Prerequisites
Fig. 8
I.I. service position
WARNING
Moving the I.I. into the service position
•
Take the service screws from the kit supplied.
•
Move the C-arm into the vertical position, with the I.I. below and the SIREPHOS above.
•
Remove both attachment screws for the I.I. and replace them with the service screws.
•
Insert the service screws at least 5 turns into the threaded holes of the I.I.
•
Hold the I.I. firmly and turn the C-arm so that the I.I. is above and the SIREPHOS is be-
low (Angulation).
•
Hold the I.I. firmly and carefully remove it from the C-arm. Be careful not to damage the
cable.
Installing the I.I.
•
Move the C-arm into the vertical position with the I.I. below, the SIREPHOS above.
•
Holding the I.I. firmly, carefully lift it up. Be careful of the cable.
•
Holding the I.I. firmly, move the C-arm so that the Sirephos is above and the I.I. is below
(Angulation).
•
Carefully insert the I.I. into the C-arm. When doing this, be careful not to damage the ca-
ble.
•
Remove both the service screws from the threaded holes of the I.I. and reinstall both the
attachment screws.
Siemens AG
Medical Solutions
Risk of injuries on mechanical parts!
If not observed, slight to medium injuries, especially to the hands,
can occur.
Observe the notes concerning the C-arm carbon fiber structure.
SPR2-230.061.01
Rev. 13
Page 17 of 18
12.04
CS PS 24
1 - 17
1
1
1
SIREMOBIL Iso-C
System Manual