7 - 8

Guide cylinder

Fig. 8

Emergency off switch

Loosen the cover above the horizontal carriage.
Unsolder the emergency off switch.
Replace the emergency off switch.
Solder the wires on the emergency off switch.
Reinstall the cover on the horizontal carriage.
Check functioning of the emergency off switch.
- Emergency off switch pressed: vertical lift must not be possible.
- Emergency off switch released: vertical lift must be possible.

Lubricating the lifting column

Lubricate the spindle of the lifting column and the mounting of the guide cylinder once a
year.

Spindle

Lubricate the spindle with approximately 10 - 20 ml of "Optimol, Optipit" special oil.
Guide cylinder
Grease the guide cylinder with approximately 5 - 10 ml of special bearing grease
"Slick Pac PTFE" .
Then move the lifting column through the entire travel range twice to ensure that the
grease is evenly distributed.