6 - 36
3
1
1
Fig. 9
Video switcher
NOTE
NOTE
•
Switch the system off.
•
Loosen the four fastening screws of the securing bracket (1/Fig. 9).
•
Loosen the two clamping screws, but do not unscrew them completely (2/Fig. 9).
•
Tip the holding plate with the video switchers into the monitor trolley compartment
(see Fig. 10).
•
Unplug the connectors of the video switcher to be replaced.
•
Pull the video switcher to be replaced out of the holding plate from the side.
•
Depress the green button on the video switcher.
•
Push the new video switcher into the holding plate and reconnect all connectors.
•
Screw the holding plate back onto the monitor trolley.
•
Tighten the two clamping screws again.
•
Switch the system on.
•
After the system run-up, the SIMOMED HM Monitor (reference monitor) should display
the MEMOSKOP reference image. If the run-up process of the 3D reconstruction PC or
the syngo interface appears, the green button of video switcher 1 is in the wrong position
and needs to be switched over. (3/Fig. 9) points to video switcher 1.
•
Perform the IQ quick test.
SIREMOBIL Iso-C
System Manual
Replacing boards / replacing components
1
2
1
Only applies to SIREMOBIL Iso-C 3D.
Video switcher 1 (3/Fig. 9) is used to switch the video signal
between the MEMOSKOP and the 3D reconstruction PC.
Video switcher 2 (1/Fig. 10) is used to route the video signal to the
video printer.
SPR2-230.061.01
Rev. 13
Fig. 10
Page 36 of 48
12.04
CS PS 24
1
6
Siemens AG
Medical Solutions