Chapter 2: Installing the IntelliVue Smart-hopping 1.4 GHz Core Access Point

ITS4843B and ITS4843C Installation Procedure

To install the IntelliVue Smart-hopping 1.4 GHz Core Access Point:
1. Mount the IntelliVue Smart-hopping 1.4 GHz Core Access Point and Remote
Antennas within the clinic where they can communicate with 1.4 GHz IntelliVue
Patient Monitors. Note these guidelines when installing the IntelliVue Smart-
hopping Core Access Point and Remote Antennas:
• You may mount each IntelliVue Smart-hopping Core Access Point as follows:
• You may mount each IntelliVue Smart-hopping Remote Antenna as follows:
• Use category 5e (or better) Unshielded Twisted Pair cable to connect each
• The total length of UTP cable from the IntelliVue Smart-hopping access
• Use only the supplied, unmodified 74 ft. (22.6m) Coax and UTP cable
2. Attach each supplied local antenna to the Core AP chassis by inserting the
antenna into the socket connector. Orient the antenna (either straight for wall
mounting or rotate downward for ceiling tile mounting) and then screw on its
collar until the antenna is secure to the chassis.
3. Attach each supplied antenna to the Remote Antenna chassis by inserting the
antenna into the socket connector. Orient the antenna (either straight for wall
mounting or rotate downward for ceiling tile mounting) and then screw on its
collar until the antenna is secure to the chassis.
4. If you have installed Remote Antennas, connect each IntelliVue Smart-hopping
Core Access Point to its installed Remote Antenna(s) using the supplied,
unmodified 74 ft. (22.6m) Coax and UTP cable bundle(s).
2-6
-
to a wall (fixed mount) as described on page 2-9
-
to a wall (CA earthquake rated) as described on page 2-10
-
above a ceiling as described on page 2-12
-
below a ceiling (fixed mount) as described on page 2-15
-
below a ceiling (quick release) as described on page 2-19
-
to a wall (fixed mount) as described on page 2-24
-
to a wall (ca earthquake rated) as described on page 2-26
-
above a ceiling as described on page 2-28
-
above a ceiling (tether mount) as described on page 2-31
-
below a ceiling (fixed mount) as described on page 2-33
-
below a ceiling (quick release) as described on page 2-37
IntelliVue Smart-hopping Core Access Point to the IntelliVue Smart-hopping
infrastructure.
point-to the IntelliVue Smart-hopping Synchronization Unit-to the IntelliVue
Smart-hopping Power over Ethernet Unit-to ICN Network Switch cannot
exceed 328 ft. (100 m).