5 - 14
Concluding steps
•
Switch off the system and switch it back on again.
•
Return to the main menu by pressing ESC at the service PC.
•
Select the "Data, Backup..." menu.
•
Click "parameter" in the Backup window.
•
Then click "Backup" .
•
Enter "slot diaphragm adjusted" under "Remarks for Parameter".
•
Then click "OK" . The new position values are stored to disk.
•
You can now exit from the service program.
Collimator displays on the monitor
Prerequisites
•
Camera rotation, collimators for X-Iris and slot diaphragm must be set correctly.
WARNING
Collimator display, X-Iris
•
Connect the service PC to the serial interface of the SIREMOBIL.
•
Start the service program.
•
Select the "Adjustment, Calibrations..." menu.
•
Respond to the query with "Yes".
•
Select "Screen Display COLLIMATOR" in the "Function Groups:" combo box.
Start adjustment
•
In the "Available Functions:" window select "Init" and click on "Execute". The X-Iris is au-
tomatically opened to full format and the slot diaphragm is opened to maximum aperture.
Full format, X-Iris open
•
Zoom must be disabled.
•
Release fluoro and wait until generator has performed automatic adjustment.
•
Switch fluoro off. The X-Iris is displayed on the monitor.
•
Use the keys for opening and closing the X-Iris to set the diameter of the superimposed
circle to the size of the X-iris.
•
In the "Available Functions:" window select "Open Pos - II. Full Format" and click on "Ex-
ecute".
•
Close the status window by clicking on the OK button.
SIREMOBIL Iso-C
System Manual
X-ray radiation!
See Chapter 1, Safety Information.
Use radiation protection! Wear a lead apron!
SPR2-230.061.01
Rev. 13