Loading the operating software
Control board D1, Host
•
Connect the service PC to the serial service interface on SIREMOBIL Compact.
•
Start the service program.
•
Insert the "Parameter" diskette in drive A.
•
Select the Download menu and start the download (ARC / Fil. / Host).
•
After the download has successfully been completed (1 minute), wait for the system to
boot.
•
The SIREMOBIL Compact is operational.
NOTE
Siemens AG
Medical Engineering
If you cannot run download (e.g. power failure during download):
- Switch the SIREMOBIL Compact off and back on
- Restart download
Register 5
RXR2-130.061.02
Rev. 03
Page 1 of 2
10.03
CS PS 24
4 - 1
4
4
SIREMOBIL Compact
System Manual