Technical
Publication
Part Number 2392751-100
Revision 3
GE Medical Systems
Vivid™ 4 Service Manual
©
Copyright
2006 by GE Medical Systems

Advertisement

Table of Contents
loading

Summary of Contents for GE Vivid 4 BT03 RFI

  • Page 1 Technical Publication Part Number 2392751-100 Revision 3 GE Medical Systems Vivid™ 4 Service Manual © Copyright 2006 by GE Medical Systems...
  • Page 2 • VERSUCHEN SIE NICHT, DAS GERÄT ZU REPARIEREN, BEVOR DIESES WARNUNG KUNDENDIENST-HANDBUCH NICHT ZU RATE GEZOGEN UND VERSTANDEN WURDE. • WIRD DIESE WARNUNG NICHT BEACHTET, SO KANN ES ZU VERLETZUNGEN DES KUNDENDIENSTTECHNIKERS, DES BEDIENERS ODER DES PATIENTEN DURCH ELEKTRISCHE SCHLÄGE, MECHANISCHE ODER SONSTIGE GEFAHREN KOMMEN.
  • Page 3 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL • ESTE MANUAL DE SERVICIO SÓLO EXISTE EN INGLÉS. • SI ALGÚN PROVEEDOR DE SERVICIOS AJENO A GEMS SOLICITA UN IDIOMA QUE NO SEA EL INGLÉS, ES RESPONSABILIDAD DEL CLIENTE OFRECER UN SERVICIO DE TRADUCCIÓN.
  • Page 4 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL...
  • Page 5 GE Medical Systems personnel. In performing all electrical work on these products, GE will use its own specially trained field engineers. All of GE’s electrical work on these products will comply with the requirements of the applicable electrical codes.
  • Page 6 The contents of this publication may not be copied or duplicated in any form, in whole or in part, without prior written permission of GE Medical Systems. GE Medical Systems may revise this publication from time to time without written notice. TRADEMARKS All products and their name brands are trademarks of their respective holders.
  • Page 7 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Revision History Revision Date Reason for change October 2003 Initial Release February 2004 Hardware Modifications; Corrections July 2005 Updated System Labels; added Waste Electrical and Electronic Equipment (WEEE) Disposal warning...
  • Page 8: Table Of Contents

    GE M 2392751-100, R ™ 4 S EDICAL YSTEMS IRECTION EVISION IVID ERVICE ANUAL Table of Contents CHAPTER 1 Introduction Overview ............1 - 1 Purpose of Chapter 1 .
  • Page 9 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Electrostatic Discharge (ESD) Prevention ......1 - 24 Standards Used .
  • Page 10 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL CHAPTER 2 Pre-Installation Overview............2 - 1 Purpose of Chapter 2 .
  • Page 11 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL CHAPTER 3 Installation Overview ............3 - 1 Purpose of Chapter 3 .
  • Page 12 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Connecting the Color Printer ......3 - 27 Connecting the HP DeskJet 6122 Color Printer .
  • Page 13 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Disconnecting the Unit when Storing ......3 - 70 Preparing the Unit for Transport .
  • Page 14 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL CHAPTER 4 Functional Checks Overview............4 - 1 Purpose of Chapter 4 .
  • Page 15 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 7S Probe Image Quality Tests ........4 - 18 2D Center Noise Test .
  • Page 16 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL CHAPTER 5 Components and Function (Theory) Overview............5 - 1 Purpose of Chapter 5 .
  • Page 17 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Control ..........5 - 27 Image Port Board (IMP) .
  • Page 18 Contact GE ........
  • Page 19 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL CHAPTER 6 Service Adjustments Overview ............6 - 1 Purpose of Chapter 6 .
  • Page 20 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL CHAPTER 7 Diagnostics/Troubleshooting Overview............7 - 1 Purpose of Chapter .
  • Page 21 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Show History ........7 - 66 H/W Report .
  • Page 22 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL PC Diagnostics ........7 - 103 CPU Tests .
  • Page 23 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL CHAPTER 8 Replacement Procedures Overview ............8 - 1 Purpose of Chapter 8 .
  • Page 24 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Control Console Upper Cover (Front) Removal Procedure ..8 - 14 Control Console Upper Cover (Front) Installation Procedure ..8 - 15 Control Console Upper Cover (Rear) Replacement Procedure .
  • Page 25 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Preparation ..........8 - 28 Keypad Removal Procedure .
  • Page 26 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Tools ..........8 - 41 Preparation .
  • Page 27 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL PC2IP Board Replacement Procedure ......8 - 63 Tools .
  • Page 28 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Lower Section Components Replacement ....... 8 - 84 AC Distribution Box Replacement Procedure .
  • Page 29 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Rear Wheel Replacement Procedure ......8 - 111 Tools .
  • Page 30 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL HP 6540/3 USB Deskjet Color Printer Installation Procedure ..8 - 140 HP 6540/3 USB Deskjet Color Printer Removal Procedure ..8 - 150...
  • Page 31 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL CHAPTER 9 Renewal Parts Overview ............9 - 1 Purpose of Chapter 9 .
  • Page 32 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL CHAPTER 10 Periodic Maintenance Overview............10 - 1 Purpose of Chapter 10 .
  • Page 33 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Electrical Safety Tests ..........10 - 18 Safety Test Overview .
  • Page 34: Introduction

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Chapter 1 Introduction Section 1-1 Overview 1-1-1 Purpose of Chapter 1 This chapter describes important issues related to safely servicing the Vivid™ 4 scanner. The service provider must read and understand all the information presented here before installing or servicing a unit.
  • Page 35: Typical Users Of The Basic Service Manual

    Provides periodic maintenance procedures for the Vivid™ 4 ultrasound unit. 1-1-3 Typical Users of the Basic Service Manual This manual is intended for the following categories of users: • GE service personnel (installation, maintenance, etc.). • Hospital service personnel. • Contractors (some parts of Chapter 2 - Pre-Installation).
  • Page 36: Vivid™ 4 Models Covered In This Manual

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 1-1-4 Vivid™ 4 Models Covered in this Manual The Vivid™ 4 models documented in this manual are shown in Table 1-2 Table 1-3 below. NOTE: The difference between the two types of Vivid™ 4 BT03 models are as follows: On RFI models (supported by software version 4.2, and above), Image Port, RFT, and FEC functionality...
  • Page 37: System History - Hardware And Software Versions

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 1-1-5 System History - Hardware and Software Versions The newest generation of the Vivid™ 4 (BT03) ultrasound unit is based on the Vivid 3 (BT01 and BT02) ultrasound unit, and therefore shares much of the same hardware.
  • Page 38: Important Conventions

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Section 1-2 Important Conventions 1-2-1 Conventions Used in this Manual 1-2-1-1 Model Designations This manual covers the Vivid™ 4 ultrasound units listed in Table 1-2 on page 1-3...
  • Page 39: Standard Hazard Icons

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 1-2-1-4 Standard Hazard Icons Important information will always be preceded by the exclamation point contained within a triangle, as seen throughout this chapter. In addition to text, several different graphical icons (symbols) may be used to make you aware of specific types of hazards that could cause harm.
  • Page 40: Safety Considerations

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Section 1-3 Safety Considerations 1-3-1 Introduction The following safety precautions must be observed during all phases of operation, service and repair of this equipment. Failure to comply with these precautions or with specific warnings elsewhere in this manual, violates safety standards of design, manufacture and intended use of the equipment.
  • Page 41: Electrical Safety

    Note: When replacing UPS batteries, ensure that defective batteries are disposed of in accordance with local regulations. Alternatively, forward them to GE Medical Systems for proper disposal. 1-3-4-1 Probes All the probes for the Vivid™...
  • Page 42: Dangerous Procedure Warnings

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 1-3-5 Dangerous Procedure Warnings Warnings, such as the examples below, precede potentially dangerous procedures throughout this manual. Instructions contained in the warnings must be followed. DANGEROUS VOLTAGES, CAPABLE OF CAUSING DEATH, ARE PRESENT IN THIS EQUIPMENT.
  • Page 43: Product Labels And Icons

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Section 1-4 Product Labels and Icons The Vivid™ 4 ultrasound unit comes equipped with product labels and icons. These labels and icons represent pertinent information regarding the operation of the ultrasound unit.
  • Page 44 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Figure 1-2 Product Label and Icon Locations (Rear) 1 Main Label 2 AC Voltage Rating Label 3 GND Label Chapter 1 - Introduction 1-11...
  • Page 45: Label Descriptions

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 1-4-2 Label Descriptions The following table shows the labels and symbols that may be found on the Vivid™ 4 ultrasound unit, and provides a description of each label’s purpose and location.
  • Page 46 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Table 1-6 Product Icons (Continued) Label Name Description Location "DANGER - Risk of explosion The system is not designed for use with Indicated in the Service Manual.
  • Page 47: Vivid™ 4 External Labels

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 1-4-3 Vivid™ 4 External Labels In addition to the labels described in the previous section, additional labels may be found on the Vivid™ 4 ultrasound unit, as described in the following sections: •...
  • Page 48 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL • German: Used in all German language countries. Figure 1-5 Main Label (German) • French: Used in all French language countries. Figure 1-6 Main Label (French)
  • Page 49 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL • Spanish: Used in all Spanish language countries. Figure 1-7 Main Label (Spanish) • Portuguese: Used in all Portuguese language countries. Figure 1-8 Main Label (Portuguese)
  • Page 50 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Figure 1-9 Main Label (Portuguese) 220 - 240V • Italian: Used in all Italian language countries. Figure 1-10 Main Label (Italian) Chapter 1 - Introduction 1-17...
  • Page 51 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL • Chinese: Used in all Chinese language countries. Figure 1-11 Main Label (Chinese) • Danish: Used in all Danish language countries. Figure 1-12 Main Label (Danish)
  • Page 52 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL • Dutch: Used in all Dutch language countries. Figure 1-13 Main Label (Dutch) • Finnish: Used in all Finnish language countries. Figure 1-14 Main Label (Finnish)
  • Page 53 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL • Greek: Used in all Greek language countries. Figure 1-15 Main Label (Greek) • Japanese: Used in all Japanese language countries. Figure 1-16 Main Label (Japanese)
  • Page 54 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL • Norwegian: Used in all Norwegian language countries. Figure 1-17 Main Label (Norwegian) • Russian: Used in all Russian language countries. Figure 1-18 Main Label (Russian)
  • Page 55: Rating Labels

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL • Swedish: Used in all Swedish language countries. Figure 1-19 Main Label (Swedish) 1-4-3-2 Rating Labels Indicates the ultrasound unit’s factory preset input AC voltage as follows: •...
  • Page 56: Gnd Label

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 1-4-3-3 GND Label Indicates the protective earth (grounding) terminal. The GND label shown below is located at the rear of the unit, as shown in Figure 1-2 on page 1-11.
  • Page 57: Emc, Emi, And Esd

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Section 1-5 EMC, EMI, and ESD 1-5-1 Electromagnetic Compatibility (EMC) Electromagnetic compatibility describes a level of performance of a device within its electromagnetic environment. This environment consists of the device itself and its surroundings, including other equipment, power sources and persons with which the device must interface.
  • Page 58: Standards Used

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 1-5-3 Standards Used To fulfill the requirements of relevant EC directives and/or European Harmonized/International standards, the following documents/standards have been used: Table 1-7 Standards Used Standard/Directive...
  • Page 59: Customer Assistance

    Milwaukee, WI 53219 Customer Answer Center Phone: +1-262-524-5698 Fax: +1-414-647-4125 Europe General Imaging: +49 (212) 2802 207 GE Ultraschall Deutschland GmbH& Co. KG Cardiac: +49 (212) 2802 208 BeethovenstraBe 239 Postfach 11 05 60, D-42665 Solingen Germany Fax: +49 212 2802 431...
  • Page 60: Overview

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Chapter 2 Pre-Installation Section 2-1 Overview 2-1-1 Purpose of Chapter 2 This chapter provides the information required to plan and prepare for the installation of a Vivid™ 4 ultrasound unit.
  • Page 61: Console Requirements

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Section 2-2 Console Requirements 2-2-1 Unit Environmental Requirements Table 2-2 Environmental Requirements Relative Humidity Requirement Temperature (non-condensing) Air Pressure Operational 50 — 70% 700 — 1060 hPa 10 —...
  • Page 62: Electrical Requirements

    2-2-5 Electrical Requirements NOTE: GE Medical Systems requires a dedicated power and Ground for the proper operation of its Ultrasound equipment. This dedicated power shall originate at the last distribution panel before the system. Sites with a mains power system with defined Neutral and Live: The dedicated line shall consist of one phase, a neutral (not shared with any other circuit), and a full size Ground wire from the distribution panel to the Ultrasound outlet.
  • Page 63: Site Power Outlets

    2-2-5-5 Mains Power Plug If the unit arrives without a power plug, or with the wrong plug, contact your GE dealer. When necessary, the installation engineer will supply the locally-required power plug. 2-2-5-6...
  • Page 64: Emi Limitations

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 2-2-6 EMI Limitations Ultrasound machines are susceptible to Electromagnetic Interference (EMI) from radio frequencies, magnetic fields, and transients in the air or wiring. They also generate EMI. The Vivid™ 4 ultrasound unit complies with limits as stated on the EMC label.
  • Page 65: Probe Environmental Requirements

    RF gaskets touch metal. to a different, appropriate location. The interconnect cables are grounded and require ferrite beads and other shielding. Use GE specified harnesses and Cable length, material, and routing are all important; do not make any changes that do peripherals.
  • Page 66: Facility Needs

    (and the accompanying electrical installations) are highly sophisticated and special engineering competence is required. All electrical work on these products must comply with the requirements of applicable electrical codes. The purchaser of GE equipment must utilize only qualified personnel to perform electrical servicing of the equipment.
  • Page 67: Mandatory Site Requirements

    2-11 for network details that are required. Note: All relevant preliminary network outlets installations at the prepared site must be performed by authorized contractors. The purchaser of GE equipment must utilize only qualified personnel to perform servicing of the equipment. 2-3-3 Site Recommendations The following are (optional) site recommendations.
  • Page 68: Recommended Ultrasound Room Layout

    Recommended Ultrasound Room Layout Dedicated Power Outlets VIVID 4 Hospital Network Dedicated Analog Telephone Line for Connection to InSite GE Cabinet for Software and Manuals (optional) 36 IN. (92 CM) Scale: Each square = 1 sq ft (144 sq ins) Figure 2-1 Minimal Floor Plan 2.5m x 3m (8.2ft x 9.84 ft)
  • Page 69: Dicom Option Pre-Installation Requirements

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 2-3-4-4 DICOM Option Pre-Installation Requirements To configure the Vivid™ 4 ultrasound unit to work with other network connections, the network administrator must provide the required information, which should include the following: •...
  • Page 70: Connectivity Installation Worksheet

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Section 2-4 Connectivity Installation Worksheet Site System Information Comments: Floor: Site: Dept: Room: Type: Vivid™ 4 SN: REV: CONTACT INFORMATION Name Title E-Mail Address Phone Remote Archive Setup...
  • Page 71 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Table 2-6 Pre-Installation Check List Action Schedule at least 3 hours for installation of the system. Notify installation team of the existence of any variances from the basic installation.
  • Page 72: Overview

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Chapter 3 Installation Section 3-1 Overview 3-1-1 Purpose of Chapter 3 This chapter provides instructions for installing the Vivid™ 4 ultrasound unit. Before beginning the installation process, an appropriate site must be prepared, as described in Chapter 2 - Pre-Installation.
  • Page 73: Installation Reminders

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Section 3-2 Installation Reminders 3-2-1 Average Installation Time Once the site has been prepared, the average installation time required is shown in Table 3-2 below. Table 3-2...
  • Page 74: Safety Reminders

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 3-2-3 Safety Reminders WHEN USING ANY TEST INSTRUMENT THAT IS CAPABLE OF DANGER: OPENING THE AC GROUND LINE (I.E., METER’S GROUND SWITCH IS OPEN), DO NOT TOUCH THE UNIT! WARNING: Two people are required to unpack the unit, as it is heavy.
  • Page 75: Receiving And Unpacking The Equipment

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Section 3-3 Receiving and Unpacking the Equipment CAUTION: Please read this section fully before unpacking the Vivid™ 4 ultrasound unit. Note: The Vivid™ 4 ultrasound unit is shipped from the factory either in a wooden shipping crate, or in a cardboard shipping carton.
  • Page 76: Unpacking And Removing The Unit From The Wooden Crate

    If the crate is damaged, or if either the Drop or Tilt indicators have turned red (showing failure), please inform the GE Medical Systems sales representative immediately. In addition, mark on the shipping consignment note or packing slip/post-delivery checklist (in the “Package” column) that the Tilt and/or Drop indicators show failure.
  • Page 77 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL CAUTION: When using sharp tools to open packing materials, take care to avoid cutting or damaging the contents. Note: Unless otherwise specified, referenced items are shown in Figure 3-1 on page 3-4.
  • Page 78 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 12) Remove the two monitor foam supports (item 12 in Figure 3-4), located one on each side of the monitor. Figure 3-4 Antistatic Cover, Supports and Accessories...
  • Page 79 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 17) Press the brake pedal, located near the bottom of the front of the unit, to the right, to prevent the wheels from swiveling, as described in the Swivel Brake Label section, page 1-23.
  • Page 80: Unpacking The Cardboard Shipping Carton

    If the crate is damaged, or if either the Drop or Tilt indicators have turned red (showing failure), please inform the GE Medical Systems sales representative immediately. In addition, mark on the shipping consignment note or packing slip/post-delivery checklist (in the “Package” column) that the Tilt and/or Drop indicators show failure.
  • Page 81 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL CAUTION: When using sharp tools to open packing materials, take care to avoid cutting or damaging the contents. Note: Unless otherwise specified, referenced items are shown in Figure 3-6 on page 3-9.
  • Page 82 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Figure 3-8 Front and Side View of the Unit in the Shipping Carton 8) Pull out and remove the monitor support (item 10 in Figure 3-9 below) from under the base of the monitor.
  • Page 83 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 9) Release the front stopper (item C in Figure 3-10 below) by sliding the metal sleeve located at the locking pin’s base upwards, and then turning it towards you.
  • Page 84: Preparing For Installation

    When preparing for installation of a Vivid™ 4 system, it is important to verify that all items ordered by the customer have been received. Compare all items listed on the packing slip (shipping consignment note) with those received and report any items that are missing, back-ordered, or damaged, to your GE Medical Systems sales representative.
  • Page 85: Accessories

    3-4-3-1 Damage Inspection Checklist Visually inspect the contents of the shipping crate/shipping carton for damage. If any parts are damaged or missing, contact an authorized GE Service Representative. A Damage Inspection Checklist is provided in Table 3-6 on page 3-15.
  • Page 86 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Table 3-6 Damage Inspection Checklist Step Item Recommended Procedure Verify that the system is switched OFF and unplugged. Clean the console and Console control panel. Probe Holders Clean the gel wells with warm water and a damp cloth to remove all traces of gel.
  • Page 87: Front And Side View Of The Vivid™ 4 Ultrasound Unit

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 3-4-3-2 Front and Side View of the Vivid™ 4 Ultrasound Unit Figure 3-11 below shows the Vivid™ 4 ultrasound unit components that are visible from the front and side of the ultrasound unit.
  • Page 88 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Foot Brake: Three-position brake, as follows: • LEFT (P) locks the wheels. • MIDDLE unlocks the wheels. • RIGHT (double arrows) locks the swivel action. 10 (Not shown in...
  • Page 89: Rear View Of The Vivid™ 4 Ultrasound Unit

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 3-4-3-3 Rear View of the Vivid™ 4 Ultrasound Unit The following figure shows the Vivid™ 4 ultrasound unit components that are visible from the rear of the ultrasound unit: Figure 3-12 Rear View of the Vivid™...
  • Page 90: System Voltage Confirmation

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 3-4-4 System Voltage Confirmation 3-4-4-1 System Voltage Settings Verify that the scanner is set to the correct voltage. The Voltage settings for the Vivid™ 4 Scanner are found on a label to the right of the Power switch and External I/O, on the rear of the system.
  • Page 91 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Figure 3-14 System Configuration Technical Support Tab 5) Click OK. Note: If the voltage is not set correctly, contact an authorized GE Service Representative. 3-20 Section 3-4 - Preparing for Installation...
  • Page 92: Video Formats Confirmation

    4) Click the Technical Support tab and make sure Frequency is set to either 60Hz for NTSC, or 50Hz for PAL. 5) Click OK. Note: If the video format is not set correctly, contact an authorized GE Service Representative. Chapter 3 - Installation 3-21...
  • Page 93: Ensuring Protection From Emi

    RF gaskets touch metal. appropriate location. The interconnect cables are grounded and require ferrite beads and other shielding. Cable Use GE specified harnesses length, material, and routing are all important; do not make any changes that do not meet and peripherals.
  • Page 94: Completing The Hardware Installation

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Section 3-5 Completing the Hardware Installation It is recommended to pay attention to the system specifications and make sure the facility has been prepared in accordance with the information provided in Chapter 2 - Pre-Installation.
  • Page 95: Connecting Peripherals

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 3-5-2 Connecting Peripherals Peripheral devices, such as a VCR or printer, are connected to the Vivid™ 4 ultrasound unit using the rear panel connectors. Ensure that all peripheral devices connected to the ultrasound unit comply with national safety requirements for medical equipment, including IEC601, CSA22.2, AS3200.1 and UL544.
  • Page 96: Right Rear Panel Connectors

    For use with external peripherals. Voltages are set according to local country voltage. If additional auxiliary outlets are required, use the special cable 2. AUXILIARY AC OUTLET provided by GE. DO NOT attempt to connect additional peripherals using an external wall outlet. 3. THERMAL CIRCUIT BREAKERS Three 4A thermal circuit breakers for fuse protection.
  • Page 97: Left Rear Panel Connectors

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 3-5-2-1-2 Left Rear Panel Connectors Table 3-10 describes the connectors included in the left rear panel (shown in Figure 3-17): Figure 3-17 Left Rear Panel Connectors...
  • Page 98: Connecting The Vcr

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 3-5-2-2 Connecting the VCR 1) Place the VCR on the peripheral tray and connect the following: Table 3-11 VCR Cables From the VCR To the Left Panel...
  • Page 99: Connecting The Hp Deskjet 6122 Color Printer

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Table 3-13 Color Printer Cables To the Cables From the Color Printer underneath the Control Console Video IN R, G, B, Sync External Trigger Print Trigger...
  • Page 100: Connecting Probes

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 3-5-3 Connecting Probes The Vivid™ 4 ultrasound unit operates with various types of probes that are used for scanning patients, including flat phased, convex and linear electronic array probes. Once connected, the probes can be selected for different applications.
  • Page 101: Connecting The Ecg

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 3-5-4 Connecting the ECG The internal ECG is connected into a rectangular-shaped socket on the patient trace (I/O) panel. The patient trace (I/O) panel is located on the front of the ultrasound unit, as shown in Figure 3-11 on page 3-16.
  • Page 102 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL AHA (Americas, Japan) IEC (Europe, Asia, ROW) Figure 3-19 ECG Cable and Electrode Placement Note: For optimal ECG operation, use only electrodes that meet universal standards - see Table 3-16.
  • Page 103: Connecting The Unit To A Power Source

    GE Medical Systems. Use only the power cords, cables and plugs provided by or designated by GE Medical Systems to connect the unit to the power source. CAUTION: Verify compliance with all electrical and safety requirements and check the power cord to verify that it is intact and of hospital-grade before connecting the unit to the power source.
  • Page 104: Connecting The Ultrasound Unit To The Electrical Outlet

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 3-5-5-2 Connecting the Ultrasound Unit to the Electrical Outlet Note: To help assure grounding reliability, connect to a hospital-grade or “hospital only” grounded power outlet. If using the ultrasound unit with an external UPS system, follow all the grounding and applicable safety standards as documented both in this manual and the external UPS manufacturer’s manual.
  • Page 105: Disconnecting The Ultrasound Unit From The Electrical Outlet

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 3-5-5-3 Disconnecting the Ultrasound Unit from the Electrical Outlet CAUTION: Whenever disconnecting the Vivid™ 4 unit from the electrical outlet, always observe the safety precautions. First unplug the mains power cable from the wall outlet socket, then from the unit itself.
  • Page 106 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Figure 3-21 Shut-Down Options Screen Chapter 3 - Installation 3-35...
  • Page 107: System Configuration

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Section 3-6 System Configuration Once all the required peripherals have been installed and the unit has been switched on, configure the system settings in the System Configuration window tabs. Refer to the Vivid™ 4 User Manual for additional information about system configuration.
  • Page 108: System Tab

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 3-6-3 System Tab 1) Press Config on the alphanumeric keyboard. The System Configuration window is displayed. 2) Trackball to the System tab and press Select. The software information is displayed in the upper...
  • Page 109: Connectivity Tab

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 3-6-4 Connectivity Tab For details on the Connectivity tab, refer to the Connectivity Setup section, on page 3-52. 3-6-5 Archive Tab 1) Press Config on the alphanumeric keyboard. The System Configuration window is displayed.
  • Page 110 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 7) Select the patient ID page type from the Patient ID Page Type dropdown list, as follows: • Type A: displays patient ID, last and first name, weight, height, BSA, BP and other fields.
  • Page 111: Annotation Settings Tab

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 3-6-6 Annotation Settings Tab 1) Press Config on the alphanumeric keyboard. The System Configuration window is displayed. 2) Trackball to the Annotation Settings tab and press Select. The Annotation Settings tab is...
  • Page 112 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Whenever the Bodymark is now displayed, it will appear in the newly-selected position (until such time as this setting is changed) 5.) Activate the Duplicate Bodymark on Switch Window and/or the Bodymark Menu on Freeze checkboxes, as required.
  • Page 113: System Options Tab

    Note: Please note that all passwords should be visible underneath the control console or/and on the green password (option) certificate. If a new password (option) is required, contact the GE representative to initiate the appropriate process, so that the relevant password (option) will be installed on the system.
  • Page 114 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Note: If an option was not initially purchased or installed, it may be acquired and/or installed later. During option installation, type the relevant password in the appropriate Option window (see...
  • Page 115: Printers Tab

    Note: Please note, the printer indicated in the Currently Installed Printers field, is the recommended and approved printer to be used with the Vivid™ 4 system. Other generic printers are not approved, and GE Medical Systems will not guarantee their correct operation on this system.
  • Page 116: Vcr/Ecg Tab

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 3-6-9 VCR/ECG Tab 1) Press Config on the alphanumeric keyboard. The System Configuration window is displayed. 2) Trackball to the VCR/ECG tab and press Select. The VCR/ECG tab is displayed, as shown below:...
  • Page 117 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 5) In the ECG and Phono Options area, complete the following, as required: • Select each connected option including, ECG Exist, ECG External, Phono Exist and/or Phono Filter.
  • Page 118: Technical Support Tab

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 3-6-10 Technical Support Tab The System Configuration Technical Support tab is used when new hardware has been installed, or when it necessary to enable new functionality. The appearance of the Technical Support tab differs in accordance with the specific Vivid™...
  • Page 119 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Table 3-17 Recommended Vivid™ 4 Hardware Configurations - RFI Models Item Vivid™ 4 - BT03 Second Re Image Port No board Front Board V7-3 (TR-V4) Ex Keyboard...
  • Page 120: Technical Support Tab - Rft Configuration

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 3-6-10-2 Technical Support Tab - RFT Configuration Figure 3-30 Technical Support Tab - RFT Configuration Monitoring - should always be enabled. (this provides constant monitoring of critical parameters to ensure safe operation of the system).
  • Page 121 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Table 3-18 Recommended Vivid™ 4 Hardware Configurations - RFT Models Item Vivid™ 4 - BT03 Second Re Image Port Revision E8/F8 Front Board V7-3 (TR-V4) Ex Keyboard...
  • Page 122: Technical Support History Tab

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 3-6-11 Technical Support History Tab The Technical Support History is where you would record any historical technical support information that may be relevant to a problem, or any other maintenance operations performed.
  • Page 123: Connectivity Setup

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Section 3-7 Connectivity Setup 3-7-1 Introduction The Vivid™ 4 ultrasound unit can be connected to various connectivity devices, such as DICOM devices and EchoPac servers. The following sections describe how to connect the system to a remote archive/work station or a DICOM service, using a TCP/IP connection.
  • Page 124: Local Network Connection To Echopac Pc Workstation

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 3-7-2-2 Local Network Connection to EchoPAC PC Workstation Do not use the fifth connector and the switch can be in any position, Vivid7 Vivid3 EchoPAC PC Use only Standard non-crossed Ethernet cabling.
  • Page 125: Setting Up For Connectivity

    EchoPAC. This function (which is configured on EchoPAC, not on the Vivid™ 4 system) is used to export examination results to the Hospital Information System (HIS), through HL7 protocol. Users requiring this option should contact the GE representative. 1) Press Config on the alphanumeric keyboard. The System Configuration window is displayed.
  • Page 126 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 4) Enable the DICOM printer option, as required (under the Services sub-tab) - see Figure 3-36. Figure 3-36 Setting up DICOM Print 5) Enable any other remote server, as required (for example, Services, My Computer, select...
  • Page 127 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Figure 3-37 Setting up Remote Server for MPEGvue 6) When done, trackball to the OK button and press Select. 3-56 Section 3-7 - Connectivity Setup...
  • Page 128: Setting Up The Network Connection

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 3-7-4 Setting Up the Network Connection 1) Press Config on the alphanumeric keyboard. The System Configuration window is displayed. 2) Trackball to the Connectivity tab and press Select. The Connectivity tab is displayed, with the...
  • Page 129 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 5) Trackball to the Local Settings tab and press Select. The Local Settings tab is displayed, as shown below. Figure 3-40 Local Settings Tab Note: The Computer Name field shows the serial number of your Vivid™ 4 system.
  • Page 130: Setting Up For Communication With A Prosolv Workstation Or Echopac

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 3-7-5 Setting Up for Communication with a Prosolv Workstation or EchoPac The Vivid™ 4 ultrasound unit can connect to a Prosolv DICOM viewer through a DICOM server. The unit cannot connect directly to the viewer.
  • Page 131 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 8) Enter the IP address for the DICOM server in the IP-address field. 9) Trackball to the Dicom Storage Server line in the Services table. The Services tab is expanded,...
  • Page 132 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Figure 3-44 Connectivity Tab (Settings for DICOM Print Expanded) 12) Under Settings, select the required Format, Film Orientation, etc (see Figure 3-44, above). 13.)Proceed to the next section, to perform the procedure for validating communication with the DICOM server.
  • Page 133: Validating Communication With The Dicom Server

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 3-7-5-1 Validating Communication with the DICOM Server 1) Trackball to the Views tab and press Select. The Views tab is displayed. 2) Trackball to the Dataflow button and press Select. The full list of data flows is displayed, as shown...
  • Page 134 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 5) If the connection was successfully established, trackball to the Dicom Storage Server line and press Select, then trackball to the Check button and press Select.
  • Page 135: Validating Communication With The Workstation

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 3-7-5-2 Validating Communication with the Workstation 1) Press the Archive button on the control panel. The Patient List window is displayed, as shown below: Figure 3-46 Patient List Window 2) Select the Local Archive-Int.
  • Page 136: Connecting Directly To Echopac (Or Image Vault)

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 5) Press <Ctrl+S> to view the status of the spooler, as shown below: Figure 3-47 Job Spooler The Status column can contain the following options: •...
  • Page 137 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 7) Select the EchoPAC server from the Name dropdown list. The Services tab is updated, as shown below: Figure 3-48 Services Tab - EchoPAC 8) Enter the IP address of the EchoPAC server in the IP-address field.
  • Page 138 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 10) If EchoPAC Network is required, make sure that the EchoPAC Network checkbox is selected, as shown in Figure 3-49, above, then proceed as follows: a.) Enter ECHOPAC7-OOOXXX-DB in the Database Name field.
  • Page 139: Validating Communication With The Echopac Server

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 3-7-6-1 Validating Communication with the EchoPAC Server 1) Trackball to the Views tab and press Select. The Views tab is displayed. 2) Trackball to the Dataflow button and press Select. The full list of data flows is displayed, as shown below.
  • Page 140: Validating Echopac Or Image Vault Operation

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL NOTE: IMPORTANT INFORMATION ABOUT DATAFLOWS: The system’s default dataflow is Local Archive - Int. HD. In this configuration, the Patient List is in the local archive and the images are stored on the internal hard drive.
  • Page 141: Storing And Transporting The Unit

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Section 3-8 Storing and Transporting the Unit The Vivid™ 4 ultrasound unit weighs up to 170 kg (375 lbs), depending on the configuration. To avoid equipment damage and ensure maximum safety while transporting or moving the unit, the following precautions are recommended: •...
  • Page 142: Safety Precautions For Moving The Vivid™ 4 Unit

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 3-8-3 Safety Precautions for Moving the Vivid™ 4 Unit CAUTION: Whenever moving the Vivid™ 4 ultrasound unit, always observe the following safety precautions: 1) Proceed cautiously when crossing door or elevator thresholds. Grasp the front handle grips or the back handle bar, and then push or pull.
  • Page 143: Cardboard Shipping Carton And Packaging Materials

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 3-8-5 Cardboard Shipping Carton and Packaging Materials Verify that the following packing materials are available. All these materials are required to safely package and transport the Vivid™ 4 ultrasound unit in the cardboard shipping carton for shipment by vehicle, or any other means.
  • Page 144: Packing The Unit Into The Wooden Shipping Crate

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 3-8-6 Packing the Unit into the Wooden Shipping Crate Note: For details of item numbers in this procedure, refer to the following figures: • Figure 3-1 on page 3-4 •...
  • Page 145: Assembling The Wooden Shipping Crate

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 21) Lower the control console (monitor) so that it is pressing down on the two cartons. Important! NOTICE Verify that the height from the crate base to the top of the monitor does not exceed 132cm (52in) 22) Place the pre-packed manuals in the compartment under the monitor.
  • Page 146 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 10) Hold the front wall (item 13) vertically against the front of the crate and secure it in place using eight clips (item 15). 11) Secure the hand lever (clip remover) tool (item 23) to the outside of the wooden crate, using a screw.
  • Page 147: Packing The Unit In The Cardboard Shipping Carton

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 3-8-8 Packing the Unit in the Cardboard Shipping Carton Note: For details of item numbers in this procedure, refer to the following figures: • Figure 3-1 on page 3-4 •...
  • Page 148: Assembling The Cardboard Shipping Carton

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 23) Lower the control console so that it is pressing on the two cartons, and verify that the height from the base of the case to the top of the monitor does not exceed 132cm (52in).
  • Page 149: Completing The Installation Paperwork

    - see Figure 3-52 on page 3-79. Note: Step 2 may have been completed already by GE personnel. The Product Locator Installation Card shown in Figure 3-52 may not be identical to the card(s) provided.
  • Page 150: User Manual(S)

    Check that the correct Vivid™ 4 User Manual(s) for the system (and relevant software revision), is included with the installation. Specific language versions of the Vivid™ 4 User Manual may also be available. Check with your GE Sales Representative for availability. During and after installation, supplementary documentation (such as, User Manuals and Installation Manuals for the peripherals) must be kept as part of the original system documentation.
  • Page 151: Overview

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Chapter 4 Functional Checks Section 4-1 Overview 4-1-1 Purpose of Chapter 4 This chapter describes the functional checks that must be performed either as a part of the installation procedure, or as required during servicing and periodic maintenance of the Vivid™...
  • Page 152: General Procedures

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Section 4-2 General Procedures 4-2-1 Power ON/OFF and Boot-up Tests Table 4-2 Power ON/OFF and Boot-up Tests Step Task Expected Result(s) Power up the unit by holding down the On/Off button.
  • Page 153: Functional Check

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Section 4-3 Functional Check 4-3-1 Basic Controls 4-3-1-1 Keyboard and Footswitch Tests Table 4-3 Keyboard and Footswitch Tests Step Task Expected Result(s) Start checking the alphanumeric keyboard backlights.
  • Page 154: Speakers Tests

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 4-3-1-2 Speakers Tests Table 4-5 Speakers Test Step Task Expected Result(s) Run the audio diagnostic, as described in the Audio (Doppler Sound Driver) Diagnostic Test on page 7 - 71.
  • Page 155: Video Out Test

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 4-3-2-2 Video OUT Test Table 4-8 Video OUT Test Step Task Expected Result(s) Activate a CineLoop or Freeze a scanned image. Connect the monitor to the composite VIDEO OUT 1 The image is displayed correctly.
  • Page 156: Mechanical Functions

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 4-3-3 Mechanical Functions 4-3-3-1 Console Movement Test Table 4-11 Console Movement Test Step Task Expected Result(s) Pull the front handle (located under the control console) to Ensure that you do not apply too much force to move the half-pressed position and swivel the console to the left.
  • Page 157: Back End Processor Tests

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 4-3-4 Back End Processor Tests 4-3-4-1 Internal ECG Test Table 4-13 Internal ECG Test Step Task Expected Result(s) Connect the internal ECG input to an ECG simulator or suitable substitute, using the cable set and electrode pads.
  • Page 158: Magneto Optical Drive (Mod) Test

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 4-3-4-3 Magneto Optical Drive (MOD) Test Table 4-15 MOD Test Step Task Expected Result(s) Restart the system performing a full shutdown and During the boot-up process, verify that the messages...
  • Page 159: Cd Writer Test

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 4-3-4-4 CD Writer Test Table 4-16 CD Writer Test Step Task Expected Result(s) Insert the new recordable CD X16 in the CDR drive. Press Ctrl+F11 (or Alt+D) on the keyboard and select UtilMO_CDR_App is displayed.
  • Page 160: Plug And Scan Test

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 4-3-4-5 Plug and Scan Test Table 4-17 Plug and Scan Test Task Expected Result(s) Press the On/Off button on the control console for no The system enters Standby mode.
  • Page 161: Image Testing: 2D/M/Cfm/Doppler

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Section 4-4 Image Testing: 2D/M/CFM/Doppler NOTE: To perform these tests thoroughly, it is recommended to use three types of soft tissue phantoms: RMI 403GS, RMI 404G and RMI 1425A, or an equivalent surface that can simulate the same soft tissue effects.
  • Page 162: Penetration Test

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 4-4-1-3 2D Penetration Test Table 4-20 2D Noise Penetration Test (3S) Step Task Expected Result(s) Use the Standard Imaging Phantom RMI403GS. Select a cardiac preset. Set the following parameters, and scan the phantom at position (3): •...
  • Page 163: Cfm Stationary Noise Test

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 4-4-1-5 CFM Stationary Noise Test Table 4-22 CFM Stationary Noise Test (3S) Step Task Expected Result(s) Use the vertical nail area in the Electronic Flow Phantom RMI1425A.
  • Page 164: Cfm Operation Tests

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 4-4-1-7 CFM Operation Tests Table 4-24 CFM Operation Tests (3S) Step Task Expected Result(s) Select a Cardiac preset and set the 2D Gain to 40. Activate CFM and set the following CFM parameters: •...
  • Page 165: Color M Operation Tests

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 4-4-1-8 Color M Operation Tests Table 4-25 Color M Operation Tests (3S) Step Task Expected Result(s) Select a Cardiac preset and set the following 2D parameters: •...
  • Page 166: Pw Operation Test

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 4-4-1-9 PW Operation Test Table 4-26 PW Operation Tests (3S) Step Task Expected Result(s) Select a Cardiac preset and set the 2D Gain to 40. Activate PW Mode and set the following PW parameters: •...
  • Page 167: Cw Operation Tests

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 4-4-1-10 CW Operation Tests Table 4-27 CW Operation Tests (3S) Step Task Expected Result(s) Select a Cardiac preset and set the 2D Gain to 45. Activate CW Mode and set the following CW parameters: •...
  • Page 168: Cw Noise Test

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 4-4-1-11 CW Noise Test Table 4-28 CW Noise Test (3S) Step Task Expected Result(s) With a 3S probe in the air, select a Cardiac preset and activate CW-Mode.
  • Page 169: C358 Curved Probe Image Quality Tests

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 4-4-3 C358 Curved Probe Image Quality Tests 4-4-3-1 2D Artifacts Test Table 4-30 2D Artifacts Test (C358) Step Task Expected Result(s) Use the Standard Imaging Phantom RMI403GS.
  • Page 170: Pw Operation Tests

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 4-4-3-3 PW Operation Tests Table 4-32 PW Operation Tests (C358) Step Task Expected Result(s) Select the Abdomen preset from the Abdomen preset options, and set the 2D Gain to 45.
  • Page 171: 739L Probe Image Quality Tests

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 4-4-4 739L Probe Image Quality Tests 4-4-4-1 2D Noise Uniformity Test Table 4-33 2D Noise Uniformity Test (739L) Step Task Expected Result(s) With a 739L probe in the air, select the Carotid preset from the Vascular preset options.
  • Page 172: Cfm Stationary Noise Test

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 4-4-4-3 CFM Stationary Noise Test Table 4-35 CFM Stationary Noise Test (739L) Step Task Expected Result(s) Use the vertical nail area in the Electronic Flow Phantom RMI1425A.
  • Page 173: Cfm Operation Tests

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 4-4-4-5 CFM Operation Tests Table 4-37 CFM Operation Tests (739L) Step Task Expected Result(s) Select a Cardiac preset and set the 2D Gain to 40. Activate CFM and set the following CFM parameters: •...
  • Page 174: Pw Operation Tests

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 4-4-4-6 PW Operation Tests Table 4-38 PW Operation Tests (739L) Step Task Expected Result(s) Select the Abdomen preset from the Abdomen preset options, and set the 2D Gain to 90.
  • Page 175: Probe 10S Image Quality Tests

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 4-4-5 Probe 10S Image Quality Tests 4-4-5-1 2D Center Noise Test NOTE: This test should be performed in a dark room. Table 4-39 2D Noise Center Test (10S)
  • Page 176: System Turnover Checklist

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Section 4-5 System Turnover Checklist Before returning the Vivid™ 4 ultrasound scanner to regular operational use, the System Configuration settings should be checked. In 2D-Mode, press Config on the keyboard and check the settings in...
  • Page 177 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Figure 4-1 System Tab Configuration - Example Chapter 4 - Functional Checks 4-27...
  • Page 178 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Figure 4-2 System Options Tab Configuration - Example 4-28 Section 4-5 - System Turnover Checklist...
  • Page 179 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Figure 4-3 Hospital Info Tab Configuration - Example Chapter 4 - Functional Checks 4-29...
  • Page 180 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Figure 4-4 VCR/ECG Tab Configuration - Example 4-30 Section 4-5 - System Turnover Checklist...
  • Page 181 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Figure 4-5 Archive Tab Configuration - Example Chapter 4 - Functional Checks 4-31...
  • Page 182 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Figure 4-6 Connectivity Local Settings Tab Configuration - Example 4-32 Section 4-5 - System Turnover Checklist...
  • Page 183: Site Log

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Section 4-6 Site Log Table 4-1: Vivid™ 4 Site Log Date Service Person Problem Comments Chapter 4 - Functional Checks 4-33...
  • Page 184 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Table 4-1: Vivid™ 4 Site Log (Continued) Date Service Person Problem Comments 4-34 Section 4-6 - Site Log...
  • Page 185: Overview

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Chapter 5 Components and Function (Theory) Section 5-1 Overview 5-1-1 Purpose of Chapter 5 This chapter contains block diagrams and functional explanations of the Vivid™ 4 electronic circuits.
  • Page 186: General Information

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Section 5-2 General Information The Vivid™ 4 is a phased, linear array ultrasound imaging scanner. The system is versatile and, depending upon the software, can be used for various applications, such as:...
  • Page 187: Block Diagrams

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Section 5-3 Block Diagrams 5-3-1 System Block Diagrams The following system block diagrams are included in this section: • System Block Diagram - Vivid™ 4 BT03 Systems with RFI Configuration...
  • Page 188 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL PCVIC Figure 5-2 System Block Diagram - Vivid™ 4 BT03 Systems with RFT Configuration Section 5-1 - Overview...
  • Page 189 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL SYSTEM MONITOR 17” FB200590 COLOR PRINTER 2269429 Speakers R/G/B/ Sync 2269309 2391918 Track Color Printer Trig. K.B Assy K/B A/N K/B Matrix From Ac Ball On/off...
  • Page 190 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL SYSTEM MONITOR 17“ P/N FB200590 COLOR PRINTER Speaker R/G/B/ Sync 2269309 2391918 Track Color Printer Trig. K.B Assy K/B A/N K/B Matrix From Ac Ball On/off...
  • Page 191: Front End

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Section 5-4 Front End 5-4-1 General Information The Front End includes all the boards in the Front End Card Cage, as follows: • Front End Board •...
  • Page 192 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Figure 5-5 Front End Crate Block Diagram - RFI Configuration Section 5-1 - Overview...
  • Page 193 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 64Ch 64Ch Beam Former Beam Former Front Front Board Board Board Board Board Board TR4 Boards TR4 Boards Board Board FE bus FE bus Pipeline Pipeline...
  • Page 194: Front End Bus

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 5-4-1-1 Front End Bus Most of the FE boards are controlled by the Front End Controller or RFI board via a Front End Bus (FE_BUS). The FE_BUS is a synchronous and bidirectional bus built up by 16-bit data, 6-bit device address select, 10-bit page address and 8 strobe signals controlling the data transactions.
  • Page 195: Transmitter Power Supply

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL • The Radio Frequency Interface (RFI) board controls all Front End boards (on systems with the RFI configuration). The board loads all parameters to the FB, MUX and Beamformer RAMs, it reads the probe identification, selects probe connector on TR4 boards and controls the high voltage multiplexer in linear probes.
  • Page 196: And Mla-1 Buses (Rfi Configuration Only)

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 5-4-1-4-2 MLA-0 and MLA-1 Buses (RFI configuration only) • The Mid Processors are interconnected through a data bus system called the MLA-0 and MLA-1. These are uni-directional buses, transporting data from the Beamformer to the Demodulator FPGA on the RFI.
  • Page 197 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL The probe elements receive the echo signals from different depths. These signals are amplified and conditioned at the Analog Time Gain Compensation (ATGC) control. The different operating modes dictate in which way the signals are translated, as follows: •...
  • Page 198 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Figure 5-8 Received Signal Path Block Diagram - Systems with RFI Configuration 5-14 Section 5-1 - Overview...
  • Page 199 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Figure 5-9 Received Signal Path Block Diagram - Systems with RFT Configuration Chapter 5 - Components and Function (Theory) 5-15...
  • Page 200: Front Board Assembly (Fb)

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 5-4-2 Front Board Assembly (FB) 5-4-2-1 General Description The FB Board has 64 identical receive channels. The signal in each channel is fed via a Transmit/Receive (T/R) switch to a preamplifier. The next step is a Time Controlled Gain (TGC) amplifier used to equalize the signal strength from the near-field, the mid-field and the far-field.
  • Page 201: Tr4 Board Description

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 5-4-2-2 TR4 Board Description The TR4 Board has four transmit and receive signal channels. Each channel has two pulsars for different transmitted Tx voltages, and a low-signal receiver amplifier. Both are connected to the same input/output line.
  • Page 202: Mux Board

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 5-4-3 MUX Board 5-4-3-1 General Description The MUX Board has 64 identical receive channels, and performs the following functions: • Receiving the channels and connecting them to the BF Boards.
  • Page 203: Beamformer Board (Bf)

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 5-4-4 Beamformer Board (BF) 5-4-4-1 General Description The Beamformer Board (BF Board) contains the A/D converter, the ASIC (called the Focuser [FOC]) and a Beam Adder (BA). The BF Board accepts input channels and converts each channel to digital bytes.
  • Page 204: Description Of Operation

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 5-4-4-2 Description of Operation AD C AD C FOC 2 MLA0 To next board AD C (1 of 16) AD C SRAM & Bus- FE bus...
  • Page 205: Radio Frequency Interface (Rfi) Board

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 5-4-5 Radio Frequency Interface (RFI) Board 5-4-5-1 General Description The Radio Frequency Interface (RFI) board has been introduced in the manufacture of Vivid™ 4 systems, as a result of certain hardware components having become obsolete.
  • Page 206: Data Buffer

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 5-4-5-4 Data Buffer After RF filtering, the data is written into different sliding ring buffers, dedicated to the different types of data. While data is written into the buffers sample-by-sample in vectors, multiple samples from the same range (depth) can be read out.
  • Page 207: Tx Power Supply Management

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 5-4-5-10 TX Power Supply Management The TX Power supply has two separate voltage outputs, as follows: • TX Power Supply #1 (TX1) - generates voltage levels for pulsed TX operations •...
  • Page 208 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 5-4-5-11 Table 5-3 RFI - Test Points Test Point # Description Comments EMIFB CLOCK 133 MHz Doppler Probe Sense Voltage Tempsense Output 130.8 mV/ MAX158 INT...
  • Page 209 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Table 5-3 RFI - Test Points (Continued) Test Point # Description Comments 2.5 V voltage 3.3 V voltage 5 V voltage - 5 V voltage -2.0 V voltage...
  • Page 210: Front End Controller Board (Fec)

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 5-4-6 Front End Controller Board (FEC) The Front End Controller (FEC) Board controls other boards in the Front End Crate. The control is done through a synchronous and bidirectional Bus called the Front End Bus. The following tasks are performed by the FEC Board: •...
  • Page 211: Rf And Tissue Processor Board (Rft)

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 5-4-7 RF and Tissue Processor Board (RFT) The RF and Tissue Processor Board (RFT) receives data from the BF1 Board in the Front End. Both types of data samples (RF_MODE) are communicated to the RFT Board from the Front End Controller Board, via the Front End Bus.
  • Page 212: Image Port Board (Imp)

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 5-4-8 Image Port Board (IMP) The function of the Image Port (IMP) is to receive signals from either the RFT (during scanning) or the VCR (during playback), capture them, and route them to the Back End Processor (BEP) via the PC2IP Bus.
  • Page 213: Back End Processor

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Section 5-5 Back End Processor 5-5-1 Introduction The Back End Processor (BEP) supports the operation of the Vivid™ 4 ultrasound unit and is the main controller for the unit. The BEP comprises the various components described in the following sections: •...
  • Page 214 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Figure 5-16 Back End Component Locations - Vivid™ 4 BT03 Systems with RFI Configuration 5-30 Section 5-1 - Overview...
  • Page 215 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL MO Driver MO Driver CD-R Driver CD-R Driver ECG Module ECG Module Floppy Driver Floppy Driver HD Driver HD Driver VGA Card VGA Card AGP Slot AGP Slot Plug &...
  • Page 216 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Figure 5-18 Back End Block Diagram - Vivid™ 4 BT03 Systems with RFI Configuration 5-32 Section 5-1 - Overview...
  • Page 217 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL MO Driver MO Driver ECG Module ECG Module CD-R Driver CD-R Driver Floppy Driver Floppy Driver HD Driver HD Driver S-VGA S-VGA Monitor Monitor Audio Bracket...
  • Page 218: Central Processing Unit (Cpu)

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 5-5-2 Central Processing Unit (CPU) The CPU controls and processes the internal Back End Processor operations. In addition, the CPU supports the Front End Crate via the IP card. The CPU used in the Vivid™ 4 ultrasound unit is a Pentium 4, 2GHz Intel Processor, or newer.
  • Page 219 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Note: Not drawn to scale Figure 5-20 BEP Motherboard Layout - Vivid™ 4 BT03 Chapter 5 - Components and Function (Theory) 5-35...
  • Page 220 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Table 5-4 BEP Motherboard Layout BT03 - Legend to Figure 5-20 Jumpers Description Default Setting CMOS Clear Open (Disabled) Front Side Bus Speed Pins 1-2 (Auto)
  • Page 221: Keyboard Controller

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 5-5-3 Keyboard Controller The keyboard controller controls all the extended keyboard operations, power management systems and audio control. Keyboard Controller Board Figure 5-21 Keyboard Controller Block Diagram The Keyboard Controller contains the Audio Amplifier that controls the volume of the front cover speakers.
  • Page 222: Multifunction I/O Controller

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 5-5-4 Multifunction I/O Controller The Multifunction I/O Controller provides two additional serial COMs, COM3 and COM4, that are used to support the modem functions. 5-5-5 Frame Grabber (RFI systems only)
  • Page 223: Floppy Drive

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 5-5-10 Floppy Drive The Floppy Drive supports basic data retrieval from the Vivid™ 4 ultrasound unit, such as log files, export to Microsoft Excel, small image files and other service utility operations. The floppy drive is controlled by the CPU via signals from the IDE Bus.
  • Page 224: Magneto-Optical Drive (Mod)

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 5-5-12 Magneto-Optical Drive (MOD) The MOD is used to back-up images and reports. Backing up enables storage of the images and reports for future review on Vivid™ 4 ultrasound units, or other EchoPac-compatible devices. The MOD is controlled by the CPU via the SCSI Card.
  • Page 225: Modem

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 5-5-15 Modem The Modem is a universal modem controlled by the CPU via the USB port. The modem is used for iLinq (InSite) remote servicing. This is a standard MultiTech modem that connects to an analog phone line.
  • Page 226: Pc-Vic Assembly

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 5-5-16 PC-VIC Assembly 5-5-16-1 General Description The PC-VIC assembly comprises the PC-VIC board and the I/O board. The I/O board contains inputs and outputs; it receives all input signals, filters them, and transmits them either to another external peripheral or to the PC-VIC board.
  • Page 227: Pal

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 5-5-16-2 Table 5-5 PAL Video Specifications Horizontal Timing 944 pixels Horizontal Total (HT) 152 pixels Horizontal Start (HS) 768 pixels Horizontal Active (HA) 14.7 MHz (PAL)
  • Page 228: External Peripherals

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Section 5-6 External Peripherals 5-6-1 Introduction The following external peripherals are used with the Vivid™ 4 ultrasound scanner: • VCR (connected to Internal I/O) • Black & White Video Printer (connected to Internal I/O) •...
  • Page 229: Vivid™ 4 Power Distribution

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Section 5-7 Vivid™ 4 Power Distribution 5-7-1 Electrical Power The Vivid™ 4 system can be set up to operate from the following 50/60 Hz AC voltages: 100V, 120V, 220V, 230V and 240V.
  • Page 230: Ac System

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 5-7-2 AC System The AC System (see Figure 5-26) comprises the following components: • AC Isolation Transformer. • AC Distribution Box, as described in AC Distribution Box on page 5 - 47.
  • Page 231: Ac Distribution Box

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 5-7-3 AC Distribution Box The AC Distribution Box contains an electronic board that selects the proper Input and Output system AC voltages, the load connectors, and the low voltage 12 V power supply. In addition, the AC Distribution Box controls the soft AC Start Circuit, the temperature sensor and the Fan Control Circuit.
  • Page 232: Input Ac Voltage Configuration

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 5-7-3-3 Input AC Voltage Configuration • The Internal Vivid™ 4 voltage is 230 V AC. • Back End and Front End Crates operate on 230 V AC.
  • Page 233: Front End Dc Power Distribution

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 5-7-4 Front End DC Power Distribution Figure 5-28 DC Distribution: Front End Crate (Vivid™ 4 Systems with RFI Configuration) Chapter 5 - Components and Function (Theory)
  • Page 234 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Front Bd MUX Bd (BF2) Image Port DC-PWR +5VA (15A) +5VA AVCC AVCC (4A) +10 (3A) +15 (0.6A) -15 (0.5A) AVEE (10A) AVEE AVEE AVEE DC-PWR...
  • Page 235: Front End Cooling System

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Section 5-8 Front End Cooling System 5-8-1 General Description The Vivid™ 4 Front End cooling system includes following components: • Dust Filter • Fans Other components that have an influence on the airflow is the Front End itself. Air is sucked from the inside of the system through the Filter Cover and the Filter.
  • Page 236: Common Service Platform

    5-9-1 Introduction The Service Platform contains a set of software modules that are common to many of GE’s ultrasound and cardiology systems. This web-enabled technology provides linkage to e-Services, e-Commerce, and the iCenter, making GE’s scanners more e-enabled than ever.
  • Page 237: Global Service User Interface (Gsui)

    GEMS products. 5-9-3-1 Internationalization The user interfaces provided by the service platform are designed for GE personnel and, as such, are in English only. At this time, there is no multi-lingual capability built into the Common Service Interface. 5-9-3-2...
  • Page 238: Access / Security

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 5-9-3-3 Access / Security The service interface has different access and security user levels. Users are only granted access to the tools they are authorized to use.
  • Page 239: Overview

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Chapter 6 Service Adjustments Section 6-1 Overview 6-1-1 Purpose of Chapter 6 Table 6-1 Contents in Chapter 6 Section Description Page Number Overview Input AC Voltage Configuration...
  • Page 240: Input Ac Voltage Configuration

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Section 6-2 Input AC Voltage Configuration • The internal Vivid™ 4 voltage is 230V AC. • The Back End and Front End Crates operate on 230V AC.
  • Page 241: Front End Voltages And Signal Indicators

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Section 6-3 Front End Voltages and Signal Indicators The majority of the boards in the Front End Crate have red, green and yellow LEDs, which indicate the board’s status as follows:...
  • Page 242 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Figure 6-2 Front End Crate LEDs - RFT Configuration The LEDs for each of the boards in the Front End Crate are summarized in the following subsections.
  • Page 243: Rfi Leds

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 6-3-1 RFI LEDs Table 6-3 RFI - LEDs LED # LED Color Description Normal State LD505 Demod Error - Red LD504 Lit in Scanning Mode Slow tick in idle...
  • Page 244: Front End Controller (Fec) Leds

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 6-3-3 Front End Controller (FEC) LEDs Table 6-5 Front End Controller LEDs Indicates When Lit Comments Board Failure Normally OFF Green Board OK Normally lit Yellow...
  • Page 245: Channels Multiplexer (Mux)

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 6-3-6 Channels Multiplexer (MUX) Table 6-8 Channels Multiplexer LEDs Indicates When Lit Comments Green 5V PS is OK Normally lit Green 5V PS is OK Normally lit...
  • Page 246: Back End Power Supply Voltages

    4) Click the Technical Support tab and make sure Frequency is set to either 60Hz for NTSC, or 50Hz for PAL. 5) Click OK. Note: If the video format is not set correctly, contact an authorized GE Service Representative. Section 6-4 - Back End Power Supply Voltages...
  • Page 247: Vivid™ 4 17" Monitor Operation & Image Quality Calibration

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Section 6-6 Vivid™ 4 17" Monitor Operation & Image Quality Calibration 6-6-1 Monitor Operation The settings that can be adjusted on the Vivid™ 4 17" monitor are the features that affect image quality, namely Contrast and Brightness.
  • Page 248: Image Quality Calibration

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 6-6-2 Image Quality Calibration The image quality settings on the Vivid™ 4 17" monitor may need periodic adjustment due to changes in ambient light. These adjustments can be easily made using the Contrast and Brightness buttons on the front part of the display monitor.
  • Page 249: Calibration

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Section 6-7 Calibration The Diagnostic dialog box (accessed by pressing <Ctrl+F11> or <Alt+D>, when in regular scanning mode), enables you to perform three types of calibration, as described in the following sections: •...
  • Page 250 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 2) Trackball to the Calibration button and press Select. The following calibration options are displayed: Figure 6-6 Calibration Options 6-12 Section 6-7 - Calibration...
  • Page 251: Monitor Calibration

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 6-7-2 Monitor Calibration If required, Monitor Calibration may be performed after adjusting the image quality settings, using the monitor’s control buttons, as described in the Image Quality Calibration section, page 6-10.
  • Page 252: Beamformer Calibration

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 6-7-3 Beamformer Calibration NOTE: IMPORTANT - Do not perform BF Calibration after operating in CW or PW scanning modes since validation will fail. Perform the BF Calibration procedure only after 2D scanning mode.
  • Page 253 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 3) Connect the Y/C cable from Video OUT to Video IN on the right rear panel. 4) Select the PatternB&W.bmp from the Test with one of bitmaps area. You will observe the duplication of the selected pattern.
  • Page 254: Diagnostics/Troubleshooting

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Chapter 7 Diagnostics/Troubleshooting Section 7-1 Overview 7-1-1 Purpose of Chapter This chapter describes how to set up and run diagnostic tools to locate system problems and failures.
  • Page 255: Diagnostics

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Section 7-2 Diagnostics 7-2-1 Diagnostic Tools The diagnostic tools check the system parts, as well as power supplies, temperature, fan operation, board functions, Back-end signal processing performance, keyboard operation, and so on.
  • Page 256: Accessing The Diagnostic Menu

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 7-2-3 Accessing the Diagnostic Menu The diagnostic tools are accessed from 2D-Mode by simultaneously pressing <Ctrl+F11> on the alphanumeric keyboard. The Diagnostic menu is displayed, as shown below:...
  • Page 257: Diagnostic Menu Options

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 7-2-3-1 Diagnostic Menu Options The Diagnostic menu includes the following options: • Software Version: Displays the Vivid™ 4 ultrasound unit software version number and last upgrade date.
  • Page 258: Performing Front End (Fe) Diagnostics

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Section 7-3 Performing Front End (FE) Diagnostics When performing Front End Diagnostics tests, it is strongly recommended to start with Auto Sequence, especially if the system shows signs of an intermittent problem. As described in this section, the testing sequence can be paused and restarted at any stage, providing a quick and easy method of reviewing any detected faults.
  • Page 259: Accessing The Front End Diagnostic Options

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 7-3-1 Accessing the Front End Diagnostic Options • In 2D-Mode, press <Ctrl+F11> on the alphanumeric keyboard. The Diagnostic menu is displayed, as shown in Figure 7-1 on page 7-3.
  • Page 260 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Auto Sequence commences. While the sequence of automatic testing proceeds, the name of the test currently in progress is displayed in the Test Name field and the current Loop # and Step # are shown below.
  • Page 261: Manual Options

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 7-3-1-2 Manual Options 1.) In the Diagnostic menu, trackball to the Front End button and press Select. The FE Diagnostics dialog box is displayed, as shown in...
  • Page 262 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Data Flow Map Test Description Area Test Area Comment Area Tabs Figure 7-4 FE Diagnostics Dialog Box - RFT Configuration Chapter 7 - Diagnostics/Troubleshooting...
  • Page 263 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL The FE Diagnostics dialog box includes the following options: Table 7-2 FE Diagnostics Dialog Box Options Option Description Comments Each button has an LED that indicates the...
  • Page 264: Radio Frequency Interface (Rfi) Diagnostic Tests (For Rfi Configuration)

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 7-3-2 Radio Frequency Interface (RFI) Diagnostic Tests (for RFI Configuration) 1) Access the FE Diagnostics dialog box, as described in the Accessing the Front End Diagnostic Options section, on page 7-6.
  • Page 265 If no problems were identified, proceed to re-boot the system, as prompted (Figure 7-6, above). • If no problems were identified but the problem persists, contact your local GE Medical Systems field engineer for assistance. • If errors were identified but the results were marginal, perform calibration, as described in the Calibration section, on page 6-11, and then repeat the RFI test.
  • Page 266: Image Port (Imp) Diagnostic Tests (For Rft Configuration)

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 7-3-3 Image Port (IMP) Diagnostic Tests (for RFT Configuration) 1) Access the FE Diagnostics dialog box, as described in the Accessing the Front End Diagnostic Options section, on page 7-6.
  • Page 267 Diagnostic Tests (for RFT Configuration) section, on page 7-15. • If no problems were identified but the problem persists, contact your local GE Medical Systems field engineer for assistance. • If errors were identified but the results were marginal, perform calibration, as described in the Calibration section, on page 6-11, and then repeat the IMP test.
  • Page 268: Vme Bus (Vme) Diagnostic Tests (For Rft Configuration)

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 7-3-4 VME Bus (VME) Diagnostic Tests (for RFT Configuration) The VME Bus diagnostic test checks the communication on the VME Bus between the IP card, the FEC and the RFT.
  • Page 269 Tests (for RFT Configuration) section, on page 7-17. • If no problems were identified but the problem persists, contact your local GE Medical Systems field engineer for assistance. • If errors were identified but the results were marginal, perform calibration, as described in Calibration on page 6 - 11, and then repeat the VME test.
  • Page 270: Rft Diagnostic Tests (For Rft Configuration)

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 7-3-5 RFT Diagnostic Tests (for RFT Configuration) The RFT diagnostic tests include the tests listed below: • Board Test: General RFT board test which activates the remaining tests sequentially. Each of these tests can also be performed independently.
  • Page 271 (FEC) Diagnostic Tests (for RFT Configuration) section, on page 7-19. • If no problems were identified but the problem persists, contact your local GE Medical Systems field engineer for assistance. • If errors were identified but the results were marginal, perform calibration, as described in the Calibration section, on page 6-11, and then repeat the RFT test.
  • Page 272: Front End Controller (Fec) Diagnostic Tests (For Rft Configuration)

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 7-3-6 Front End Controller (FEC) Diagnostic Tests (for RFT Configuration) The FEC diagnostic tests include the following: • Board Test: General board test which activates the remaining tests sequentially. Each of these tests can also be performed individually.
  • Page 273 Diagnostic Tests section, on page 7-21. • If no problems were identified but the problem persists, contact your local GE Medical Systems field engineer for assistance. • If errors were identified but the results were marginal, perform calibration, as described in the Calibration section, on page 6-11, and then repeat the test.
  • Page 274: Beamformer (Bf) Diagnostic Tests

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 7-3-7 Beamformer (BF) Diagnostic Tests The BF diagnostic tests enable you to check the ADC output and perform a digital beamformer test. Note: Before performing the BF tests, perform the IMP, VME Bus, RFT and FEC tests, as described beginning on page 7-13.
  • Page 275 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Data Flow Map Test Description Area Test Area Comment Area Tabs Figure 7-11 BF Tab - FE Diagnostics Dialog Box (RFI Configuration) 7-22 Section 7-3 - Performing Front End (FE) Diagnostics...
  • Page 276 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Data Flow Map Test Description Area Test Area Comment Area Tabs Figure 7-12 BF Tab - FE Diagnostics Dialog Box (RFT Configuration) 3) Proceed as follows: Trackball to one of the following buttons and press Select: •...
  • Page 277 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Figure 7-13 Ultrasound Scope Window 6) Under Auto Detect, click the Start button. Checking of the test results commences and the light adjacent to the button flashes green and continues to flash while Auto Detect is in progress. If a fault is detected, the flashing light changes to red and an error log is displayed listing the channel/s on which the test failed.
  • Page 278 Assembly (FB) Diagnostic Tests section, on page 7-26. • If no problems were identified but the problem persists, contact your local GE Medical Systems field engineer for assistance. • If errors were identified but the results were marginal, perform calibration, as described in the Calibration section, on page 6-11, and then repeat the test.
  • Page 279: Front Board Assembly (Fb) Diagnostic Tests

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 7-3-8 Front Board Assembly (FB) Diagnostic Tests The FB diagnostics tests check the FB circuits and boards. To avoid frequent rebooting of the system, these tests must be performed in the sequence described below. When performing FB tests out of sequence, follow the instructions displayed in the Comment area of the FE Diagnostics dialog box, and reboot as required.
  • Page 280 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Data Flow Map Test Description Area Test Area Comment Area Tabs Figure 7-14 F.B. Tab - FE Diagnostics Dialog Box (RFI Configuration) Chapter 7 - Diagnostics/Troubleshooting...
  • Page 281 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Data Flow Map Test Description Area Test Area Comment Area Tabs Figure 7-15 F.B. Tab - FE Diagnostics Dialog Box (RFT Configuration) 3) Trackball to the following buttons (in the order shown) and press Select: a.) Transmit Test: Checks the FE pulsers via the probe elements, as described in...
  • Page 282: Transmit Test

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 7-3-8-2 Transmit Test If a solid dark line is observed during scanning, it may indicate a bad transmitting channel. The transmit test checks the FE pulsers via the probe elements, enabling you to check the transmitting channels. The transmit test is not sensitive to the receiving channels.
  • Page 283 • If no errors were identified, proceed with the next test. • If no errors were identified but the problem persists, contact your local GE Medical Systems field engineer for assistance. • If errors were identified but the results were marginal, perform calibration, as described in the Calibration section, on page 6-11, and then repeat the test.
  • Page 284: Fla Jumper Switch

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 7-3-8-3 FLA Jumper Switch All the elements are connected through the transmit and receive jumpers. This test checks the jumpers that connect the side elements to the center elements. During this test, every probe element transmits and receives signals.
  • Page 285 Trackball to the Mark bad channel checkbox and press Select. • Trackball to Save Report and press Select, and contact a GE Medical Systems service representative. 7) Trackball to the Start Close test button and press Select. The closed FLA jumper test is performed.
  • Page 286 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Figure 7-18 Report Dialog Box - FLA Jumper Test (Closed) The Report dialog box contains the following information: • Test Result Message: Displays the test result.
  • Page 287 • If no errors were identified, proceed with the next test. • If no errors were identified but the problem persists, contact your local GE Medical Systems field engineer for assistance. • If errors were identified but the results were marginal, perform calibration, as described in the Calibration section, on page 6-11, and then repeat the test.
  • Page 288: Receive Test

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 7-3-8-4 Receive Test The Receive Test checks each amplifier output against the predefined output level. This system test is performed in Phased Array Probe Simulation mode. The FEC board injects analog signals to the Tx amplifiers inputs after the Tx switch.
  • Page 289 Trackball to the Mark bad channel checkbox and press Select. • Trackball to Save Report and press Select, and contact a GE Medical Systems service representative. 6) Trackball to the Low Freq Test button and press Select. The receive test is performed in low...
  • Page 290 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Figure 7-20 Report Dialog Box - Receive Test Low Frequency The Report dialog box contains the following information: • Test Result Message: Displays the test result.
  • Page 291 • If no errors were identified, proceed with the next test. • If no errors were identified but the problem persists, contact your local GE Medical Systems field engineer for assistance. • If errors were identified but the results were marginal, perform calibration, as described in the Calibration section, on page 6-11, and then repeat the test.
  • Page 292: Tr Switch Test

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 7-3-8-5 TR Switch Test If a solid line is observed during scanning, it may indicate a bad TR switch. This test checks the transmit switches. 1) Connect a 3S probe and lock it in place.
  • Page 293 • If no errors were identified, proceed with the next test. • If no errors were identified but the problem persists, contact your local GE Medical Systems field engineer for assistance. • If errors were identified but the results were marginal, perform calibration, as described in the Calibration section, on page 6-11, and then repeat the test.
  • Page 294: Pencil Probe Test

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 7-3-8-6 Pencil Probe Test If a solid black line is observed during scanning, it may indicate a bad transmitting channel. The pencil probe test checks the transmitting channels associated relays to the pencil probe port. This test is not for sensitive receiving channels.The test has two phases, namely the open pencil probe test and the...
  • Page 295 Trackball to the Mark bad channel checkbox and press Select. Repeat as required to mark additional channels. • Trackball to Save Report and press Select, and contact a GE Medical Systems service representative. 6) Trackball to the Start Close test and press Select. The pencil probe test (close) is performed and...
  • Page 296 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Figure 7-23 Report Dialog Box - Pencil Probe Test (Close) The Report dialog box includes the following: • Test Result Message: Displays the test result. •...
  • Page 297 • If no errors were identified, proceed with the next test. • If no errors were identified but the problem persists, contact your local GE Medical Systems field engineer for assistance. • If errors were identified but the results were marginal, perform calibration, as described in the Calibration section, on page 6-11, and then repeat the test.
  • Page 298: Port Test

    • If no errors were identified, proceed with the next test. • If no errors were identified but the problem persists, contact your local GE Medical Systems field engineer for assistance. • If errors were identified but the results were marginal, perform calibration, as described in the Calibration section, on page 6-11, and then repeat the test.
  • Page 299: Ram Test

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 7-3-8-8 RAM Test The RAM test enables you to check the RAM that stores the delays for the transmitted ultrasound signals to the form focal beam.
  • Page 300 • If no errors were identified, proceed with the next test. • If no errors were identified but the problem persists, contact your local GE Medical Systems field engineer for assistance. • If errors were identified but the results were marginal, perform calibration, as described in the Calibration section, on page 6-11, and then repeat the test.
  • Page 301: Mux Diagnostic Tests

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 7-3-9 MUX Diagnostic Tests NOTE: The Select Socket and Select Supply options are not relevant for this test. 7-3-9-1 Accessing the MUX Diagnostic Test Options 1) Access the FE Diagnostics dialog box, as described in the...
  • Page 302 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Data Flow Map Test Description Area Test Area Comment Area Tabs Figure 7-27 MUX Tab - FE Diagnostics Dialog Box (RFT Configuration) Chapter 7 - Diagnostics/Troubleshooting...
  • Page 303: Normal Channels Diagnostic Tests

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 3) Trackball to each of the following buttons in turn (in the order stated below) and press Select to perform the specific test: • Normal channels: Test the output channels from the input channels, as described in the Normal Channels Diagnostic Tests below.
  • Page 304 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Figure 7-28 Normal Channels Diagnostic Test Chapter 7 - Diagnostics/Troubleshooting 7-51...
  • Page 305 Trackball to the Mark bad channel checkbox and press Select. • Trackball to Save Report and press Select, and contact a GE Medical Systems service representative. 5) Trackball to the Low Freq Test button and press Select. The normal channels diagnostic test is...
  • Page 306 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Figure 7-29 Report Dialog Box - Normal Channels Test Low Frequency The Report dialog box contains the following information: • Test Result Message: Displays the test result.
  • Page 307 • If no errors were identified, proceed with the next test. • If no errors were identified but the problem persists, contact your local GE Medical Systems field engineer for assistance. • If errors were identified but the results were marginal, perform calibration, as described in the Calibration section, on page 6-11, and then repeat the test.
  • Page 308: Expanded Low Channels Diagnostic Tests

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 7-3-9-3 Expanded Low Channels Diagnostic Tests Tests output signals in CW-Mode only, when the input channels are connected to the output channels. If one channel is missing, the output will be -6db. The input signals come from either a 3S or a 7S probe.
  • Page 309: Expanded High Channels Diagnostic Tests

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 7-3-9-4 Expanded High Channels Diagnostic Tests Tests the output signals from the input of the high channels. If one channel is missing, the output will be -6db. The input signals come from either a 3S or a 7S probe.
  • Page 310: Port Test

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 7-3-9-5 Port Test The Port test is a general logic control test. Figure 7-32 Port Test (RFI Configuration) Chapter 7 - Diagnostics/Troubleshooting 7-57...
  • Page 311 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Figure 7-33 Port Test (RFT Configuration) 7-58 Section 7-3 - Performing Front End (FE) Diagnostics...
  • Page 312: Advanced Board Tests

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 7-3-9-6 Advanced Board Tests The advanced board tests check two additional parts of the MUX Board: TEE probe control and Monitoring, as follows: TEE probe control enables you to check different positions of the probe; by changing the probe angle, you are able to verify its normal operation.
  • Page 313 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Figure 7-35 TEE Probe Test Dialog Box 5) Verify the test results, and then trackball to the OK button and press Select. 7-60 Section 7-3 - Performing Front End (FE) Diagnostics...
  • Page 314: 7-3-9-6-2 Ac Power Input Test

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 7-3-9-6-2 AC Power Input Test The AC power input test measures the AC input voltage deviation from the nominal input voltage (accuracy by percentage). 1) Access the MUX tab of the FB Diagnostics dialog box, as described in the...
  • Page 315 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 4) Trackball to the AC Power Input button and press Select. The AC Power Input dialog box is displayed: Figure 7-37 AC Power Input Dialog Box 5) Trackball to the Show button and press Select to display the voltage.
  • Page 316: 7-3-9-6-3 Lvps Voltage Test

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 7-3-9-6-3 LVPS Voltage Test The LVPS test measures the LVPS voltages. The voltage range is displayed next to the measured voltage. All the voltages can be measured with DVM on the front board test point.
  • Page 317: 7-3-9-6-4 Hvps Voltage Test

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 7-3-9-6-4 HVPS Voltage Test The HVPS test measures the Tx voltages. The current values are displayed next to the measured values. All the voltages can be measured with DVM on the front board test point.
  • Page 318: 7-3-9-6-5 Temp Test

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 7-3-9-6-5 Temp Test The Temp test displays the temperature at the Front End crate, as well as the temperature of the AC distribution box. 1) Access the MUX tab of the FB Diagnostics dialog box, as described in the...
  • Page 319: 7-3-9-6-6 Show History

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 7-3-9-6-6 Show History The Show History option in the View Power dialog box displays selected test results by year, month and day. 1) Access the MUX tab of the FB Diagnostics dialog box, as described in the...
  • Page 320 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 7-3-10 H/W Report The H/W Report button in the FE Diagnostics dialog box enables you to generate a report displaying the Front End boards, software versions and hardware revisions.
  • Page 321: Current Report

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 7-3-11 Current Report The H/W Report button in the FE Diagnostics dialog box enables you to generate a report of the current test and failure results.
  • Page 322: Performing Back End Diagnostics On The System

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Section 7-4 Performing Back End Diagnostics on the System The Back End Diagnostics enable you to test the functionality of various Back End hardware components. Note: The Back End Diagnostic Tests may be performed independently, regardless of sequence.
  • Page 323 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Note: The following test options are available in the HW Diagnostics menu. These tests may be performed independently, regardless of sequence. • Audio: Enables you to test the doppler sound, as described in the...
  • Page 324: Accessing The Back End Diagnostic Options

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 7-4-2 Audio (Doppler Sound Driver) Diagnostic Test The audio test enables you to test the Doppler sound at different PRFs and frequencies. 1) Access the HW Diagnostics menu, as described in the Accessing the Back End Diagnostic Options section, on page 7-69.
  • Page 325: Ecg/Phono Diagnostic Test

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 7-4-3 ECG/Phono Diagnostic Test The ECG and Phono tests enable you to check the output of the ECG and Phono options. 7-4-3-1 ECG Diagnostic Test 1) Connect the internal ECG input to yourself or to an appropriate ECG simulator, using the cable set and electrode pads.
  • Page 326: Phono Test

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 7-4-3-2 Phono Test 1) Connect the media microphone to the microphone input on the ECG board. 2) Access the HW Diagnostics menu, as described in the Accessing the Back End Diagnostic Options section, on page 7-69.
  • Page 327: External Keyboard Diagnostic Test

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 7-4-4 External Keyboard Diagnostic Test The external keyboard diagnostic tests enable you to test the buttons and lights on the keyboard in different modes. Note: To test the keys on the alphanumeric keyboard, refer to the...
  • Page 328: Automatic Lights Test

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Figure 7-49 KB Panel Tab - Extended Keyboard Test Window 5) Press each of the control panel buttons (except for Select, Menu and the On/Off button), and verify that the correct buttons are activated on-screen in the KB Panel tab.
  • Page 329: Manual Lights Test

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 7-4-4-3 Manual Lights Test 1) Access the HW Diagnostics menu, as described in the Accessing the Back End Diagnostic Options section, on page 7-69. 2) Trackball to the External Keyboard button and press Select. The Extended Keyboard Test window...
  • Page 330: Keyboard Diagnostic Test

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 7-4-5 Keyboard Diagnostic Test The keyboard diagnostic test enables you to check the functionality of the alphanumeric keyboard keys. 1) Access the HW Diagnostics menu, as described in the Accessing the Back End Diagnostic Options section, on page 7-69.
  • Page 331: Media Driver Diagnostic Test

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 7-4-6 Media Driver Diagnostic Test The media driver diagnostic test enables you to check the functionality of the MOD, if installed, as well as the functionality of the CD Writer, if installed.
  • Page 332: Cd Writer Diagnostic Test

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 11) Trackball to the <GetVolInfo> button and press Select. The message Media in drive G has following properties: Media Label = TEST_LABEL, Media File-System Name = FAT is displayed.
  • Page 333: Computer Diagnostic Test

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 7-4-7 Computer Diagnostic Test The computer diagnostic tests enable you to perform a benchmark test for the correct frame rate, as well as check the memory usage.
  • Page 334: Check Disk/Bus

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Figure 7-54 Benchmark Testing Results In the information displayed under CPU Description, Memory and Frame Rate, make sure that the values displayed in the upper line (which indicates current computer results) are higher or equal to those indicated in the lower line.
  • Page 335 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Figure 7-55 Disk/Bus Bench Window Disk Bench: In the information displayed, make sure that the speed values displayed are greater than 23MB/sec. This test checks the speed of information transfer between the hard disk and IDE BUS.
  • Page 336: Ups Test

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 7-4-8 UPS Test The UPS Test is used to verify proper operation of the uninterrupted power supply (UPS) and its connection to the system. This procedure is performed in Standby Mode.
  • Page 337: Checking The Network Adaptors From Windows Device Manager

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 7-4-9 Checking the Network Adaptors from Windows Device Manager NOTE: This check is performed on the system while in Technical Support Mode, as described below. Make sure the T-plug (dongle) is inserted into the parallel or USB port on the Vivid™...
  • Page 338: Common Service Interface

    NOTE: The Common Service Interface uses the Windows version of PC-Doctor. The Service Platform contains a set of software modules that are common to many of GE Medical Systems’ ultrasound and cardiology systems. This web-enabled technology provides linkage to e-Services, e-Commerce and the iCenter, making GE Medical Systems’ scanners more e-enabled than ever.
  • Page 339: Service Log In

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 7-5-2-1 Service Log In 1) From the System Configuration window, make sure that iLinq (the Common Service platform) is enabled under the Technical Support tab, as shown in Figure 7-58 below.
  • Page 340: Access And Security

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Figure 7-59 Profile Manager Window 4) Select InSite from the Profile Name drop-down list to confirm the connection to the local host if you are running the application as a field engineer from a remote location, select Remote from the Profile Name drop-down list.
  • Page 341 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 7) Trackball to the Okay button and press Select. The GEMS Service Home Page window is displayed, as shown below: Figure 7-61 GEMS Service Home Page Window The GEMS Service Home Page window contains details of the system location, the system IP address, the application status and other status information about the system.
  • Page 342: Error Logs Page

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 7-5-2-2 Access and Security The service interface has various different security and user levels. Users are only granted access to the tools they are authorized to use, as described in the following table:...
  • Page 343: Log Viewer

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 7-5-3-1 Log Viewer When the Log Viewer option is selected in the left pane of the Error Logs page, the following five categories are displayed, each of which enable you to view a number of different system logs.
  • Page 344 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL • Power: Displays all the power logs, including errors, monitoring dates and values, as shown below: Figure 7-65 Power Logs • Temperature: Displays all the temperature logs, including errors, monitoring dates and values,...
  • Page 345: 7-5-3-1-2 Utilities

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 7-5-3-1-2 Utilities The Utilities option on the Logs Viewer menu enables you to access the following log utilities: • Plot Log: Enables you to view the results of the Temperature or Power logs in graphical format.
  • Page 346: 7-5-3-1-3 Search

    Figure 7-69 Search for Text in Log or Page 7-5-3-1-4 Filter The Filter option on the Logs Viewer menu is available to users with the GE Service access level. Select the System Logs option to select default options that you want to filter from the System...
  • Page 347: 7-5-3-1-5 Exit

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 7-5-3-1-5 Exit The Exit option on the Log Viewer menu enables you to exit the Log Viewer. Select the Exit Log Viewer option to return to the Common Service Desktop.
  • Page 348 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Each system re-boot will generate a new log file; each log file can be viewed by selecting the appropriate link, as shown in the example in...
  • Page 349 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL If required, a log file can be saved using the Save function from this screen (Figure 7-73) (i.e, when you are connected to the system remotely), or you may save the file to a floppy disk on the system itself.
  • Page 350: Diagnostics Page

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 7-5-4 Diagnostics Page The service platform uses a web-controlled user interface to provide access to common service components and perform diagnostics. There are two types of diagnostics: •...
  • Page 351: 7-5-4-1-1 Instructions Frame

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 7-5-4-1 Diagnostics Page When a test is selected in the left pane of the Diagnostics page, the right pane is divided into three frames, as shown in Figure 7-74 on page 7-98.
  • Page 352 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL • Reset Power Supply: Enables you to reset the Front End power supply: Figure 7-76 Reset Power Supply 7-100 Section 7-5 - Common Service Interface...
  • Page 353 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL • Restart Scanner Software: (System Shutdown) Enables you to remotely shutdown or restart the system. Figure 7-77 Restart Scanner Software Chapter 7 - Diagnostics/Troubleshooting 7-101...
  • Page 354 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL • Save System State to Log: This performs the same system operation as that initiated when pressing <Ctrl> F or <Alt> D. Figure 7-78 Save System State to Log...
  • Page 355: Pc Diagnostics

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 7-5-4-3 PC Diagnostics When PC Diagnostics is selected in the left pane of the Diagnostics page, the diagnostic tests described in the following sections can be performed: •...
  • Page 356: 7-5-4-3-2 Hard Drive Tests

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 7-5-4-3-2 Hard Drive Tests When PC Diagnostics is selected in the left pane of the Diagnostics page, the following hard drive diagnostic tests can be performed: •...
  • Page 357: 7-5-4-3-5 Video Test

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 7-5-4-3-5 Video Test When PC Diagnostics is selected in the left pane of the Diagnostics page, the following video diagnostic tests can be performed. These tests use 18 test patterns to fill the video buffer, testing your graphics acceleration and text output •...
  • Page 358: 7-5-4-3-9 Audio Test

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 7-5-4-3-9 Audio Test The audio test option is not enabled. 7-5-4-3-10 Monitor Test When PC Diagnostics is selected in the left pane of the Diagnostics page, the following monitor...
  • Page 359: Front End Diagnostics

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 7-5-4-4 Front End Diagnostics When the Front End Diagnostics option is selected in the left pane of the Diagnostics page, the same tests as those performed by pressing <Ctrl+F11> during regular operation, can be conducted. Both...
  • Page 360 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Figure 7-82 Non-Interactive Tests (Systems with RFT Configuration) For information on performing a specific Non-interactive test, refer to the appropriate section, as follows: • - see...
  • Page 361 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Figure 7-83 Interactive Tests NOTE: Interactive tests are performed on the Front Board only. Chapter 7 - Diagnostics/Troubleshooting 7-109...
  • Page 362 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL For information on performing a specific Interactive test, refer to the appropriate section, as follows: • Transmit Test Transmit Test on page 7 - 29. •...
  • Page 363: Calibration Page

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 7-5-5 Image Quality Page When the Image Quality page is selected, as shown below, you can verify and calibrate image quality. Figure 7-84 Image Quality Page Note: This page is not populated in this version.
  • Page 364: Configuration Page

    When the Configuration page is selected, you can configure the GSUI (InSite). No changes should be made in this page without the help of GE Medical Systems Service Personnel, as any modifications made by untrained personnel can cause damage to the ultrasound unit.
  • Page 365: Utilities Page

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 7-5-8 Utilities Page When the Utilities page is selected (as shown below), a variety of Windows 2000 utility tools are available to indicate the status of the system, in addition to various other tools.
  • Page 366 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Figure 7-88 System Shutdown 7-114 Section 7-5 - Common Service Interface...
  • Page 367: Scanner Utilities

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 7-5-8-2 Scanner Utilities The functions accessed under Scanner Utilities are described below. • System Shutdown There are three options provided (described on previous page - also see...
  • Page 368 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL • Clean UserDefs used to clean all user-defined settings and restore the default settings Figure 7-90 Clear User-defined Settings CAUTION: THIS PROCESS IS IRREVERSIBLE! Before proceeding, make sure you have a complete backup of the current settings if they will be required in the future.
  • Page 369: Prodiags Configuration

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 7-5-8-3 ProDiags Configuration 7-5-8-3-1 ProDiags Browser 1) Select the Prodiags Browser under the Configuration Utility option. The following message is displayed. Wait a few minutes, until the following screen is displayed.
  • Page 370 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Figure 7-92 ProDiags Browser - Task The Task tab indicates all the tasks that are available for activation. This also indicates which tasks are scheduled, and their type.
  • Page 371 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Figure 7-93 ProDiags Browser - Time Slot The Time Slot tab indicates the time slots available for setting tasks. Chapter 7 - Diagnostics/Troubleshooting 7-119...
  • Page 372 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Figure 7-94 ProDiags Browser - Log and Result The Log & Result tab indicates the logs and results of any tasks that have been performed. At any of the above-mentioned tabs, you can Add or Delete a task from the list of displayed tasks.
  • Page 373: 7-5-8-3-2 File Transfer Configuration

    IVID ERVICE ANUAL 7-5-8-3-2 File Transfer Configuration The File Transfer Configuration utility can be used to transfer files to the GE Service center. Follow the on-screen instructions as shown below. Figure 7-95 File Transfer Configuration Chapter 7 - Diagnostics/Troubleshooting 7-121...
  • Page 374: 7-5-8-3-3 Ilinq Configuration Utility

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 7-5-8-3-3 iLinq Configuration Utility When the iLinq Configuration Utility option is selected in the left pane of the Configuration page (see Figure 7-96 below) - the InSite Interactive Platform Configuration window is displayed, as shown in Figure 7-97 on page 7-123.
  • Page 375 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Click Accept and select the modem device as the communication device to the OLC in the Device Connection tab, also shown below. Figure 7-97 InSite Interactive Platform Configuration...
  • Page 376 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Contact the OLC to receive all the setup information required for the InSite Checkout tab, as shown below. When all the settings have been entered, click the CHECKOUT NOW button twice. The OLC will attempt to communicate with your ultrasound unit via the modem.
  • Page 377 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL iLinq IP Save/Restore The iLinq IP Save/Restore utility, which is accessed from the System Configuration window (under the Technical Support Tab), will allow you to save the iLinq IP settings to the hard disk, as described below.
  • Page 378 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL iLinq IP Restore Procedure: When required (for example, after software re-installation), you can perform Restore of iLinq IP settings from the hard disk, as follows: 1) From the System Configuration window (under the Technical Support Tab - refer to...
  • Page 379: Replacement Page

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 7-5-9 Replacement Page When the Replacement page is selected, as shown below, part replacement details and ordering information is displayed, as described in Chapter 9 - Renewal Parts.
  • Page 380: Automatic Error Log

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Section 7-6 Automatic Error Log The Vivid™ 4 automatically logs the software activity, sequences and error messages that occur during normal operation. Each day’s activity is compressed into a single daily logger report, with a time stamp appearing in each line of the report.
  • Page 381: Saving The Logger Report

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 7-6-2 Saving the Logger Report The daily logger reports are retained in the system for 90 days, during which time they can be recalled. In addition, each report can be downloaded to a floppy disk, for example in order to be sent to the servicing center.
  • Page 382 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL SYSTEM MONITOR 17” FB200590 COLOR PRINTER 2269429 Speakers R/G/B/ Sync 2269309 2391918 Track Color Printer Trig. K.B Assy K/B A/N K/B Matrix From Ac Ball On/off...
  • Page 383 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL SYSTEM MONITOR 17“ P/N FB200590 COLOR PRINTER Speaker R/G/B/ Sync 2269309 2391918 Track Color Printer Trig. K.B Assy K/B A/N K/B Matrix From Ac Ball On/off...
  • Page 384: Overview

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Chapter 8 Replacement Procedures Section 8-1 Overview 8-1-1 Purpose of Chapter 8 This chapter provides replacement procedures for system parts. Table 8-1 Contents in Chapter 8...
  • Page 385 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Section 8-2 Cover Replacement Procedures 8-2-1 Overview of Covers Figure 8-1 Vivid™ 4 Ultrasound Unit Table 8-2 Vivid™ 4 Ultrasound Unit Covers (Legend to Figure 8-1)
  • Page 386: Preparations

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 8-2-1-1 Preparations Shut down the Vivid™ 4 ultrasound unit, as described in Chapter 3 - Installation. DANGER: Electrical hazards exist at several points in the system. Familiarize yourself with all hazardous voltages and high current levels before removing any of the covers.
  • Page 387: Side Covers Replacement Procedures

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 8-2-2 Side Covers Replacement Procedures 8-2-2-1 Tools Use the appropriate Phillips-type screw drivers, as indicated in the side cover replacement procedures. 8-2-2-2 Preparations Shut down the Vivid™ 4 ultrasound unit, as described in Chapter 3 - Installation.
  • Page 388: Front Cover And Air Filter Replacement Procedures

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 8-2-3 Front Cover and Air Filter Replacement Procedures 8-2-3-1 Tools Use the appropriate Phillips-type screw drivers, as indicated in the front cover and air filter replacement procedures.
  • Page 389: Front Cover Installation Procedure

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 4) Remove the three screws located across the front of the top cover, shown below: Figure 8-5 Top Cover Screws (Lower Section) 5) In order to free the front cover, grip the cover by its lower sides, and push upwards while pulling it towards you.
  • Page 390: Rear Cover Replacement Procedures

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 8-2-4 Rear Cover Replacement Procedures 8-2-4-1 Tools Use the appropriate Phillips-type screw drivers, as indicated in the rear cover replacement procedures. 8-2-4-2 Preparations Shut down the Vivid™ 4 ultrasound unit, as described in Chapter 3 - Installation.
  • Page 391: Connector Panels Cover Replacement Procedures

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 8-2-5 Connector Panels Cover Replacement Procedures 8-2-5-1 Tools Use the appropriate Phillips-type screw drivers, in indicated in the connector panels cover replacement procedures. 8-2-5-2 Preparations Shut down the Vivid™ 4 ultrasound unit, as described in Chapter 3 - Installation.
  • Page 392: Top Cover (Lower Section) Replacement Procedures

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 8-2-6 Top Cover (Lower Section) Replacement Procedures 8-2-6-1 Tools Use the appropriate Phillips-type screw drivers, and a wire cutter as indicated in the replacement procedures for the lower section of the top cover.
  • Page 393 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 7) Cut the cable tie holding the AC cable (P/N 2253013) to the other cables, and unplug the AC cable from the AC Outlet in the AC Distribution Box, as shown below:...
  • Page 394 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 10) Remove the four screws that secure each connector panel to the top cover. 11) Gently pull the rear panels backwards, leaving them hanging loosely on their cables, as shown...
  • Page 395: Top Cover Lower Section Installation Procedure

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 8-2-6-4 Top Cover Lower Section Installation Procedure 1) Return the gas spring cover section back to its original position in the top cover. 2) Align the opened gap of the gas spring cover to be in parallel with the opened gap of the top cover collar.
  • Page 396: Gas Spring Cover Replacement Procedure

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 8-2-7 Gas Spring Cover Replacement Procedure 8-2-7-1 Tools Use the appropriate Phillips-type screw drivers, as indicated in the gas spring cover replacement procedures. 8-2-7-2 Preparation Shut down the Vivid™ 4 ultrasound unit, as described in Chapter 3 - Installation.
  • Page 397: Control Console Upper Cover (Front) Replacement Procedure

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 8-2-8 Control Console Upper Cover (Front) Replacement Procedure NOTE: The Vivid™ 4 control console upper cover (new type) has two components - Front (P/N 2405642) and Rear (P/N 2405643).
  • Page 398: Control Console Upper Cover (Front) Installation Procedure

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 8-2-8-4 Control Console Upper Cover (Front) Installation Procedure 1) Raise the control console to its maximum height. 2) Return the upper cover (front section) to the Vivid™ 4 control console, making sure it is placed in the correct position (with screw holes properly aligned).
  • Page 399: Control Console Upper Cover (Rear) Replacement Procedure

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 8-2-9 Control Console Upper Cover (Rear) Replacement Procedure NOTE: The Vivid™ 4 control console upper cover (new type) has two components - Front (P/N 2405642) and Rear (P/N 2405643).
  • Page 400: Control Console Upper Cover (Rear) Installation Procedure

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 8-2-9-4 Control Console Upper Cover (Rear) Installation Procedure 1) Raise the control console to its maximum height. 2) Return the control console upper cover (rear section) to the Vivid™ 4 control console, making sure it is placed in the correct position (with the screw holes properly aligned and the aperture for the monitor correctly placed).
  • Page 401: Control Console Lower Cover Replacement Procedure

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 8-2-10 Control Console Lower Cover Replacement Procedure 8-2-10-1 Tools Use the appropriate Phillips-type screw drivers, as indicated in the bottom cover replacement procedures. 8-2-10-2 Preparation Shut down the Vivid™ 4 ultrasound unit, as described in Chapter 3 - Installation.
  • Page 402: Control Console Lower Rear Cover Replacement Procedure

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 8-2-11 Control Console Lower Rear Cover Replacement Procedure 8-2-11-1 Tools Use the appropriate Phillips-type screw drivers, as indicated in the replacement procedures of the upper section of the control console’s bottom cover.
  • Page 403: Front Handle Replacement Procedure

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 8-2-12 Front Handle Replacement Procedure 8-2-12-1 Tools Use the appropriate Phillips-type screw drivers and an 8mm wrench, to replace the front handle. 8-2-12-2 Preparation Shut down the Vivid™ 4 ultrasound unit, as described in Chapter 3 - Installation.
  • Page 404: Rear Handle Replacement Procedure

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 8-2-13 Rear Handle Replacement Procedure 8-2-13-1 Tools Use the appropriate Phillips-type screw drivers and an 8mm Allen wrench, to replace the rear handle. 8-2-13-2 Preparation Shut down the Vivid™ 4 ultrasound unit, as described in Chapter 3 - Installation.
  • Page 405: Control Console Components Replacement

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Section 8-3 Control Console Components Replacement The Vivid™ 4 Control Console comprises various components, the replacement procedures for each of which is described as follows: •...
  • Page 406: Vivid™ 4 17" Monitor Rear Cover Installation

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 5.) Unscrew and remove each of the 4 screws that fasten the Monitor Rear Cover in position (two each side; one upper and one lower), as shown in Figure 8-21.
  • Page 407: Vivid™ 4 17" Monitor Replacement Procedure

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 8-3-2 Vivid™ 4 17" Monitor Replacement Procedure CAUTION: The Vivid™ 4 17" monitor is heavy. At least two people must be available to lift the monitor. Attempts to move this unit by one person alone, could result in personal injury, and/or damage to the system.
  • Page 408: Vivid™ 4 17" Monitor Installation

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 6.) Loosen and remove the four screws on the monitor securing bracket (the metal cover was removed in step 4 to provide easy access) - refer Figure 8-23 below.
  • Page 409: Keyboard Replacement Procedure

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 8-3-3 Keyboard Replacement Procedure 8-3-3-1 Tools Use the appropriate flat and Phillips-type screw drivers, a wire cutter and the appropriate Hex wrench keys as indicated in the keyboard replacement procedure.
  • Page 410: Keyboard Installation Procedure

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 8) Lift out the keyboard, completely removing it from the machine chassis. 8-3-3-4 Keyboard Installation Procedure 1) Return the keyboard to its original position. 2) Push the keyboard as far in as it can go, carefully aligning the forked metal aligning plate with the...
  • Page 411: Keypad Replacement Procedure

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 8-3-4 Keypad Replacement Procedure 8-3-4-1 Tools Use a small flat screw driver, as indicated in the keypad replacement procedure. 8-3-4-2 Preparation Shut down the Vivid™ 4 ultrasound unit, as described in Chapter 3 - Installation.
  • Page 412: Keycaps (External Keyboard) Replacement Procedure

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 8-3-5 Keycaps (External Keyboard) Replacement Procedure NOTE: This section describes the replacement procedure for ATGC slider Keycaps on the external keyboard. ATGC Slider Keycaps are supplied in a Keycap Kit - for details of the Part Number, refer to Table 9-6 on page 9-8.
  • Page 413: Rotary Knob (External Keyboard) Replacement Procedure

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 8-3-6 Rotary Knob (External Keyboard) Replacement Procedure NOTE: This section describes the replacement procedure for a parameter adjustment Rotary Knob on the external keyboard. Rotary knobs are supplied as part of the Keycap Kit - for details of the Part Number,...
  • Page 414: Rotary Knob Installation Procedure

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 8-3-6-4 Rotary Knob Installation Procedure 1) Using a new rotary knob, insert the metal tension plate (removed in step 4 above) into the centre of the knob, aligning the plate correctly with the notches on each side of the hole. The plate, which is...
  • Page 415: Trackball Replacement Procedure

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 8-3-7 Trackball Replacement Procedure 8-3-7-1 Tools Use the appropriate flat and Phillips-type screw drivers, a wire cutter and the appropriate Hex wrench keys as indicated in the trackball replacement procedure.
  • Page 416 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 3.) Carefully return the keyboard to its original position taking care to align the forked metal aligning plate with the central nut; tighten the M5 nut.
  • Page 417: Speaker Replacement Procedure

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 8-3-8 Speaker Replacement Procedure 8-3-8-1 Tools Use the appropriate flat and Phillips-type screw drivers, a wire cutter and the appropriate Hex wrench keys as indicated in the speaker replacement procedure.
  • Page 418: Front End Parts Replacement

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Section 8-4 Front End Parts Replacement 8-4-1 Front End Boards Replacement Procedure NOTE: The procedures below provide the instructions for replacing the following boards: FB, MUX, BF, FEC, RFT, IMP, or RFI.
  • Page 419: Front End Boards Installation Procedure

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 4) Gently pull the required card from the crate. CAUTION: When removing the card, ESD may cause damage to the card. Always have the ESD hand strip connected to the machine chassis and to your hand.
  • Page 420: Tr4 Boards Replacement Procedure

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 8-4-2 TR4 Boards Replacement Procedure NOTE: The procedure below provides the instructions for replacing any of the TR4 boards ( P/N 2253036). 8-4-2-1 Tools Use the appropriate flat and Phillips-type screw drivers and an ESD wrist band strap as indicated in the TR4 Boards replacement procedure.
  • Page 421: Tr4 Boards Installation Procedure

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 4.) Turn the FE Board over, then remove the brackets that secure the TR4 boards in place as shown Figure 8-33 below. Securing Bracket Securing Bracket...
  • Page 422: Dc Power Supply Replacement Procedure

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 8-4-3 DC Power Supply Replacement Procedure 8-4-3-1 Tools Use the appropriate flat and Phillips-type screw drivers and a wire cutter, as indicated in the DC power supply replacement procedure.
  • Page 423: Tx Power Supply Replacement Procedure

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 8-4-4 TX Power Supply Replacement Procedure 8-4-4-1 Tools Use the appropriate flat and Phillips-type screw drivers, a wire cutter, and the appropriate Hex wrenches as indicated in the Tx power supply replacement procedure.
  • Page 424: Front End Crate Replacement Procedure

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 8-4-5 Front End Crate Replacement Procedure 8-4-5-1 Tools Use the appropriate flat and Phillips-type screw drivers, a wire cutter and the appropriate Hex wrenches as indicated in the front end crate replacement procedure.
  • Page 425: 2392751-100, Revision

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 10) Disconnect the two Ground cables from the crate - see Figure 8-35 on page 8-41. 11) Carefully remove the front end crate. CAUTION: To avoid damaging the Ground cables, pull the crate out very slowly.
  • Page 426: Preparation

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 8-4-6 Fan Replacement Procedure 8-4-6-1 Tools Use the appropriate flat and Phillips-type screw drivers, a wire cutter and the appropriate Hex wrenches as indicated in the fan replacement procedure.
  • Page 427: Fan Installation Procedure

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 8-4-6-4 Fan Installation Procedure 1) Wrap the spiral wrapping to hold the fan wires together. 2) Reconnect the fan wires to the block terminal using the same color coding as the fan that was removed.
  • Page 428: Back End Parts Replacement

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Section 8-5 Back End Parts Replacement Replacement of internal Back End Processor components should be done by qualified personnel only. When replacing internal BEP components, ensure replacement is with new components of the same type, and hardware version and place them in the PCI slots as shown in the appropriate figure below.
  • Page 429 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL NOTE: IMPORTANT - When replacing the BEP, the new Back End Processor will be supplied in a basic configuration with only the following components ready installed: •...
  • Page 430 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Figure 8-40 Back End Component Locations - Systems with RFT Configuration Chapter 8 - Replacement Procedures 8-47...
  • Page 431: Back End Processor Replacement Procedure

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 8-5-2 Back End Processor Replacement Procedure 8-5-2-1 Tools Use the appropriate flat and Phillips-type screw drivers and a wire cutter, as indicated in the Back End Processor replacement procedure.
  • Page 432 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Note: When replacing a BEP, one must re-install software using the software provided with the system. Do NOT use the software supplied with the BEP unless it is a version later than the software currently installed on the system.
  • Page 433: Back End Processor Installation Procedure

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL PCB: Frame Grabber (RFI systems only) - see Frame Grabber Board Removal Procedure on page 8 - 64. PCB: Keyboard Ctrl. Card - see Keyboard Control Board Removal Procedure on page 8 - 65.
  • Page 434 HASP key is connected to the parallel port. Make sure that the nominal voltage and range is set correctly in the Technical Support Tab under System Configuration. (This must be done by an authorized GE representative). Software version 4.1 or above must be re-installed - see the Software Installation/Upgrade Procedure on page 8 - 113.
  • Page 435 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 21.) If available on the system, connect the Network cable (Part No. 2292972) to the Network socket [9]. 22.) Connect the Back End Audio BKT EXT cable (Part No. 2258803) to the Audio IN/OUT socket [10b].
  • Page 436 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL The dialog box displays the Technical Support Tab as shown in Figure 8-43. RFI-configured systems RFT-configured systems Figure 8-43 System Configuration - Technical Support Tab 37.)Select the appropriate settings, as follows:...
  • Page 437: Bep2 Cover And Octopus Card Holder Replacement Procedure

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 8-5-3 BEP2 Cover and Octopus Card Holder Replacement Procedure NOTE: BEP2 refers to a Pentium 4-based BEP (newer type). 8-5-3-1 Tools Use the appropriate flat and Phillips-type screw drivers, as indicated in the BEP cover replacement procedure.
  • Page 438 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Figure 8-45 Front Cover - BEP2 Figure 8-46 Front Cover Removal - BEP2 Chapter 8 - Replacement Procedures 8-55...
  • Page 439: Bep2 - Front Cover Installation Procedure

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 8-5-3-6 BEP2 - Front Cover Installation Procedure 1) Return the front cover to the front of the BEP and press the two latches on the side, in order to push the cover back into position.
  • Page 440: Plug & Scan Board Replacement Procedure

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 8-5-4 Plug & Scan Board Replacement Procedure 8-5-4-1 Tools Use the appropriate flat and Phillips-type screw drivers and a wire cutter, as indicated in the Plug & Scan board replacement procedure.
  • Page 441: Plug & Scan Battery Replacement Procedure

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 8-5-5 Plug & Scan Battery Replacement Procedure NOTE: IMPORTANT - If the Plug & Scan battery is replaced, ensure that you properly transport and dispose of the old battery according to the procedures established by your facility, company, or regional regulatory agency.
  • Page 442: Plug & Scan Battery Installation Procedure

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 6) The battery is fastened with velcro strips to the inner wall of the BEP. Remove the battery by pulling it away from the velcro. Figure 8-49 Plug & Scan Battery Removed 8-5-5-4 Plug &...
  • Page 443: Vga Agp Board Replacement Procedure

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 8-5-6 VGA AGP Board Replacement Procedure DO NOT TOUCH ANY BOARDS WITH INTEGRATED CIRCUITS PRIOR TO CAUTION: TAKING THE NECESSARY ESD PRECAUTIONS: 1. ALWAYS CONNECT YOURSELF, VIA AN ARM-WRIST STRAP, TO THE ADVISED ESD CONNECTION POINT LOCATED ON THE REAR OF THE SCANNER (TO THE RIGHT OF THE POWER CONNECTOR).
  • Page 444: Scsi Board Replacement Procedure

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 8-5-7 SCSI Board Replacement Procedure DO NOT TOUCH ANY BOARDS WITH INTEGRATED CIRCUITS PRIOR TO CAUTION: TAKING THE NECESSARY ESD PRECAUTIONS: 1. ALWAYS CONNECT YOURSELF, VIA AN ARM-WRIST STRAP, TO THE ADVISED ESD CONNECTION POINT LOCATED ON THE REAR OF THE SCANNER (TO THE RIGHT OF THE POWER CONNECTOR).
  • Page 445 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 7) Return the Vivid™ 4 rear and left and right side covers - see the Rear Cover Installation Procedure on page 8 - 7 and the Side Covers Installation Procedure on page 8 - 4.
  • Page 446: Pc2Ip Board Replacement Procedure

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 8-5-8 PC2IP Board Replacement Procedure 8-5-8-1 Tools Use the appropriate flat and Phillips-type screw drivers and a wire cutter as indicated in the PC2IP board replacement procedure.
  • Page 447: Frame Grabber Board Replacement Procedure

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 8-5-9 Frame Grabber Board Replacement Procedure NOTE: The procedures described in this section are only applicable to Vivid™ 4 systems with the RFI configuration. 8-5-9-1 Tools Use the appropriate flat and Phillips-type screw drivers and a wire cutter as indicated in the frame grabber board replacement procedure.
  • Page 448: Keyboard Control Board Replacement Procedure

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 8-5-10 Keyboard Control Board Replacement Procedure 8-5-10-1 Tools Use the appropriate flat and Phillips-type screw drivers and a wire cutter as indicated in the keyboard control board replacement procedure.
  • Page 449: Cdrw Drive Replacement Procedure

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 8-5-11 CDRW Drive Replacement Procedure 8-5-11-1 Tools Use the appropriate flat and Phillips-type screw drivers and a wire cutter. When installing a new CDRW Drive on systems that have software versions earlier than 4.1, it is necessary to install the CD-RW patch (P/N 2413448) that is supplied as part of the CD-RW Kit, as shown below.
  • Page 450: Cdrw Drive Installation Procedure

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 8-5-11-4 CDRW Drive Installation Procedure Note: Before installing the new CDRW Drive, make sure the jumper settings closest to the data cable are shorted as follows:...
  • Page 451: Mo Drive Replacement Procedure

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 8-5-12 MO Drive Replacement Procedure 8-5-12-1 Tools Use the appropriate flat and Phillips-type screw drivers and a wire cutter. 8-5-12-2 Preparation Shut down the Vivid™ 4 ultrasound unit, as described in Chapter 3 - Installation.
  • Page 452: Mo Drive Installation Procedure

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 8-5-12-4 MO Drive Installation Procedure Note: Before installing the new MO Drive, make sure the MO Drive jumper settings are shorted as shown below: Settings required:...
  • Page 453: Ecg Module Replacement Procedure

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 8-5-13 ECG Module Replacement Procedure NOTE: In this section, the replacement procedure for an ECG Module for BEP1 and BEP2 systems are provided in separate sub-sections.
  • Page 454 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL ECG Module Figure 8-53 ECG Module - BEP2 8) Remove the four screws that secure the ECG Module in position (two from each side of the exposed...
  • Page 455: Bep2 - Ecg Module Installation Procedure

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 8-5-13-4 BEP2 - ECG Module Installation Procedure NOTE: BEP2 refers to a Pentium 4-based BEP (newer type). 1.) Slide the new ECG Module into the front of the BEP, pushing it in until it aligns with other modules at the front of BEP.
  • Page 456: Pc-Vic Replacement Procedure

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 8-5-14 PC-VIC Replacement Procedure NOTE: This procedure is performed for systems with a BEP2 (Pentium 4-based BEP, newer type). 8-5-14-1 Tools Use the appropriate flat and Phillips-type screw drivers and a wire cutter as indicated in the PC-VIC replacement procedure.
  • Page 457: Pc-Vic Installation Procedure

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Figure 8-57 PC-VIC Removal 7.) Remove the PC-VIC. 8-5-14-4 PC-VIC Installation Procedure 1) Place a new PC-VIC in the PC-VIC module holder. 2) Return and tighten the four screws (two each side) on the side of the holder - refer to Figure 8-57, above.
  • Page 458: Bep2 Power Supply Replacement Procedure

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 8-5-15 BEP2 Power Supply Replacement Procedure NOTE: This procedure is performed for systems with a BEP2 (Pentium 4-based BEP, newer type). 8-5-15-1 Tools Use the appropriate flat and Phillips-type screw drivers and a wire cutter as indicated in the BEP2 Power Supply replacement procedure.
  • Page 459 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 8.) From the rear of the BEP2, remove the four screws that fasten the power supply in position on the BEP, as shown in Figure 8-59.
  • Page 460 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 10.)Gently slide the power supply a short distance backwards (away from the BEP casing), while tilting it downwards and pulling it to the side. NOTE: For clarification, the numbers appearing in square brackets in the following steps correspond to the...
  • Page 461 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 17.)Disconnect the power cable from the Floppy Disk (labelled [6] in Figure 8-61); refer also to Figure 8-62, above. 18.)Disconnect the mains power cable (4-pin flat connector [7] - Part No. 2252981) from the rear of the BEP, cutting any cable ties as necessary.
  • Page 462: Bep2 Power Supply Installation Procedure

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 8-5-15-4 BEP2 Power Supply Installation Procedure Note: When performing this procedure, it is recommended to work with the BEP2 lying on its side (as shown in Figure 8-60 on page 8-76), to provide easier access.
  • Page 463 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 4.) Re-connect the mains power cable [B] to the motherboard (beside the ECG location). 5.) Re-connect the mains power cable (4-pin flat connector [C] - Part No. 2252981) to the rear of the BEP, and fasten with cable ties.
  • Page 464: Hard Disk Replacement Procedure

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 8-5-16 Hard Disk Replacement Procedure 8-5-16-1 Tools Use the appropriate flat and Phillips-type screw drivers and a wire cutter as indicated in the Hard Disk replacement procedure.
  • Page 465: Hard Disk Installation Procedure

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Figure 8-67 Hard Disk Removal 2 6.) Disconnect all the cables at the rear of the hard disk. 7.) Remove the hard disk. 8-5-16-4 Hard Disk Installation Procedure Note: Before installing a new hard disk, make sure the jumper settings are as shown below: 1) Place the new hard disk in position in the hard disk module holder.
  • Page 466 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 6) Install the BEP side cover, as described in the BEP2 - Side Cover Installation Procedure on page 8 - 54. 7) Return the Vivid™ 4 left cover, as described in the Side Covers Installation Procedure on page 8 - 4.
  • Page 467: Lower Section Components Replacement

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Section 8-6 Lower Section Components Replacement 8-6-1 AC Distribution Box Replacement Procedure 8-6-1-1 Tools Use the appropriate flat and Phillips-type screw drivers and a wire cutter as indicated in the AC Distribution Box replacement procedure.
  • Page 468: Ac Distribution Box Installation Procedure

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 8-6-1-4 AC Distribution Box Installation Procedure 1) Insert the AC Distribution Box through the rear of the machine and then push it into its original position.
  • Page 469: Ac Input Box Replacement Procedure

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 8-6-2 AC Input Box Replacement Procedure 8-6-2-1 Tools Use the appropriate flat and Phillips-type screw drivers and a wire cutter as indicated in the AC Input Box replacement procedure.
  • Page 470: Keyboard Or Monitor Cable Replacement Procedure

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 8-6-3 Keyboard or Monitor Cable Replacement Procedure 8-6-3-1 Tools Use the appropriate flat and Phillips-type screw drivers, a wire cutter and the appropriate Allen keys as indicated in the keyboard or monitor cable replacement procedure.
  • Page 471: Ac, Bep Or Fe Cable Replacement Procedure

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 8-6-4 AC, BEP or FE Cable Replacement Procedure Use the appropriate flat and Phillips-type screw drivers, a wire cutter, the appropriate Hex wrenches and the appropriate Allen keys as indicated in the cable replacement procedure.
  • Page 472: Gas Spring Cable Replacement Procedure

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 8-6-5 Gas Spring Cable Replacement Procedure NOTE: This section describes the installation and removal procedures for the Gas Spring Cable (Part No. 2268877) only. For instructions on replacing the Swivel Release Cable, refer to the...
  • Page 473 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Figure 8-72 Gas Spring Rod and Cable Holder - Close-up Views 10) Working from the rear of the unit, remove the securing screw [A] from the cable holder [B] located on the gas spring rod.
  • Page 474 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 13) Using a combination wrench, release the gas spring cable from the cable retaining bracket by loosening the securing nut, as shown in Figure 8-74 below..
  • Page 475: Gas Spring Cable Installation Procedure

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 8-6-5-3 Gas Spring Cable Installation Procedure 1) Raise the control console to its maximum height. 2) Route the gas spring cable along its original path in the machine. The end of the cable with the adjustable sleeve should be next to the up/down/swivel handle at the front of the unit;...
  • Page 476 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 14) Return the control console lower rear cover, as described in the Control Console Lower Rear Cover Installation Procedure on page 8 - 19. 15) Raise the control console to its maximum height.
  • Page 477: Swivel Release Cable Replacement Procedure

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 8-6-6 Swivel Release Cable Replacement Procedure NOTE: This section describes the installation and removal procedures for the Swivel Release Cable (Part No. 2392103) only. For instructions on replacing the Gas Spring Cable, refer to the...
  • Page 478 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Figure 8-77 Swivel Release Cable and Vertical Axis Welding (from Side) 8) Working from the left side of the unit, release the swivel release cable [A] from the cable retaining bracket [B] located on the vertical axis welding [C] by pressing the tensioning plate [D] downwards (to release tension on the cable).
  • Page 479 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Slide bracket in this direction Cable retaining bracket Loosen cap screws here Swivel release cable Figure 8-79 Swivel Release Cable and Retaining Bracket 10) Using a combination wrench, release the swivel release cable from the cable retaining bracket by...
  • Page 480: Swivel Release Cable Installation Procedure

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 8-6-6-3 Swivel Release Cable Installation Procedure 1) Raise the control console to its maximum height. 2) Route the swivel release cable along its original path in the machine. The end of the cable with the adjustable sleeve should be next to the up/down/swivel handle at the front of the unit;...
  • Page 481: Up/Down/Swivel Handle Replacement Procedure

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 8-6-7 Up/Down/Swivel Handle Replacement Procedure NOTE: This section describes the removal and installation procedures for the metal up/down/swivel handle (Part No. 237002). Up/Down/Swivel Handle (used for control console up/down/swivel movement) Figure 8-82 Vivid™...
  • Page 482 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 5) Loosen and remove the two M4 screws from the keyboard matrix (one on each of the left and right corners - see Figure 8-84 below). Lift out the keyboard, completely removing it from the machine chassis.
  • Page 483 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Note: For clarification, Figure 8-86 below shows a close-up view of the up/down/swivel handle, gas spring cable, and swivel release cable fitted in position, as seen from underneath .
  • Page 484 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Cable End Cable Stopper Grooved Slot on Handle (shown in detail) Remove cable end from here Figure 8-88 Detail of Cable End and Slot on Handle...
  • Page 485: Up/Down/Swivel Handle Installation Procedure

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 8-6-7-4 Up/Down/Swivel Handle Installation Procedure NOTE: This describes the procedure for installing the Up/Down/Swivel Handle (Part No. 237002). 1) Place the new Up/Down/Swivel handle into the correct position on the handle mounting brackets...
  • Page 486: Gas Spring Replacement Procedure

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 8-6-8 Gas Spring Replacement Procedure 8-6-8-1 Tools Use the appropriate flat and Phillips-type screw drivers, a wire cutter, the appropriate open-ring wrenches and a Hex M8x80mm screw as indicated in the gas spring replacement procedure.
  • Page 487 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Figure 8-92 Gas Spring Lower End Cable - Close-up View 9) Remove the retaining washer [K] from the pin [L] at the lower end of the gas spring.
  • Page 488: Gas Spring Installation Procedure

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 8-6-8-4 Gas Spring Installation Procedure NOTE: All the gas spring components indicated by letters appearing in square brackets [ ] in the following steps can be seen in...
  • Page 489: Front Wheel Replacement Procedure

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 8-6-9 Front Wheel Replacement Procedure NOTE: This section describes the installation and removal procedures for the earlier-type of front wheel (Part No. 2269437) and the newer improved type (Part No. 2392789). The installation procedure is slightly different for each type of front wheel, as described in the procedures below.
  • Page 490 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Figure 8-94 Releasing the Spring Washer from the Center of the Transverse Arm 3) Remove the four bolts (three wrench bolts and one Allen bolt) that secure the front wheel to the...
  • Page 491 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 8-6-9-4 Front Wheel Removal - Procedure 2 WARNING: When performing this procedure, the machine must be safely supported on two wooden blocks (or on a jack capable of supporting the weight of the machine). Do not lean on the machine, or leave it unattended without support, when the wheels have been removed.
  • Page 492 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Transverse Arm (Wheel Position Bar) Figure 8-97 Removing the Transverse Arm 6) Remove the four bolts (three wrench bolts and one Allen bolt - see Figure 8-98 below) that secure one of the front wheels to the chassis, and remove the wheel.
  • Page 493 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 8-6-9-5 Front Wheel Installation - Procedure 1 Note: The machine will remain stable, even after one wheel has been removed. However, for a safer working environment, it is recommended to support the unit with wooden blocks or a jack.
  • Page 494: Rear Wheel Replacement Procedure

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 8-6-10 Rear Wheel Replacement Procedure NOTE: This section describes the installation and removal procedure of the earlier-type of rear wheel (Part No. 2269436) and the newer improved type (Part No. 2389102). The installation procedure is the same for each type of rear wheel, as described in the procedure below.
  • Page 495 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 8-6-10-4 Rear Wheel Installation Procedure WARNING: Do not lean on the machine, or leave it unattended without support, if a wheel has been removed. WARNING: When BOTH rear wheels are being replaced, the machine must be safely supported on two wooden blocks (or on a jack capable of supporting the weight of the machine).
  • Page 496: Software Loading

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Section 8-7 Software Loading 8-7-1 Software Installation/Upgrade Procedure Note: It is recommended to back-up the archive and the settings as described in the Vivid™ 4 User Manual, before commencing software installation (or whenever upgrading the software).
  • Page 497: Peripherals

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Section 8-8 Peripherals 8-8-1 B/W Video Printer Replacement Procedure NOTE: This section describes the installation and removal procedure for the Sony UP-895MDW B/W Video Printer. Figure 8-100 SONY B/W Video Printer...
  • Page 498: B&W Printer Installation Procedure

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL All switches in DOWN position Figure 8-101 B/W Video Printer, DIP Switches on the Rear Panel 8-8-1-4 B&W Printer Installation Procedure 1.) Place the B/W Video Printer on the right side of the Vivid™ 4 ultrasound unit’s top surface (below the control console - as shown in Figure 8-102) and next to the VCR (if present), allowing sufficient room to access the cables at the rear of the printer.
  • Page 499 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 2.) Connect the Remote cable, the B/W Video IN cable, and the Power cable, as shown in Figure 8-103 below. Remote cable: Video IN Cable: From: Left Connectors Panel,...
  • Page 500: Setting The Switches On The Paper Tray

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 8-8-1-4-1 Setting the Switches on the Paper Tray 1.) Press the OPEN button in the front of the B/W Video Printer. 2.) Ensure that the switches on the Paper Tray are set as shown in Figure 8-104 below.
  • Page 501: Mitsubishi Vcr Replacement Procedure

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 8-8-2 Mitsubishi VCR Replacement Procedure NOTE: This section describes the installation and removal procedure of the following MITSUBISHI VCR models: • HS-MD3000E (PAL) - P/N: 066E9513 •...
  • Page 502: Mitsubishi Vcr Installation Procedure

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 8-8-2-4 Mitsubishi VCR Installation Procedure 1.) Place the VCR on the left side of the Vivid™ 4 ultrasound unit’s top surface (below the control console - as shown in...
  • Page 503 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 5.) Power ON the VCR by pushing the power switch on VCR front panel. Power ON/OFF Switch Figure 8-108 Mitsubishi VCR Front Panel - Power Switch 6.) Press Config then, from the System Configuration window, activate the VCR / ECG tab.
  • Page 504 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 8.) From the Video Mode drop-down list, choose PAL or NTSC (according to video system in your region), then press OK to apply the changes. 9.) Proceed to the VCR Setup procedure, described below.
  • Page 505: Mitsubishi Vcr Setup Procedure

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 8-8-2-4-1 Mitsubishi VCR Setup Procedure NOTE: This procedure must be performed on initial Mitsubishi VCR installation. NOTE: The Mitsubishi VCR can be configured using the Menus on the front panel...
  • Page 506 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Figure 8-112 Menu 1 Settings 4.) Select the <Menu2> settings, according to the configuration shown in Figure 8-113 below , then press the Enter button to continue to the next menu.
  • Page 507 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Figure 8-115 RS-232C Menu Settings 7.) Skip the <Elapsed Time> (Figure 8-116) by pressing the Enter button. Figure 8-116 Elapsed Time Menu 8.) Connect the RS-232C cable and the Audio IN and Audio OUT cables between the appropriate...
  • Page 508: Mitsubishi Vcr Removal Procedure

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 8-8-2-5 Mitsubishi VCR Removal Procedure 1.) Remove the Connectors Panels Cover as described in the Connector Panels Cover Removal Procedure on page 8 - 8. 2.) If optional shelf (P/N H45011CN) is present, refer to the...
  • Page 509: Sony Vcr Replacement Procedure

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 8-8-3 Sony VCR Replacement Procedure NOTE: This section describes the installation and removal procedure of the SONY SVO-9500MDP2 and SVO-9500MD2. The SVO-9500MDP2 (PAL) and SVO-9500MD2 (NTSC) are alike, except for the Color Signal system.
  • Page 510: Sony Vcr Installation Procedure

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 6.) Verify that the switches on the VCR rear panel are as shown in Figure 8-120 below. If necessary, adjust accordingly. 1 2 3 4 5 6...
  • Page 511 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 2.) Connect the RS-232C cable, the Audio and S-Video IN and OUT cables, and the Power cord, as shown in Figure 8-122 below. S-VIDEO OUT cable...
  • Page 512: Sony Vcr Removal Procedure

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 6.) Re-install the Connectors Panels Cover as described in the Connector Panels Cover Installation Procedure on page 8 - 8. 7) Reconnect the mains AC power cable from the wall outlet.
  • Page 513: Panasonic Vcr Replacement Procedure

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 8-8-4 Panasonic VCR Replacement Procedure NOTE: This section describes the installation and removal procedure of the Panasonic AG-MD-835E (PAL) and AG-MD-835P (NTSC) video cassette recorders. The installation and removal procedures are the same for both models.
  • Page 514: Panasonic Vcr Installation Procedure

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Pull Front Switch Lock Figure 8-124 Panasonic VCR Switches Setting 8-8-4-4 Panasonic VCR Installation Procedure 1.) Raise the control console to maximum height. 2.) Place the VCR on the Vivid™ 4 ultrasound unit’s top surface (left side, below the control console, as...
  • Page 515 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 3.) Connect the Panasonic VCR RS-323 adaptor (Part No. 066E8200) to the RS-232C cable, and then to the RS-323 socket on the rear of the VCR, as shown in Figure 8-126 below.
  • Page 516 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 5.) Place the VCR in its correct final position. 6.) Secure the VCR by fastening the belt around it, or together with the B/W Printer (if present).
  • Page 517: Panasonic Vcr Removal Procedure

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 8-8-4-5 Panasonic VCR Removal Procedure 1.) Remove the Connectors Panels Cover as described in the Connector Panels Cover Removal Procedure on page 8 - 8. 2.) Release the B/W printer by undoing the fastening belt around it (together with the VCR, if present).
  • Page 518: Sony Up-21Md Color Video Printer Replacement Procedure

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 8-8-5 Sony UP-21MD Color Video Printer Replacement Procedure NOTE: This section describes the installation and removal procedures of the Sony UP-21MD Color Video Printer. The Sony UP-21MD Color Video Printer has an NTSC/PAL Selector on the rear panel.
  • Page 519 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 5.) Pull out the Color Printer cables, found in the side compartment as shown in Figure 8-129 below. Figure 8-129 Color Printer Cables in Side Compartment...
  • Page 520 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 7) Connect the RGB and SYNC cables, Color Printer Remote cable, and AC power cable (hanging out of the Bottom Metal Cover slot - see Figure 8-129...
  • Page 521: Sony Up-21Md Color Video Printer Removal Procedure

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 8) Slide the UP-21MD Color Video Printer into the side compartment, as shown in Figure 8-132 below. UP-21MD Color Video Printer Figure 8-132 Sony UP-21MD Color Video Printer Installed on the Vivid™ 4...
  • Page 522: Tools

    GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 8-8-6 HP 6540/3 USB Deskjet Color Printer Replacement Procedure for Vivid™ 4 BT03 Systems NOTE: IMPORTANT - Whether the Vivid™ 4 system is currently using 4.3 or 4.6 software, it is necessary to...
  • Page 523 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 8-8-6-4 HP 6540/3 USB Deskjet Color Printer Installation Procedure 1.) Place the HP6540/3 USB Deskjet Color Printer on the Vivid™ 4 bottom shelf allowing sufficient room to access the cables at the rear of the printer, or on a stable surface located not far from the Vivid™...
  • Page 524 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 6.) When the Found New Hardware Wizard opens, click Cancel, see Figure 8-135 below. Figure 8-135 Found New Hardware Wizard 7.) Insert the HP6540/3 driver installation CD (P/N:2415381) provided with the printer into the CD-ROM drive.
  • Page 525 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Figure 8-137 Printer Installation Process Commences 9.) When the Add Printer Wizard opens, click Next - see Figure 8-138. Figure 8-138 Add Printer Wizard 10.)When prompted, verify that Local Printer is selected. If not, activate the Local printer radio button...
  • Page 526 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Figure 8-139 Verify Local Printer 11.)Click Next. 12.)From the displayed list of available ports, choose one of the available USB ports, as shown in Figure 8-140.
  • Page 527 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 14.)From the list of Printers, highlight the HP Deskjet 6500 Series - see Figure 8-141. Figure 8-141 Printer Manufacturer and Model 15.)Click Next. 16.)When prompted, choose Yes if you want this printer to be the default printer - see Figure 8-142.
  • Page 528 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 18.)When the following message appears, verify that Do not share this printer is selected, as shown Figure 8-143. Figure 8-143 Printer Sharing Message 19.)Click Next. 20.)In response to the Print a test page? message, choose No and then click Next - see Figure 8-144.
  • Page 529 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL The Add Printer Wizard displays the following message: Figure 8-145 New Printer Added 21.)Click Finish to close the Wizard. The Printer Properties Dialog opens with the General Tab activated, as shown in Figure 8-146.
  • Page 530 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL The following message is displayed: Figure 8-147 Test Page Printing 23.)Confirm that test page was printed properly, then click OK. The Printer Properties Dialog opens with the Ports Tab activated, as shown in Figure 8-148.
  • Page 531 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL The Printer Properties Dialog opens with the Advanced Tab activated, as shown in Figure 8-149. Figure 8-149 Printer Properties Dialog Box - Advanced Tab 25.)Click on Printing Defaults.
  • Page 532 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Figure 8-150 Printing Defaults Dialog Box 26.)Set up the printing defaults (Paper size, Paper type etc..) to your liking, then click OK. The following message is displayed: Figure 8-151 Printer Installed and Checked Message 27.)Press any key to reboot the system and then eject the CD from the CD tray.
  • Page 533 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Figure 8-152 System Configuration Window - Printers Tab 29.)Confirm that HP Deskjet 6500 Series has been added to Currently Installed Printers list, then click OK. 8-8-6-5 HP 6540/3 USB Deskjet Color Printer Removal Procedure 1) Turn OFF power to the USB printer.
  • Page 534 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Chapter 9 Renewal Parts Section 9-1 Overview 9-1-1 Purpose of Chapter 9 This chapter gives you an overview of replacement parts for the Vivid™ 4 ultrasound unit.
  • Page 535 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Section 9-2 List of Abbreviations • Assy - Assembly • - Back End Processor • Ctrl - Control • - Front End Processor • FRU 1 - Replacement part available in parts hub •...
  • Page 536 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Section 9-3 Renewal Parts Lists and Diagrams 9-3-1 Mechanical Hardware Parts Figure 9-1 Mechanical Hardware Parts - Diagram 1 Table 9-2 Mechanical Hardware Parts - Diagram 1...
  • Page 537 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Figure 9-2 Mechanical Hardware Parts - Diagram 2 Section 9-3 - Renewal Parts Lists and Diagrams...
  • Page 538 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Table 9-3 Mechanical Hardware Parts - Diagram 2 Item Part Name Part Number Description Cover Right 2258600 Cover Top (Lower Section) 2258602 Control Console Upper Cover (Front) V4...
  • Page 539 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Figure 9-3 Mechanical Hardware Parts - Diagram 3 Table 9-4 Mechanical Hardware Parts - Diagram 3 Item Part Name Part Number Description Filter (Air) 2301690 Cover Gas Spring/Sleeve V4...
  • Page 540 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Figure 9-4 Mechanical Hardware Parts - Diagram 4 Table 9-5 Mechanical Hardware Parts - Diagram 4 Item Part Name Part Number Description Wheel (Castor) - Rear...
  • Page 541 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Table 9-6 Mechanical Hardware Parts - Not Illustrated Item Part Name Part Number Description Swivel Release Cable 2392103 Monitor Service Cover 2392137 Screw Kit 2253052 Keycaps Kit V4...
  • Page 542 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 9-3-2 AC System Parts Figure 9-5 AC System Parts Chapter 9 - Renewal Parts...
  • Page 543 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Table 9-7 AC System Parts Item Part Name Part Number Description Input AC Box and Circuit Breaker 2252994 AC Voltage Distribution Box 2399304-2 Connector AC Voltage Primary 100-120V...
  • Page 544 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 9-3-3 Front End Parts Figure 9-6 Front End Parts - Diagram 1 Table 9-8 Front End Parts - Diagram 1 Item Part Name Part Number Description...
  • Page 545 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 1002 1001 1004 1003 1005 Figure 9-7 Front End Parts - Diagram 2 Table 9-9 Front End Parts - Diagram 2 Item Part Name Part Number...
  • Page 546 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 1011 1012 1013 1014 Figure 9-8 Front End Parts - Diagram 3 Table 9-10 Front End Parts - Diagram 3 Item Part Name Part Number Description...
  • Page 547 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 9-3-4 Back End Parts 1022 1021 1023 Figure 9-9 Back End Parts - Diagram 1 Table 9-11 Back End Parts - Diagram 1 Item Part Name...
  • Page 548 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 1032 1031 1034 1033 1035 Part 2269309 Speakers (pair) Part EP200132 Isolation Box (network) Figure 9-10 Back End Parts - Diagram 2 Table 9-12 Back End Parts - Diagram 2...
  • Page 549 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 1051 1053 1052 1054 1055 1057 Figure 9-11 Back End Parts - Diagram 3 Table 9-13 Back End Parts - Diagram 3 Item Part Name Part Number...
  • Page 550 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 1062 1063 1061 1065 1064 1066 1067 1069 1068 Figure 9-12 Back End Parts - Diagram 4 Chapter 9 - Renewal Parts 9-17...
  • Page 551 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Table 9-14 Back End Parts - Diagram 4 Item Part Name Part Number Description 1061 PC Video Encoder Control Assembly 2340279 (PC-VIC) 1062 Hard Disk Drive (HDD) 40 to 100 GB...
  • Page 552 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 9-3-5 Cables Table 9-15 Cables Item Part Name Part Number Description 1101 Cable: VGA Out 2252964 1102 Cable: VGA VIC/ PC 2252965 1103 Cable: Input Box / AC Distribution Box...
  • Page 553 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Table 9-15 Cables (Continued) Item Part Name Part Number Description Cable: 1112 2252981 Keyboard/ VIC Internal Power/ Back-End Cable: Back End / Keyboard Matrix 1113 2252977...
  • Page 554 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Table 9-15 Cables (Continued) Item Part Name Part Number Description 1122 Cable: AC Dist. / Rear Panel Right 2253014 AC DIS. BOX REAR PANEL RIGHT J7/8/...
  • Page 555 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Table 9-15 Cables (Continued) Item Part Name Part Number Description 1133 Cable: Speakers / Back End 2258804 Cable: 1134 ON/OFF Motherboard / Keyboard 2258805 Controller 1135...
  • Page 556 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Table 9-15 Cables (Continued) Item Part Name Part Number Description 1146 Cable: Power Front End 2276875 1147 Cable: Power Back End 2276876 Cable: 1148 Network Rear Panel Right...
  • Page 557 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 9-3-6 Software Table 9-16 Software Item Part Name Part Number Description S/W Kit - contains s/w for systems: 1301 2415386 Vivid™ 3, Vivid™ 4, and Vivid-i™...
  • Page 558 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 9-3-7 Probes Table 9-17 Probes Item Part Name Part Number Description 1401 Heart Sound Microphone H45001NE 1402 Probe: C358 H40212LC Note: Not applicable for use in Japan...
  • Page 559 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Section 9-4 Peripherals Table 9-18 Approved Peripherals Item Part Name Part Number Description VCR Mitsubishi HS-MD3000E S-VHS 1501 H45011PT RS232C \PAL VCR Mitsubishi HS-MD3000U S-VHS 1502...
  • Page 560 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Section 9-5 System Block Diagrams SYSTEM MONITOR 17” FB200590 COLOR PRINTER 2269429 Speakers R/G/B/ Sync 2269309 2391918 Track Color Printer Trig. K.B Assy K/B A/N K/B Matrix...
  • Page 561 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL SYSTEM MONITOR 17“ P/N FB200590 COLOR PRINTER Speaker R/G/B/ Sync 2269309 2391918 Track Color Printer Trig. K.B Assy K/B A/N K/B Matrix From Ac Ball 2253043...
  • Page 562 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Chapter 10 Periodic Maintenance Section 10-1 Overview 10-1-1 Purpose of Chapter 10 This chapter describes Periodic Maintenance (PM) procedures for the Vivid™ 4 ultrasound scanner and its peripherals.
  • Page 563 The results of Quality Control testing, corrective action, and the effects of corrective action, must be documented and maintained on site. Your GE Medical Systems Service Representative can help you with establishing, performing and maintaining records for a Quality Assurance program.
  • Page 564 Vivid™ 4 ultrasound scanner as recommended in the Periodic Maintenance Schedule, in order to retain the high levels of safety, dependability, and system performance. Your GE Medical Systems Service Representative has an in-depth knowledge of your Vivid™ 4 ultrasound scanning system and can best provide competent, efficient service. Please contact us for further information, and to schedule routine maintenance servicing by our qualified professional FSEs.
  • Page 565 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Table 10-2 Periodic Maintenance Schedule Service at Indicated Time Daily Weekly Monthly Annually Notes •* Clean Probes * or before each use • Clean Probe Holders •...
  • Page 566 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Examine the following on a monthly basis: • Connectors on cables - for any mechanical defects. • The entire length of electrical and power cables - for cuts or abrasions.
  • Page 567 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Section 10-4 Tools Required 10-4-1 Special Tools, Supplies and Equipment 10-4-1-1 Specific Requirements for Periodic Maintenance Table 10-3 Overview of Requirements for Periodic Maintenance Tool Part Number...
  • Page 568 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Section 10-5 System Periodic Maintenance 10-5-1 Preliminary Checks The preliminary checks take approximately 15 minutes to perform. Refer to the Vivid™ 4 User Manual whenever necessary.
  • Page 569 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 10-5-2 Functional Checks The functional checks take approximately 60 minutes to perform. Refer to the Vivid™ 4 User Manual whenever necessary. NOTE: Refer to Chapter 4 - Functional Checks, for additional details about the functional checks described in this section.
  • Page 570 If any peripherals or options are not part of the system configuration, the following checks can be omitted. Refer to the Vivid™ 4 User Manual for a list of approved peripherals/options. Table 10-6 GE Approved Peripheral/Hardware Option Functional Checks Step Item Description Verify record/playback capabilities of the VCR.
  • Page 571 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 10-5-4 Cleaning 10-5-4-1 General Cleaning Table 10-8 General Cleaning Step Item Description External On a weekly basis, moisten a soft, non-abrasive folded cloth with a mild, general purpose, Panels non-abrasive soap and water solution or general purpose disinfectant.
  • Page 572 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Table 10-9 Air Filter Cleaning Step Item Description Remove the air filter located at the front of the scanner, above the probe connector sockets. Air Filter Clean the air filter as follows: shake it in an area away from the system, wash with a mild soapy solution, rinse, and air dry.
  • Page 573 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 10-5-5 Physical Inspection Table 10-10 Physical Checks Step Item Description Labeling Verify that all system labeling is present and in readable condition. Scratches and Dents Inspect the console for dents, scratches or cracks. Use touch-up paint, as required.
  • Page 574 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 10-5-6 Diagnostic Checks (Optional) To complete the periodic maintenance checks, perform the diagnostic tests as described in Chapter 7 - Diagnostics/Troubleshooting. • Review the system error log for any problems.
  • Page 575 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Section 10-6 Probe Maintenance WARNING: Any evidence of wear on a probe indicates that it must not be used. Improper handling may easily damage ultrasound probes. See the Vivid™ 4 User Manual and also refer to the probe manufacturer’s handling instructions, for more details.
  • Page 576 Ensure that you are completely familiar with the proper care of GE Medical Systems probes. WARNING: Any evidence of wear on a probe indicates that it must not be used. Ultrasound probes can be easily damaged by improper handling.
  • Page 577 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Do not soak or wipe the probe lens with any product listed above (see Warning). Doing so could result in irreparable damage to the probe. Follow the care instructions supplied with each probe. In order to disinfect or sterilize probes without damaging them: •...
  • Page 578 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 10-6-4-2-2 Cleaning and Sterilizing Probes Used Internally Table 10-13 Cleaning & Sterilizing Procedure for Probes (TEE) Used Internally Step Description To protect yourself from blood-borne diseases, wear approved, non-allergic disposable gloves.
  • Page 579 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Section 10-7 Electrical Safety Tests 10-7-1 Safety Test Overview The electrical safety tests in this section are based on and conform to NFPA 99 (For USA) and IEC 60601-1 Medical Equipment Safety Standards.
  • Page 580 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 10-7-2 GEMS Current Leakage Limits The following limits are summarized for NFPA 99 (for USA) and IEC 60601-1 Medical Equipment Safety Standards. These limits are GEMS standards and in some cases are lower than the above standards listed.
  • Page 581 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 10-7-3 Outlet Test Wiring Arrangement - USA & Canada Test all outlets in the area for proper grounding and wiring arrangement by plugging in the neon outlet tester and noting the combination of lights that are illuminated.
  • Page 582 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 10-7-4-1 Ground Continuity Testing - Generic Procedure Follow these steps to test the Ground wire resistance: 1.) Turn the Vivid™ 4 unit OFF. 2.) Connect one of the Ohmmeter probes to the tested AC wall outlet (Ground pin).
  • Page 583 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 10-7-5 Chassis Current Leakage Test 10-7-5-1 Definition This test measures the current that would flow in a grounded person who touched accessible metal parts of the bedside station, if the Ground wire should break. The test verifies the isolation of the power line from the chassis.
  • Page 584 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 10-7-5-3 Data Sheet for Chassis Source Current Leakage The Chassis Source Current Leakage Test passes when all readings measure less than the values shown Table 10-14. Record all data on the Vivid™...
  • Page 585 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 10-7-6 Isolated Patient Lead (Source) Leakage – Lead-to-Ground 10-7-6-1 Definition This test measures the current which would flow to Ground from any of the isolated ECG leads. The meter simulates a patient who is connected to the monitoring equipment and is grounded by touching some other grounded surface.
  • Page 586 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 10-7-7 Isolated Patient Lead (Source) Leakage – Lead-to-Lead NOTE: Refer to the procedure in the IEC 60601-1. 10-7-8 Isolated Patient Lead (Sink) Leakage - Isolation Test NOTE: Refer to the procedure in the IEC 60601-1.
  • Page 587 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 10-7-8-1 Data Sheet for ECG Current Leakage The test passes when all readings measure less than the values shown in the tables below. Record all data on the Vivid™...
  • Page 588 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 10-7-9 Probe Current Leakage Test 10-7-9-1 Definition This test measures the current that would flow to Ground (from any of the probes) through a patient who is being scanned, and who becomes grounded by touching some other grounded surface.
  • Page 589 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 10-7-9-3 No Meter Probe Adapter Procedure Figure 10-6 Check Without Probe Adapter Follow these steps to test each transducer for leakage current. 1.) Turn the Vivid 3 unit OFF.
  • Page 590 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL 10-7-9-4 Data Sheet for Transducer Source Current Leakage The test passes when all readings measure less than the values shown in Table 10-15 Table 10-16. Record all data on the Vivid™...
  • Page 591 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Section 10-8 Excessive Current Leakage 10-8-1 Possible Causes of Excessive Current Leakage 10-8-1-1 Chassis Fails Check the Ground on the power cord and plug for continuity. Ensure the Ground is not broken, frayed, or intermittent.
  • Page 592 M-Mode Function Probe Holders Applicable Software Options External I/O Applicable Hardware Options Wheels, Brakes & Swivel Locks Control Console Cables and Connectors Monitor GE-approved Peripherals (VCR, CD-RW, MOD, Printers) Measurement Accuracy GE-approved Peripherals Comments: Chapter 10 - Periodic Maintenance 10-31...
  • Page 593 GE M EDICAL YSTEMS 2392751-100, R ™ 4 S IRECTION EVISION IVID ERVICE ANUAL Vivid™ 4 Electrical Safety Inspection Certificate Max Value Value Electrical Test Performed Comments Allowed Measured Outlet (correct Ground & wiring config.) System Ground Continuity Chassis Source Current Leakage - Probe...
  • Page 594 GE MEDICAL SYSTEMS GE Medical Systems-Americas: Fax 414.544.3384 P.O. Box 414; Milwaukee, Wisconsin 53201-0414, U.S.A. GE Medical Systems-Europe: Fax 33.1.40.93.33.33 Paris, France GE Medical Systems-Asia: Fax 65.291.7006 Singapore GE Vingmed Ultrasound: Fax: +47 3302 1350 Horten, Norway...