Memoskop C-E 100 - Siemens SIREMOBIL Iso-C Troubleshooting Manual

Table of Contents

Advertisement

108

MEMOSKOP C-E 100

NOTE
NOTE
Replace the MEMOSKOP C-E 100. The attachment screws are accessible on the
underside of the storage device.
Check the software status of the MEMOSKOP. Download the system software to the
MEMOSKOP if necessary.
Use the MEMOSKOP Service Software to import the language file.
Set the MEMOSKOP C-E 100 to the correct video frequency and configure it for the
number of existing monitors.
See
(Configuring the video frequency / p. 77)
number of existing monitors / p.
Configure the following parameters in Technical Setup for MEMOSKOP:
Number of Monitors
Hardcopy
Edge 1 coef set
Edge 2 coef set
Motion detection setup
Date / Time format
Keyboard
SIREMOBIL Iso-C
Replacing boards and components
Check
the
software
MEMOSKOP. If the software status of the MEMOSKOP differs
from that of the system, the system software status must be
downloaded to the MEMOSKOP. The MEMOSKOP software
status can be read out in the 2nd line of the User Setup. The
first number describes the MEMOSKOP type.
CE-100/ C/ C & MOD; 4=C-SUB/ C-SUB & MOD).
The MEMOSKOP C-E 100 image storage device is a new
image storage device that works only with the correct num-
ber of images. For this reason, if the image storage was set
to its default values (Menu "Diagnostics", "Load Factory
Defaults"), the parameter "Maximum image numbers" must
be set to 100 images.
Read and record the parameters for MEMOSKOP C-E 100 if
you can select User Setup. After replacing the MEMOSKOP,
configure the MEMOSKOP C-E 100 according to the parame-
ters noted.
77)."
Configure 1 for SIREMOBIL with 1 monitor. Configure 2 for
SIREMOBIL with 2 monitors.
Check whether 5000 is programmed in "4 Max Hardcopy
time". If not, program this value.
Check whether 3 is programmed; if not, program this value.
Check whether 8 is programmed, If not, program this value.
Check whether the following are programmed "MD1:
Threshold" 100 and for "MD2: Threshold" 120. If not, pro-
gram these values.
Set the country-specific format.
Set to the keyboard layout being used.
SPR2-230.840.01.01.02
08.08
status
of
the
and
(MEMOSKOP, configuration for the
Page 108 of 148
CS PS SP
6.14
newly
delivered
(2=CE; 3=
© Siemens, 2008