Siemens SIREMOBIL Iso-C Troubleshooting Manual page 63

Table of Contents

Advertisement

Adjustments / Programs
NOTE
Connect the service PC to the serial interface of the SIREMOBIL Iso-C.
Start the service program.
Select the Adjustment, Calibrations menu.
Answer the query "yes".
In the Combo box "Function Groups:" select "Camera Rotation" .
In the window "Available Functions:" select "1. Init (Disable Limits)" and click "Execute".
For the query "Do you want to reset the Unit", click "No".
Attach the center cross or equivalent to the image intensifier.
X
Start fluoro and rotate the camera to the +180° position (with the camera rotation key
clockwise). The image contents should rotate counterclockwise.
Stop fluoro, in the window "Available Functions:" select "2. +180 Degree Position" and
click "Execute".
For the query "Do you want to reset the Unit", click "No".
X
Start fluoro and set camera rotation to the 0° position. The image (left/right / up/down)
corresponds to
Stop fluoro, in the window "Available Functions:", select "3. 0 Degree Position" and click
"Execute" .
For the query "Do you want to reset the Unit", click "No".
X
Start fluoro and rotate the camera to the -180° position (with the counterclockwise cam-
era rotation key). The image contents will rotate clockwise.
Stop fluoro, in the window "Available Functions:", select "4. -180 Degree Position" and
click "Execute" .
For the query "Do you want to reset the Unit", click "No".
In the window "Available Functions:" , select "5. Init Off & Store Positions" and click
"Execute".
For the query "Do you want to reset the Unit", click "Yes".
Return to the service program main menu with the ESC key.
Select the Data, Backup... menu.
Select Parameter in the Backup window.
Click "Backup".
In "Remarks for Parameter", enter "camera rotation set".
Click "OK". The new position values for camera rotation will be stored to the disk.
Switch the SIREMOBIL Iso-C off and on again.
© Siemens, 2008
For internal use only
With the right camera rotation key (the arrow on the key is
pointing clockwise), the marker on the monitor will rotate
clockwise. The image contents will rotate counterclockwise.
With the left camera rotation key (the arrow on the key is
pointing counterclockwise), the marker superimposed on the
monitor will rotate counterclockwise. The image contents will
rotate clockwise.
(Fig. 20 / p.
62).
SPR2-230.840.01.01.02
08.08
Page 63 of 148