5 - 10

Collimator, X-Iris

NOTE
WARNING
Attach the center cross in the center of the image intensifier input screen.
Connect the service PC to the serial interface on board D1.
Start the service program.
Select the Adjustment, Calibrations menu.
Answer the query with "yes".
Select "X-Iris" in the "Function Groups" combo box.
In the "Available Functions:" window select "1. Init (Limits Off / Autosearch)" and click on
Execute.
Close the status window by clicking on the OK button.
For the query "Do you want to reset the unit," click "No".
The X-Iris can now be completely opened and closed with the X-Iris keys.
Select I.I. survey format.
During fluoroscopy, adjust the X-Iris by opening and closing the X-Iris keys so that the
collimator blades are still visible along the edge of the I.I.
Stop fluoroscopy.
Select "2. II. - Full Format Position" on the service PC and click on "Execute" .
Close the status window by clicking on the OK button.
For the query "Do you want to reset the unit," click "No".
Switch to zoom format.
During fluoroscopy, adjust the X-Iris by opening and closing the X-Iris keys so that the
collimator blades are still visible along the edge of the I.I.
Stop fluoroscopy.
Select "3. II. - Zoom Format Position" on the service PC and click on "Execute" .
Close the status window by clicking on the OK button.
For the query "Do you want to reset the unit," click "No".
Open the X-Iris completely.
Select "4. CEX Position" on the service PC and click on "Execute" .
For the query "Do you want to reset the unit", click "No" .
Close the X-Iris completely.
SIREMOBIL Compact
System Manual
Because of the mechanical hysteresis of the semi-transparent
slot diaphragm, open the collimator blades to maximum aperture,
prior to adjusting the radiation iris.
X-ray radiation!
See Chapter 1, Safety Information.
Protect yourself against radiation exposure. Wear a lead apron.