GE M
S
EDICAL
YSTEMS
D
2392751-100, R
IRECTION
7-3-9-6
Advanced Board Tests
The advanced board tests check two additional parts of the MUX Board: TEE probe control and
Monitoring, as follows:
-
-
7-3-9-6-1 TEE Probe Control
The TEE probe control test checks the position setting of the probe, as well as its normal operation.
1) Access the MUX tab of the FB Diagnostics dialog box, as described in the
Diagnostic Test Options
2) Trackball to the Advanced board tests button and press Select. The Advanced System Tests
options are displayed, as shown in
3) Select the radio button next to the appropriate socket.
4) Trackball to the TEE Probe Control button and press Select. The TEE Probe Test dialog box is
displayed, as shown in
3
EVISION
TEE probe control enables you to check different positions of the probe; by changing the
probe angle, you are able to verify its normal operation.
Monitoring enabled you to check the HVPS and LVPS voltages, and the AC Power Input, as
well as temperatures.
section, on page 7-48.
Figure 7-34 MUX Advanced System Tests Options
Figure
Chapter 7 - Diagnostics/Troubleshooting
Figure 7-34
on page 7-59.
7-35, with all the probe settings.
V
™ 4 S
M
IVID
ERVICE
ANUAL
Accessing the MUX
7-59