Default Edge
Monitor split
Monitor text
Maximum image number
Disable dose rate selection
in user setup
Unit serial number
Product name
Default program change-
able
Software backward com-
patible
Configure the following parameters in User Setup for MEMOSKOP:
Operating mode parame-
ters
VCR output
Hardcopy Release
Text monitor
Split Screen
Display of time
Set Date / Time
Hospital Name
Storage of Dose on disk
Dose on Hardcopy output
Perform the IQ quick test.
SIREMOBIL Iso-C
Replacing boards and components
Configure 1. Program a different edge enhancement value
only upon customer request.
Configure Yes.
Configure Yes.
n.a.
Program for "No" when equipped with an dose-area product
measurement device. The menu "Storage of Dose on disk"
and the menu "Dose on Hardcopy output" are visible in
User Setup. Otherwise set to "Yes".
Program the serial number of the SIREMOBIL. Do not enter
the leading zero.
Program "SIREMOBIL Iso-C".
Default: Yes. Depending on country-specific regulations
configure Yes or No.
Configure "No".
Change the parameters programmed for the organ pro-
gram upon customer request.
Configure to Monitor A.
For SIREMOBIL with 1 monitor: Configure to monitor A. For
SIREMOBIL with 2 monitors: Configure to monitor B.
For SIREMOBIL with 1 monitor: Configure to monitor A. For
SIREMOBIL with 2 monitors: Configure to both monitors.
Configure horizontally. Configure vertically only upon cus-
tomer request.
Configure Yes.
Set the current date and time.
Enter the hospital name.
If the dose-area product measurement device is present
(Diamentor), configure Yes. Otherwise, configure No.
If the dose-area product measurement device is present
(Diamentor), configure Yes. Otherwise configure No.
SPR2-230.840.01.01.02
08.08
Page 106 of 148
CS PS SP
© Siemens, 2008