Replacing boards / replacing components

Checks and adjustments

Check the I.I. electrode voltages according to the test protocol for the I.I.
- Do not readjust in cases where there are only slight deviations from the values on the
test protocol (measurement device tolerances).
Check that the camera optics are centered to the I.I. output and adjust them if necessary;
refer to "Replacing the camera optics ", subitem "Checking the centering of the camera
optics to the I.I. output " and "Centering the camera optics to the I.I. output".
Check the camera rotation and readjust it if necessary.
Check the setting of the radiation iris and readjust it if necessary.
Check the dose rate and readjust it if necessary.
Perform the IQ Quick test.
- When doing this, check the resolution first and, if necessary, readjust the optical
resolution of the Compact optics.
Complete the country-specific acceptance. (§16 Partial acceptance... /DHHS...)

Replacing the compact optics

WARNING
1
2
Fig. 5

Removing the Compact optics

Turn the C-arm so that the I.I. is at the top.
Loosen and remove the I.I. cover.
Rotate the camera in the "0" position (6/Fig. 6).
Siemens AG
Medical Engineering
Electrical voltage!
See Chapter 1, Safety Information.
Prior to removing the mini voltage supply, the system must be
switched off for at least 3 minutes until the high voltage in the
system and in the I.I. mini voltage supply dissipates.
4
3
5
3
4
Register 5
RXR2-130.061.02
Rev. 03
6
5
5
3
Fig. 6
Page 9 of 18
10.03
CS PS 24
6 - 9
6
6
2
5
4
3
1
6
SIREMOBIL Compact
System Manual