The resolution (minimum end voltage step) is:
Resolution
The voltage change can range from 0% (V HGT
peak to peak amplitude setting in .04% steps, when V LEN
It is best to keep V LEN as small as possible. This reduces the number of discrete steps
the 3314A must make to record the waveform and improves resolution. Also, if total
V LEN for a waveform is very great, the waveform may drift appreciably.
Vector Length. The length of a vector in time is determined by d
vector) and V LEN (variable from 1 to 127 for each vector). To determine each vector's
time, multiply its V LEN by d t.
The marker indicates the vector available for V HGT and V
sified by the Z Axis output.
Pressing V HGT or V
for quick reviews or edits of an ARB wave.
Effects of changing V lEN