Fig. 14
Switch ON the SIREMOBIL Iso-C and release the horizontal brake (the LED in the
switch lights).
Move the horizontal carriage so that the opening in the horizontal carriage,
(1/Fig.14) as shown in Fig.14 is located opposite the opening in the bearing support (2/
Fig.14).
Insert the cable bundle (3/Fig.14) between the horizontal carriage and the bearing
support in the direction of the C-arm. When doing this, be careful not to damage or crimp
the cables!
NOTICE
Subsequently move the horizontal carriage carefully forward out of the bearing support
until the horizontal lift brake (1/Fig.15)with the plate (2/Fig.15) is accessible.
Note the following:
- Be careful not to crimp or damage the cables.
- The C-arm must remain centered on the work table. Move the work table, if necessary.
- Secure the cover with the switches to prevent it from falling.
- When removing the horizontal carriage, keep it centered to the ball bearings. Do not tilt
it.
- You may have to adjust the position of the vertical lifting column until the horizontal
carriage moves easily.
Switch OFF the SIREMOBIL Iso-C.
Disconnect the plug connections for the horizontal brake.
Remove the plate (2/Fig.15).
Remove the horizontal brake cable from the plate and lift the horizontal brake out of the
bearing support.
SIREMOBIL Iso-C
Replacing the horizontal lift brake
3
Two persons are required to remove the horizontal carriage. One person
should be at the bearing surface of the C-arm. The second person at the
horizontal carriage.
When removing the horizontal carriage, make sure that none of the
cables is crimped or damaged.
SPR2-230.076.01
Rev. 01
2
Fig. 15