6 - 4
Reconnect the plugs between the VIDEOMED DC and the Compact optics and secure
them with cable ties. In doing this, make sure that the cable or the cable ties do not ex-
tend beyond the edge of the board and that they are located in the recess of the
VIDEOMED DC board.
Leave the I.I. in the service position.

Checks and adjustments

Switch the system on and rotate the camera from +220 degrees to -220 degrees.
When doing this, the cable in the cable spool must allow for unhindered rotation over the
entire rotation range.
Check the camera rotation and, if necessary, readjust it. The camera should be able to
rotate 220° +3° or -220° -3° (Check this using the display).
Check the dose rate and, if necessary, readjust it.
Perform the IQ Quick test. First check the resolution. If necessary, adjust the optical
sharpness at the Compact optics.
If the "Modularis" system is present, perform the isocenter check.

Basic system operating panel

Disconnect the plugs for the operating panel from board D1.
Remove the cable clamps to the operating panel.
Remove the Allen screws on the mounting ring (swivel joint of the operating panel).
Replace the operating panel .
Reinstall the mounting ring.
Reconnect the cable.
Reattach the cable clamp.
Perform a function test of the operating panel.
SIREMOBIL Iso-C
System Manual
Replacing boards / replacing components
SPR2-230.061.01
Rev. 13
Page 4 of 48
12.04
CS PS 24