6 - 8
Fig. 3
Remove the cover on the front side of the C-arm by removing the screws (I.I. side, refer
to Fig. 3).
Move the I.I. into the service position. Refer to chapter 1, I.I. service position.
Remove both attachment screws for the control section in the I.I. flange of the C-arm (ar-
row/Fig. 4).
Carefully remove the control section from the C-arm and disconnect all the plugs.
Disconnect the protective conductor.
Remove the VIDEOMED DC camera from the I.I.
Refer to Replacing boards / Replacing components VIDEOMED DC.
SIREMOBIL Iso-C
System Manual
Replacing boards / replacing components
SPR2-230.061.01
Rev. 13
Fig. 4
Page 8 of 48
12.04
CS PS 24