Download Print this page

Specifications - HP 16442A User Manual

Test fixture

Advertisement

Specifications

The "supplemental information" and "typical" entries, in the following specifications are
not warranted, but provide useful information about the functions and performance of the
instruments.
The following specifications data is specified at 23 ±5
Functions
HP 16442A lets you test various shaped DUTs. HP 16442A has:
6 SMU channels (when using non-Kelvin connections).
3 SMU channels (when using Kelvin connections).
2 VSU channels.
2 VMU channels.
2 PGU channels.
1 GNDU channel.
interlock (Intlk) 6-pin connector.
Voltage and current range
Channel Maximum Voltage Maximum Current
SMU
200 V
VSU
40 V
VMU
40 V
PGU
40 V
GNDU
40 V
C and 50 % relative humidity.
1 A
100 mA
200 mA (AC peak)
1.6 A
HP 16442A User's Gude 21