6 - 24
MEMOSKOP C-E
NOTE
NOTE
Replace the MEMOSKOP C-E.
The attachment screws are on the underside of the storage device.
Check the software status of the MEMOSKOP. Download the system software to the
MEMOSKOP if necessary.
Use the MEMOSKOP Service Software to import the language file.
Configure the MEMOSKOP C-E for the correct video frequency and the number of moni-
tors present.
Refer to chapter 5 "MEMOSKOP C-E without Keyboard, Emulation ...", "Configuring the
Video frequency" and "MEMOSKOP-Configuration for the number of existing monitors".
Configure the following parameters in Technical Setup for the MEMOSKOP:
Number of Monitors
Hardcopy
Edge 1 coef set
Edge 2 coef set
Motion detection setup
Date / Time format
Keyboard
Display of realtime clock
Display of rotated images
Monitor zoom
Default LUT
Default Edge
Monitor split
Monitor text
SIREMOBIL Iso-C
System Manual
Replacing boards / replacing components
Check the software status of the newly delivered MEMOSKOP.
If the software status of the MEMOSKOP differs from that of the
system, the system software status must be downloaded to the
MEMOSKOP.
The MEMOSKOP software status can be read out in the 2nd line
of the User Setup.
The first number describes the MEMOSKOP type.
(2=CE; 3= CE-100/ C/ C & MOD; 4=C-SUB/ C-SUB & MOD).
Read and record the parameters for MEMOSKOP C-E if you are
able to select User setup. After replacing the Memoskop, con-
figure the MEMOSKOP C-E according to the parameters recorded.
Configure 1 for SIREMOBIL with 1 monitor.
Configure 2 for SIREMOBIL with 2 monitors.
Check whether 5000 is programmed in "4 Max Hardcopy
time". If not, program this value.
Check whether 3 is programmed, if not, program this value.
Check whether 8 is programmed, If not, program this value.
Check whether the following are programmed MD1:
Threshold 100 and at MD2: Threshold 120.
If not, program these values.
Set to country-specific format.
Set to the keyboard layout being used.
Configure Yes. Configure No only upon customer request.
Configure Yes.
Configure No.
Configure 2. Program a different LUT curve only upon cus-
tomer request.
Configure 1. Program a different Edge enhancement value
only upon customer request.
Configure Yes.
Configure Yes.