Adjustments / Programs
Fig. 3
Fig. 4

Camera rotation

WARNING
NOTE
NOTE
Siemens AG
Medical Solutions
X-ray radiation!
See Chapter 1, Safety Information.
Observe the need for radiation protection. Wear a lead apron.
The keys for horizontal and vertical image reversal must be off
prior to adjusting camera rotation. Camera rotation should be set
in the 0° position. Fig. 3, Fig. 4 and Fig. 5 shows the 0° position.
With the right camera rotation key (the arrow on the key is point-
ing clockwise), the marker on the monitor will rotate clockwise.
The image contents will rotate counterclockwise.
WIth the left camera rotation key (the arrow on the key is pointing
counterclockwise), the marker superimposed on the monitor will
rotate counterclockwise. The image contents will rotate clock-
wise.
SPR2-230.061.01
Rev. 13
Patient left position
Patient right position
Fig. 5
Page 19 of 30
12.04
CS PS 24
5 - 19
5