Frequency Converter Fu (Vertical Lift) - Siemens SIREMOBIL Iso-C Service Instructions Manual

Hide thumbs Also See for SIREMOBIL Iso-C:
Table of Contents

Advertisement

Replacing boards / replacing components
M13
FU
Fig. 2

Frequency converter FU (Vertical lift)

WARNING
Remove the Plexiglas cover over the plug field/clamp field of the frequency converter.
Disconnect the power plug.
Remove the cable clamps and cable ties at and in front of the frequency converter.
Disconnect all the cables at the frequency converter after you have documented this.
Replace the frequency converter after removing the Allen screws.
(Refer to arrows, Fig. 2.)
NOTE
Reconnect all the cables.
Reconnect the power plug.
Secure the cables with cable ties.
Reattach the cables with the cable clamps.
Perform a function test of the column vertical lift.
Siemens AG
Medical Solutions
M14
Electrical voltage!
See Chapter 1, Safety Information.
Switch system power supply off and disconnect the power plug.
Prior to installing the new frequency converter, remove the bridge
on X5 between PTC and -Uzk.
SPR2-230.061.01
Rev. 13
Page 5 of 48
12.04
CS PS 24
6 - 5
6
SIREMOBIL Iso-C
System Manual