I.I. and Compact optics
NOTE
Firmly grasping the I.I., swivel the C-arm so that the I.I. is below and the SIREPHOS is
above.
Place the I.I. carefully on the C-arm flange.
Remove both threaded pins from the I.I. housing.
Lift the I.I. toward the top off the C-arm, turn it around and place it on a suitable surface.
(I.I. is standing on the grid.)
Remove the retainer clip for attaching the plugs and cable guide from the I.I.
Remove the Compact optics and place them on a dust-free surface.
Reattach the Compact optics and the retainer to the new I.I.
If the "MODULARIS" system or a laser light localizer are present, remove the centering
disk attached above the grid.
Remove the grid. Desolder the wires for the cassette contact.

Installation

Compact optics

NOTE
Reattach the Compact optics to the new I.I.
When doing this, thread the anode wire through the recess in the optics.
The camera optics must lie against the eccentric screws.
Reinstall the O-ring and the cap nut for the anode cable.
The cable length should be 70 mm from the beginning of the white wire to the cap nut.
Reattach the retainer clip for attaching the plug and the cable guide to the I.I.
SIREMOBIL Iso-C
System Manual
Replacing boards / replacing components
When removing the Compact optics, make sure that no dust or
dirt particles contaminate the I.I. output window.
Do not loosen the eccentric screws (2/Fig. 7) on the edge of the
Compact optics!
These are used to center the Compact optics to the I.I. output
screen.
When installing the Compact optics and the VIDEOMED DC, make
sure that no dust or dirt particles contaminate the I.I. output win-
dow or the Compact optics.
SPR2-230.061.01