Adjustments / Programs
•
The mechanical limits are checked and the position values are automatically stored.
•
Camera rotation is automatically moved to the 0 degree position.
•
The X-Iris is opened to max.aperture.
•
Switch off Zoom .
•
Attach a centering cross or equivalent object to the I.I. for evaluating the position of the
fluoro image.
WARNING
•
Switch on fluoro and move the slot diaphragm to the 0 degree position using the keys for
rotating the slot diaphragm. The blades should be in the vertical on the monitor screen.
•
Switch off fluoro.
•
In the window "Available Functions:" select "0 DEGREE POSITION" and click "Execute".
•
Close the status window by clicking on the OK button.
•
Respond to the query "Do you want to reset the unit" by clicking "No" .
•
Switch on fluoro.
•
Open the slot diaphragm to where the blades begin to disappear at the edge of the im-
age.
•
Switch off fluoro.
•
In the window "Available Functions:" select "FULL FORMAT POSITION" and click "Exe-
cute".
•
Close the status window by clicking on the OK button.
•
Respond to the query "Do you want to reset the unit" by clicking "No".
•
Select the zoom format.
•
Switch on fluoro.
•
Open the slot diaphragm to where the blades begin to disappear at the edge of the im-
age.
•
Switch off fluoro.
•
In the window "Available Functions:" select "ZOOM FORMAT POSITION" and click "Ex-
ecute".
•
Close the status window by clicking on the OK button.
•
Respond to the query "Do you want to reset the unit" by clicking "No" .
•
In the window "Available Functions:" select "Enable circle mask" and click "Execute"
•
Respond to the query "Do you want to reset the unit" by clicking "No" .
•
In the window "Available Functions:" select "INIT OFF & STORE POSITION" and click
"Execute".
•
Close the status window by clicking on the OK button.
•
Respond to the query "Do you want to reset the unit" by clicking "Yes". The adjustment
values are stored.
Siemens AG
Medical Solutions
X-ray radiation!
See Chapter 1, Safety Information.
Observe the need for radiation protection. Wear a lead apron.
SPR2-230.061.01
Rev. 13
Page 13 of 30
12.04
CS PS 24
5 - 13
SIREMOBIL Iso-C
System Manual