7 - 8
Guide cylinder
Fig. 8
Emergency off switch
•
Loosen the cover above the horizontal carriage.
•
Unsolder the emergency off switch.
•
Replace the emergency off switch.
•
Solder the wires on the emergency off switch.
•
Reinstall the cover on the horizontal carriage.
•
Check functioning of the emergency off switch.
- Emergency off switch pressed: vertical lift must not be possible.
- Emergency off switch released: vertical lift must be possible.
Lubricating the lifting column
Lubricate the spindle of the lifting column and the mounting of the guide cylinder once a
year.
Spindle
•
Lubricate the spindle with approximately 10 - 20 ml of "Optimol, Optipit" special oil.
Guide cylinder
•
Grease the guide cylinder with approximately 5 - 10 ml of special bearing grease
"Slick Pac PTFE" .
•
Then move the lifting column through the entire travel range twice to ensure that the
grease is evenly distributed.
SIREMOBIL Iso-C
System Manual
Spindle
SPR2-230.061.01
Rev. 13
12.04
Brakes / Movements
Page 8 of 10
CS PS 24
7
7
7
7
Siemens AG
Medical Solutions