INSTALLATION INSTRUCTIONS
3
.
Make sure that the line fuse is compatible with the voltage selected.
less than
2141AOO200
2141 AOO200
and greater
Using the wrong fuse value or fuse type will not protect the
circuitry inside the 3314A and may result in damage to your
3314A.
4. Connect the power cord. Please contact your local -hp- Sales Office if you
have any questions.
5. Set the LINE switch to the ON position. This switch is locat ed in the upper left
hand corner of the front panel. The following actions will occur:
-a 2 second count down to allow electrical stabilization and test the front
panel LEOs
-a CAL ALL to generate a full complement of calibration constants
-if the calibration failed, the appropriate error number will be displayed for
1/2 second and the 3314A will try to calibrate itself for another
seconds or until a calibration is successful
6.
Connect the 3314A to an oscilloscope as shown in Figure 1 and verify that
the 3314A's output is a 1kHz, 10mVp-p
to cycle the 3314A's function to squarewave, triangle, off and back to
sinewave. The FUNCTION key is located in the lower right hand corner of the
front panel.
More detailed installation procedures are located in Section
Manual.
Table 1. Une Fuse
v __
100/120V
O.SA SLO BLOW
220/240V
OAA SLO BLOW
100f120V
1.0A
SLO BLOW
220/240V
0
.5A
SLO BLOW
sinewave.
4
(cant)
Press the FUNCTION key