5 - 14

Concluding steps

Switch off the system and switch it back on again.
Return to the main menu by pressing ESC at the service PC.
Select the "Data, Backup..." menu.
Click "parameter" in the Backup window.
Then click "Backup" .
Enter "slot diaphragm adjusted" under "Remarks for Parameter".
Then click "OK" . The new position values are stored to disk.
You can now exit from the service program.

Collimator displays on the monitor

Prerequisites

Camera rotation, collimators for X-Iris and slot diaphragm must be set correctly.
WARNING

Collimator display, X-Iris

Connect the service PC to the serial interface of the SIREMOBIL.
Start the service program.
Select the "Adjustment, Calibrations..." menu.
Respond to the query with "Yes".
Select "Screen Display COLLIMATOR" in the "Function Groups:" combo box.

Start adjustment

In the "Available Functions:" window select "Init" and click on "Execute". The X-Iris is au-
tomatically opened to full format and the slot diaphragm is opened to maximum aperture.

Full format, X-Iris open

Zoom must be disabled.
Release fluoro and wait until generator has performed automatic adjustment.
Switch fluoro off. The X-Iris is displayed on the monitor.
Use the keys for opening and closing the X-Iris to set the diameter of the superimposed
circle to the size of the X-iris.
In the "Available Functions:" window select "Open Pos - II. Full Format" and click on "Ex-
ecute".
Close the status window by clicking on the OK button.
SIREMOBIL Iso-C
System Manual
X-ray radiation!
See Chapter 1, Safety Information.
Use radiation protection! Wear a lead apron!
SPR2-230.061.01
Rev. 13
Adjustments / Programs
Page 14 of 30
12.04
CS PS 24
5
5
5
5
5
5
Siemens AG