Ac Input Box Replacement Procedure; Tools; Preparation; Ac Input Box Removal Procedure - GE Vivid 4 BT03 RFI Service Manual

Table of Contents

Advertisement

GE M
S
EDICAL
YSTEMS
D
2392751-100, R
IRECTION
8-6-2

AC Input Box Replacement Procedure

8-6-2-1

Tools

Use the appropriate flat and Phillips-type screw drivers and a wire cutter as indicated in the AC Input
Box replacement procedure.
8-6-2-2

Preparation

Shut down the Vivid™ 4 ultrasound unit, as described in
8-6-2-3

AC Input Box Removal Procedure

1) Unplug the AC line power cable.
2) Remove the Vivid™ 4 left, right and rear covers, as described in the
Procedure
3) Unplug the internal power cable from the AC Distribution Box.
4) Unscrew the Ground cable from the AC Distribution Box.
5) Remove the two screws on both the left and the right sides that secure the AC Input Box to the
chassis.
6) Remove the AC Input Box.
8-6-2-4

AC Input Box Installation Procedure

1) Return the AC Input Box to its original position.
2) Tighten the two screws on both the left and the right sides that secure the AC Input Box to the
chassis.
3) Tighten the screw that connects the Ground cable to the AC Distribution Box.
4) Reconnect the internal power cable to the AC Distribution Box.
5) Install the Vivid™ 4 rear, left and right covers, as described in the
on page 8 - 7 and the
8-86
3
EVISION
on page 8 - 4, and the
Figure 8-69 AC Input Box Screws
Side Covers Installation Procedure
Section 8-6 - Lower Section Components Replacement
Chapter 3 -
Rear Cover Removal Procedure
V
™ 4 S
IVID
Installation.
Side Covers Removal
on page 8 - 7.
Rear Cover Installation Procedure
on page 8 - 4.
M
ERVICE
ANUAL

Advertisement

Table of Contents
loading

Table of Contents