Siemens SIREMOBIL Iso-C Troubleshooting Manual page 127

Table of Contents

Advertisement

Replacing boards and components
Reattach the mounting plate including the motor. In doing so, place the toothed belt
over the pulley again, ensuring that the actual value potentiometer and its pulley are not
damaged or bent.
Retighten the toothed belt of the orbital drive
thumb pressure on the opposite side, the belt should give approx. 2 to 3 mm.
Perform a functional check of the orbital drive. To do this, perform a 3D scan.
© Siemens, 2008
For internal use only
SPR2-230.840.01.01.02
08.08
((1/Fig. 40 / p.
126)). When depressed by
Page 127 of 148