Replacing boards / replacing components

Multispot 2000

WARNING
Before installing the new Multispot 2000, set the correct refresh rate.
Refer to Table 1.
50Hz refresh rate
Jumper
X4
X5
X6
X7
X8
X9
X10
X11
Tab. 1
Disconnect the SIREMOBIL Compact line voltage plug from the receptacle.
Replace the Multispot 2000.
The screws securing the Multispot 2000 are accessible on the underside of the camera.
Perform a camera function test after the replacement.
Generate and program the exposure parameters for the Multispot 2000 according to the
IQ Quick test (refer to the log book). Record the new exposure parameters in the IQ
Quick test instructions.

Area dose product measurement device

Remove the SIREPHOS cover (see Chapter 1, "Covers").
Replace the electronics board and the dose measurement chamber.
Perform the calibration of the area/dose product measurement device. Refer to Chapter
5 "Calibrating the Area Dose Product Measurement".
Test the accuracy of the area dose product measurement device. Refer to Chapter 5
"Testing the accuracy of the area dose product measurement device.
Re-attach the SIREPHOS cover.
Seal the SIREPHOS cover using sealing compound (see Chapter 1, "Covers").
Siemens AG
Medical Engineering
Electrical voltage!
See Chapter 1, Safety Information.
Switch the system power supply off and disconnect the power
plug.
Position
b
open
b
b
b
b
b
b
Register 5
RXR2-130.061.02
Rev. 03
60Hz refresh rate
Jumper
Position
X4
a
X5
open
X6
a
X7
a
X8
a
X9
a
X10
a
X11
a
Page 17 of 18
10.03
CS PS 24
6 - 17
6
6
SIREMOBIL Compact
System Manual