Philips DreamStation 2 User Manual page 42

Hide thumbs Also See for DreamStation 2:
Table of Contents

Advertisement

Message
Message
type
SD Card
Alert
There was a problem reading
your SD card. Please remove
and reinsert.
SD Card
Alert
Your SD card is full. Please
refer to manual or contact your
provider.
Service Required
Safe State
Please contact support.
(therapy button ring blinks
continuously}
Notification
Notification
The attached power supply
is operating from a battery
and does not support
humidification.
Indicates an
Check Power
incompatible power supply is
(light blinks
attached.
continuously)
Low voltage
Low Voltage
(light blinks
continuously)
SD Card
Notification
Your SD card has been
removed. Please reinsert to
record therapy data.
- 38 -
Possible cause/additional action
Device cannot read the SD card. A problem may
exist with the SD card, it was removed during a
writing activity, or it was inserted incorrectly.
Remove and reinsert the SD card.
If alert continues, contact your provider.
Remove SD card and replace with a new card, or
contact your provider for a new SD card.
Indicates an error which enters device into "Safe
State." This allows power to remain on but airflow is
disabled.
Disconnect device from power. Reattach power cord
to restore power. If the alert continues to occur,
contact your home care provider.
Disconnect the battery pack and use a compatible
power supply or use the device without
humidification.
Switch to a Philips-provided power supply that
is capable of supporting therapy. An 80W power
supply is required to support humidification.
Confirm a compatible Philips Respironics power
supply is attached. Switch to compatible power
supply if needed. If battery is being used, ensure
battery is adequately charged.
Indicates SD card has been removed from therapy
device and not reinserted before the start of the
current therapy session.
Reinsert the SD card to record therapy data.