Replacing the horizontal lift brake
1
Fig. 6
1
Fig. 8
Remove the cable clamp (1/Fig.6).
Remove the cable clamp (2/Fig.6) and remove the cable ties (2/Fig.6).
Remove the cable ties around the cable bundle (5/Fig.5).
Remove the cover plates (1/Fig.7).
Remove the nut (1/Fig.8). (30mm socket)
Remove the three Allen screws (2/Fig.8) and remove the angulation brake completely.
When doing this, make sure to hold onto the 3 screws of the angulation brake or use
adhesive tape to prevent them from falling. Fig. 9 shows the complete angulation brake.