Replacing boards / replacing components
E Monitor split
F Time
G Date / Time setting
H Hospital name
I Storing the dose
K Dose documentation
Perform the IQ quick test.

Memoskop C

NOTE
Replace the Memoskop C.
The attachment screws are accessible on the underside of the storage device.
Set the Memoskop C to the correct video frequency and configure it for the number of
existing monitors.
Refer to Chapter 5 "Configuring the video frequency " and "MEMOSKOP - Configuration
for the number of existing monitors".
Configure the following parameters in Technical Setup for Memoskop:
3 Number of monitors
4 Hardcopy setup
5 Edge 1 coef set
6 Edge 2 coef set
7 Motion detection setup
10 Date / Time format
12 Keyboard
13 Language
15 Display of realtime clock Configure Yes. Select No only upon customer request.
16 Display of rotated images Configure Yes.
17 Monitor zoom
18 Monitor split
19 Monitor text
20 Maximum image number Configure 100.
Siemens AG
Medical Engineering
Configure horizontally. Configure vertically only upon cus-
tomer request.
Configure Yes.
Set the current date and time.
Program the hospital name.
If the area dose product meter is present, (Diamentor) con-
figure Yes. Otherwise, configure No.
If the area dose product meter is present, (Diamentor) con-
figure Yes. Otherwise configure No.
Read and record the parameters for Memoskop C if you can
select User setup. After replacing the Memoskop, configure the
Memoskop C-E according to the parameters noted.
Configure 1 for SIREMOBIL Compact with 1 monitor.
Configure 2 for SIREMOBIL Compact with 2 monitors.
Check whether 5500 is programmed in "4 Max Hardcopy
time". If not, program this value.
Check whether 3 is programmed, If not, program this value.
Check whether 8 is programmed, If not, program this value.
Check whether the following are programmed "MD1:
Threshold" 100 and for "MD2: Threshold" 120. If not, pro-
gram these values.
Set the country-specific format.
Set to the keyboard layout being used.
Set for language of the country.
Configure No.
Configure Yes for SIREMOBIL Compact with 1 monitor.
Configure No for SIREMOBIL Compact with 2 monitors.
Configure Yes.
Register 5
RXR2-130.061.02
Rev. 03
Page 15 of 18
10.03
CS PS 24
6 - 15
6
SIREMOBIL Compact
System Manual