Prerequisites

Attaching the cover

Re-fasten or connect the protective ground wire, the cable of the keypad (brake buttons)
and the cable of the emergency stop button.
CAUTION
Expand the cover at the rear end (see Fig. 2) far enough that, as shown in Fig. 3, it lies
obliquely over the horizontal carriage in the edge, 1/Fig. 3.
Press on the top surface of the cover without grasping the sheet metal edges, until this
slides over the lower edge of the horizontal slide. Here, make sure that the cover does
not slide out from the edge (1 / Fig. 3).
Check the operation of the emergency stop button and of the brake buttons.
Fasten the cover with the 6 Allen screws (1/ Fig. 2).
MODULARIS only:
Tightening force of the Docking Plate 18 Nm must be observed (see service instructions
for MODULARIS)
Siemens AG
Medical Solutions
Danger of injury when dismantling / assembling mechanical
parts! Non-compliance can lead to minor up to moderately severe
injuries, especially to the hands!
When the cover is attached and pressed down, it is under
mechanical stress and snaps into place over the bottom edge of
the horizontal carriage.
Expand the cover at the rear end (see Fig. 2) far enough that, as
shown in Fig. 3, it lies obliquely over the horizontal carriage in the
edge, 1/Fig. 3.
Press on the top surface of the cover without grasping the sheet
metal edges, until this slides over the lower edge of the horizontal
slide.
SPR2-230.061.01
Rev. 13
Page 13 of 18
12.04
CS PS 24