4 6 9 0 S E R I E S | T U R B I D I T Y S E N S O R | O I/AT S 4 1 0 - E N R E V. A

Setup procedure for the de-bubbler

For systems that include a de-bubbler, refer to Figure 6:
1 Open the 'sample in' isolating valve
2 Adjust the sample flow through the turbidity system with the flow regulating valve
Note
It is recommended to use the sample regulating valves together with a flow indicator to ensure easy maintenance and consistent
performance. These devices are not supplied with the 4690 Turbidity systems.
Dimensions in mm (in)
Drain outlet
500 (19.7) minimum
Tundish
– see Notice 2
A
Sample in
Figure 6 Typical system installation for 4690 series turbidity systems
NOTICE
1 To prevent degassing of the sample, which can cause very erratic readings, do not exceed this measurement.
2 This is the minimum installation distance that ensures adequate flow rate through the sensor. Increase this distance if you
use long tubing or small-bore tubing.
A
to keep the overflow from the de-bubbler at a minimum.
De-bubbler
150 (5.9) maximum
– see Notice 1
B
B
.
Sample outlet
connector
4690 series
Tundish
Inlet/isolating
valve
Drain valve
Optional
150 (5.9) maximum
– see Notice 1
9