Replacing boards and components
Enable Bridge
Delay 3D-PC/ DICOM
Bridge
Interface to 3D-PC
Configure the following parameters in User Setup for MEMOSKOP:
Operating mode parame-
ters
VCR output
Hardcopy Release
Text monitor
Split screen
Display of time
Set Date / Time
Hospital Name
Storage of Dose on disk
Dose on Hardcopy output
Perform the IQ quick test.
NOTE
© Siemens, 2008
For internal use only
If the option "3D Reconstruction" or the option "DICOM
Bridge" is installed or a MOD drive is present, configure
Yes. Otherwise configure No.
Only available if option "3D Reconstruction" is installed.
Configure 120.
Only available if option "3D Reconstruction" is installed.
Configure Yes.
Change the programmed parameters only upon customer
request.
Configure to monitor A.
For SIREMOBIL with 1 monitor: Configure to monitor A. For
SIREMOBIL with 2 monitors: Configure to monitor B.
For SIREMOBIL with 1 monitor: Configure to monitor A. For
SIREMOBIL with 2 monitors: Configure to both monitors.
Configure horizontally. Configure vertically only upon cus-
tomer request.
Configure Yes.
Set the current date and time.
Enter the hospital name.
If dose-area product measurement device is present (Dia-
mentor), configure Yes. Otherwise configure No.
If dose-area product measurement device is present (Dia-
mentor), configure Yes. Otherwise configure No.
With present MODULARIS system, the parameters for the
Litho program have to be set again according to the LITHOS-
TAR MODULARIS Service Instructions.
SPR2-230.840.01.01.02
08.08
Page 113 of 148
CS PS SP
113