Use the plastic cable bracket in the wall-mounted housing to ensure that the cables are laid
securely.
Cable damage
Make sure that the cables are carefully laid and secured again after you change an
analyzer module or other component. The cables may otherwise be damaged as a result of
friction or other mechanical effects.
36
Analyzer module printed circuit board
Analog current interface
Connecting option module OM 2.1 to analyzer modules 1 and 2
existing 26-pin ribbon cable.
CAUTION
Operating Instructions, 01/2013, A5E31930403-01
Wall-mounted device