Memoskop C-Sub/ Memoskop C-Sub & Mod - Siemens SIREMOBIL Iso-C Troubleshooting Manual

Table of Contents

Advertisement

114
MEMOSKOP C-SUB/ MEMOSKOP C-SUB & MOD
NOTE
NOTE
Check the software status of the MEMOSKOP. Download the system software to the
MEMOSKOP if necessary.
Use the MEMOSKOP Service Software to import the language file.
Replace the MEMOSKOP C-SUB.
The attachment screws are accessible on the underside of the storage device.
Set the MEMOSKOP C-SUB to the correct video frequency and configure it for the
number of existing monitors.
See
(Configuring the video frequency / p. 77)
number of existing monitors / p.
Configure the following parameters in Technical Setup for MEMOSKOP:
Number of Monitors
Hardcopy
Edge 1 coef set
Edge 2 coef set
Motion detection setup
Date / Time format
Keyboard
Language
Display of realtime clock
Display of rotated images
SIREMOBIL Iso-C
Replacing boards and components
Check
the
software
MEMOSKOP. If the software status of the MEMOSKOP differs
from that of the system, the system software status must be
downloaded to the MEMOSKOP. The MEMOSKOP software
status can be read out in the 2nd line of the User Setup. The
first number describes the MEMOSKOP type.
CE-100/ C/ C & MOD; 4=C-SUB/ C-SUB & MOD).
Read and record the parameters for MEMOSKOP C-SUB if
you can select User setup. After replacement, configure the
MEMOSKOP C-SUB, according to the parameters recorded.
77).
Configure 1 for SIREMOBIL with 1 monitor. Configure 2 for
SIREMOBIL with 2 monitors.
Check whether 5000 is programmed in "4 Max Hardcopy
time". If not, program this value.
Check whether 3 is programmed; if not, program this
value.
Check whether 8 is programmed, If not, program this
value.
Check whether the following are programmed "MD1:
Threshold" 100 and for "MD2: Threshold" 120. If not, pro-
gram these values.
Set the country-specific format.
Set to the keyboard layout being used.
Set for language of the country.
Configure Yes. Select No only upon customer request.
Configure Yes.
SPR2-230.840.01.01.02
08.08
status
of
the
and
(MEMOSKOP, configuration for the
Page 114 of 148
CS PS SP
6.16
newly
delivered
(2=CE; 3=
© Siemens, 2008
For internal use only

Advertisement

Table of Contents
loading

Table of Contents