Replacing boards / replacing components
•
Reattach the mounting plate incl. motor. In doing so, place the toothed belt over the pul-
ley again, ensuring that the actual value potentiometer and its pulley are not damaged or
bent.
•
Retighten the toothed belt of the orbital drive (1/Fig. 14). When depressed by thumb
pressure on the opposite side, the belt should give approx. 2 to 3 mm.
•
Perform a function check of the orbital drive. To do this, perform a 3D scan.