GE Vivid 4 BT03 RFI Service Manual page 305

Table of Contents

Advertisement

GE M
S
EDICAL
YSTEMS
D
2392751-100, R
IRECTION
The Report dialog box contains the following information:
4) Check the test result message and proceed as follows:
5) Trackball to the Low Freq Test button and press Select. The normal channels diagnostic test is
performed in low frequency mode, and the test results are displayed in the Report dialog box, as
shown in
7-52
3
EVISION
Test Result Message: Displays the test result.
TR4 Modules: Displays a graphic representation of the front board (four groups of TR4
sockets). Each group consists of four TR4 sockets, each socket containing four vectors
(channels), as follows:
Each vector is represented by a button (highlighted in green) that displays the number of
the socket to which the TR4 is connected.
Each button is highlighted yellow when selected.
Each button is highlighted red if errors were detected for one or more channels.
Checkboxes: Enable you to manipulate bad channels:
Show Only Bad Channels: Select this checkbox to display only bad channels in the
Display area.
Mark bad channel: Select this checkbox to manually mark the currently selected channel
as bad.
If one or more channels were identified as bad, trackball to the Save Report button and press
Select, then contact a GE Medical Systems Service Representative.
If no channels were identified as bad, mark the channels suspected as being bad as follows:
Trackball to each of the channels in the Display area and press Select.
Trackball to the Mark bad channel checkbox and press Select.
Trackball to Save Report and press Select, and contact a GE Medical Systems service
representative.
Figure 7-29
on page 7-53:
Section 7-3 - Performing Front End (FE) Diagnostics
V
™ 4 S
M
IVID
ERVICE
ANUAL