Replacing boards / replacing components

Dose rate

Check the dose rate and, if necessary, adjust it.

Checking the camera optics centering at the I.I. output

WARNING
Mark the center of the I.I. input as precisely as possible using a small washer and adhe-
sive tape.
If there is a laser light localizer, the I.I. attachment with the lead cross can also be used.
Select the survey format.
Move the camera into the 0° position.
Switch fluoro on briefly.
Mark the center of the washer or lead cross on the monitor (refer to Fig. 9).
Rotate the camera in the -220° position.
Mark the center again on the monitor (refer to Fig. 9).
Rotate the camera in the in +220° position.
Mark the center again on the monitor (refer to Fig. 9).
Place an 8 mm washer over the three markings.
All three markings must be located within the 8mm diameter of the washer. An 8 mm
washer has a somewhat larger inner diameter of 8.4 mm, therefore the markings must
clearly lie within the inner diameter and should not touch the edge of the washer.
If the diameter is e 8mm, the following centering adjustment does not have to be per-
formed.
Continue with "checks and adjustments".
If the markings exceed > 8 mm, proceed as follows.
Siemens AG
Medical Solutions
X-ray radiation!
See Chapter 1, Safety Information.
Use radiation protection! Wear a lead apron.
SPR2-230.061.01
Rev. 13
Page 21 of 48
12.04
CS PS 24
6 - 21
6
6
SIREMOBIL Iso-C
System Manual