Adjustment

WARNING
Select the I.I. survey format.
Center the dose measurement chamber on the I.I. input screen.
Select "Dose rate adjustment" on the service PC.
Click on "Update Start".
The following operating modes will be automatically selected:
- kV/mA curve S2 (Antiisowatt 5mA)
- Manual kV fine adjustment (to increase the accuracy of adjustment)
Release fluoro.
Activate dose rate control.
Switch fluoro off.
Select KV stop.
Install the dose measurement chamber on the I.I. input.
Release fluoro.
Using the KV+ and KV- keys, set the dose rate to 330 nGy/s +/- 5% (Dose rate* grid
factor). This should produce a kV value of from 70 kV to 80 kV. If necessary, switch off
fluoro and change the prefiltration. After changing the prefiltration, set the dose rate
again.
Switch fluoro off.
Remove the dose measurement chamber from the I.I. input.
Switch fluoro on.
Press the >0< key on the operating console of the SIREMOBIL Compact. This starts the
automatic adjustment of the TV-Iris collimator.
Wait until the message "done" is displayed in the "AUTO REGULATION" line (Service
PC). The new TV-Iris position values will be stored in the Host RAM.
The "BRIGHTNESS DEVIATION" value will be displayed in the "BRIGHTNESS
DEVIATION" line and should not deviate by more than +/- 3.
Switch fluoro off.
Click on "Update Stop" on the service PC and wait until the "CAL" message disappears
on the 7-segment display to the right. This stores the TV-Iris position value in the Host
EEPPROM.
Close the "Monitoring" window on the service PC by clicking on the "cancel" button.
SIREMOBIL Compact
System Manual
X-ray radiation!
See Chapter 1, Safety Information.
Protect yourself against radiation exposure. Wear a lead apron.
Register 5
RXR2-130.061.02
Rev. 03
Adjustments / Programs
Page 6 of 22
10.03
CS PS 24
5
Siemens AG
Medical Engineering

Advertisement

Table of Contents
loading

Table of Contents